The Tribology of Explanted Hip Resurfacings Following Early Fracture of the Femur
Source abstract
Abstract
A recognized issue related to metal-on-metal hip resurfacings is early fracture of the femur. Most theories regarding the cause of fracture relate to clinical factors but an engineering analysis of failed hip resurfacings has not previously been reported. The objective of this work was to determine the wear volumes and surface roughness values of a cohort of retrieved hip resurfacings which were removed due to early femoral fracture, infection and avascular necrosis (AVN). Nine resurfacing femoral heads were obtained following early fracture of the femur, a further five were retrieved due to infection and AVN. All fourteen were measured for volumetric wear using a co-ordinate measuring machine. Wear rates were then calculated and regions of the articulating surface were divided into “worn” and “unworn”. Roughness values in these regions were measured using a non-contacting profilometer. The mean time to fracture was 3.7 months compared with 44.4 months for retrieval due to infection and AVN. Average wear rates in the early fracture heads were 64 times greater than those in the infection and AVN retrievals. Given the high wear rates of the early fracture components, such wear may be linked to an increased risk of femoral neck fracture.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Orthopedic Infections and Treatments · Total Knee Arthroplasty Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Biomedical Engineering and Mechanics, Virginia Tech, Blacksburg, VA 24061, USA
North Tees Explant Centre (NTEC), Farndale House, University Hospital of North Tees, England TS19 8PE, UK
Joint Replacement Unit, University Hospital of North Tees, England TS19 8PE, UK
Department of Orthopaedics & Trauma, Queen Elizabeth University Hospital, Glasgow G51 4TF, UK
School of Mechanical and Systems Engineering, Newcastle University, England NE1 7RU, UK
References (47)
Show source-provided references
1. (2014). National Joint Registry for England, Wales and Northern Ireland, Pad Creative Ltd.. 11th Annual Report.
2. Metal-on-metal hip joint tribology10.1243/095441105X69114 ↗
3. Analysis of 118 second-generation metal-on-metal retrieved hip implants10.1302/0301-620X.81B1.0810046 ↗
4. The problem is osteolysis
5. Influence of simulator kinematics on the wear of metal-on-metal hip prostheses10.1243/0954411011533481 ↗
6. Ten different hip resurfacing systems: Biomechanical analysis of design and material properties10.1007/s00264-008-0607-y ↗
7. Elastohydrodynamic lubrication analysis of metal-on-metal hip-resurfacing prostheses10.1016/S0021-9290(02)00422-0 ↗
8. Tribological analysis of failed resurfacing hip prostheses and comparison with clinical data10.1243/13506501JET484 ↗
9. Elastohydrodynamic lubrication of elliptical contacts for materials of low elastic modulus. I: Fully flooded conjunction
10. Tribological principles in metal-on-metal hip joint design10.1243/095441105X63255 ↗
11. Metal-on-metal resurfacing of the hip in patients under the age of 55 years with osteoarthritis10.1302/0301-620X.86B2.14600 ↗
12. Wienroth, M., McCormack, P., and Joyce, T. (2014). Precaution, governance and the failure of medical implants: The ASR™ hip in the UK. Life Sci. Soc. Policy, 10.10.1186/s40504-014-0019-2 ↗
13. Blood metal ion concentrations after hip resurfacing arthroplasty: A comparative study of articular surface replacement and Birmingham hip resurfacing arthroplasties10.1302/0301-620X.91B10.22308 ↗
14. The Durom large diameter head acetabular component: Early results with a large-diameter metal-on-metal bearing10.1302/0301-620X.92B2.22653 ↗
15. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip10.1302/0301-620X.90B9.20785 ↗
16. Reduced articular surface of one-piece cups: A cause of runaway wear and early failure10.1007/s11999-010-1383-8 ↗
17. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement
18. Langton, D.J., Sidaginamale, R.P., Joyce, T.J., Natu, S., Blain, P., Jefferson, R.D., Rushton, S., and Nargol, A.V.F. (2013). The clinical implications of elevated blood metal ion concentrations in asymptomatic patients with MoM hip resurfacings: A cohort study. BMJ Open, 3.10.1136/bmjopen-2012-001541 ↗
19. Australian_Orthopaedic_Association (2014). National Joint Replacement Registry, Commonwealth Government. Annual Report 2014.
20. Birmingham hip arthroplasty: Five to eight years of prospective multicenter results10.1016/j.arth.2008.07.016 ↗
21. Birmingham hip resurfacing arthroplasty: A minimum follow-up of five years10.1302/0301-620X.87B2.15030 ↗
22. Birmingham hip resurfacing: A minimum follow-up of ten years10.1302/0301-620X.93B1.24134 ↗
23. Australian_Orthopaedic_Association (2010). National Joint Replacement Registry, Commonwealth Government. Annual Report 2010.
24. Metal-on-metal hybrid surface arthroplasty: Two to six-year follow-up study10.2106/00004623-200401000-00006 ↗
25. The nonoperative management of periprosthetic fractures associated with the Birmingham hip resurfacing procedure10.1016/j.arth.2004.08.007 ↗
26. Femoral neck fractures following Birmingham hip resurfacing: A national review of 50 cases10.1302/0301-620X.87B4.15498 ↗
27. The five-year results of the Birmingham hip resurfacing arthroplasty: An independent series10.1302/0301-620X.90B4.19648 ↗
28. Femoral neck fractures after metal-on-metal total hip resurfacing: A prospective cohort study10.1016/j.arth.2007.05.017 ↗
29. Risk factors affecting outcome of metal-on-metal surface arthroplasty of the hip10.1097/00003086-200401000-00015 ↗
30. A study of implant failure in metal-on-metal surface arthroplasties10.1097/01.blo.0000238777.34939.82 ↗
31. A study of the wear of explanted metal-on-metal resurfacing hip prostheses10.1016/j.triboint.2010.04.006 ↗
32. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear10.1302/0301-620X.92B1.22770 ↗
33. Influence of cup orientation on the wear performance of metal-on-metal hip replacements10.1243/09544119JEIM518 ↗
34. Tribology and wear of metal-on-metal hip prostheses: Influence of cup angle and head position10.2106/JBJS.H.00485 ↗
35. The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements10.1243/0954411011533724 ↗
36. Characterization of the running-in period in total hip resurfacing arthroplasty: An in vivo and in vitro metal ion analysis10.2106/JBJS.H.00437 ↗
37. The effect of acetabular cup position on wear of a large-diameter metal-on-metal prosthesis studied with a hip joint simulator10.1016/j.triboint.2012.10.011 ↗
38. The effect of “running-in” on the tribology and surface morphology of metal-on-metal Birmingham hip resurfacing device in simulator studies10.1243/09544119JEIM63 ↗
39. Quantification of self-polishing in vivo from explanted metal-on-metal total hip replacements10.1016/j.triboint.2010.04.007 ↗
40. Clinical and radiographic results of metal-on-metal hip resurfacing with a minimum ten-year follow-up10.2106/JBJS.I.01715 ↗
41. Volumetric wear assessment of failed metal-on-metal hip resurfacing prostheses10.1016/j.wear.2011.07.009 ↗
42. Adverse reaction to metal debris following hip resurfacing: The influence of component type, orientation and volumetric wear10.1302/0301-620X.93B2.25099 ↗
43. Reducing metal ion release following hip resurfacing arthroplasty10.1016/j.ocl.2011.01.006 ↗
44. Roughness parameters10.1016/S0924-0136(02)00060-2 ↗
45. The effects of material combination and lubricant on the friction of total hip prostheses10.1016/S0043-1648(00)00377-X ↗
46. Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hip10.1302/0301-620X.89B4.18054 ↗
47. (v) Biotribology10.1016/j.cuor.2005.09.005 ↗