Experimental Biomechanical Analysis of the Bone-to-Implant Connection in Single-Piece Implants
Source abstract
Abstract
The mechanical properties of dental implants are critical for their durability. The purpose of this study was to determine the maximum force required to induce full pull-out of a titanium implant from the bone and to characterize the mechanical behavior during this process. First, pull-out tests were performed on monolithic implants embedded in bovine ribs and foam blocks that mimic the mechanical parameters of human bone, allowing a quantitative evaluation of implant–bone interface strength and a comparison of geometric variants. Second, the extraction process was recreated in a three dimensional finite element model incorporating nonlinear interface contact and parameterization, enabling the reproduction of load–displacement curves; the results obtained showed good agreement with the experiment. Third, the fracture surfaces were observed macroscopically and by scanning electron microscopy/energy dispersive spectroscopy. The results demonstrated significant distinctions in the forces required to extract implants with varying thread geometries, clearly indicating the impact of implant design on their mechanical stability. The presented FEM-based methodology provides a reliable tool to study mechanical interactions at the implant–bone interface. The findings obtained can improve our understanding of implant behavior in biological systems and provide a basis for further optimization of their design.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Mechanics, Materials and Biomedical Engineering, Faculty of Mechanical Engineering, Wroclaw University of Science and Technology, Smoluchowskiego 25, 50-372 Wroclaw, Poland
Department of Lightweight Elements Engineering, Foundry and Automation, Faculty of Mechanical Engineering, Wroclaw University of Science and Technology, Smoluchowskiego 25, 50-372 Wroclaw, Poland
Independent Researcher, 405300 Gherla, Romania
Independent Researcher, Rzeszowska 2, 68-200 Zary, Poland
References (39)
Show source-provided references
1. Osseointegrated Oral Implants10.1902/jop.1988.59.5.287 ↗
2. Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws
3. Brånemark, P.I., Zarb, G.A., and Albrektsson, T. (1985). Patient selection and preparation. Tissue-Integrated Prostheses: Osseointegration in Clinical Dentistry, Quintessence Publishing Co.
4. Divisions of available bone in implant dentistry
5. Correlation of cone beam CT-derived bone density parameters with primary implant stability assessed by peak insertion torque and periotest in the maxilla10.1016/j.jcms.2019.01.002 ↗
6. Heimes, D., Schiegnitz, E., Kuchen, R., Kämmerer, P.W., and Al-Nawas, B. (2021). Buccal Bone Thickness in Anterior and Posterior Teeth—A Systematic Review. Healthcare, 9.10.3390/healthcare9121663 ↗
7. Wang, S.H., Shen, Y.W., Fuh, L.J., Peng, S.L., Tsai, M.T., Huang, H.L., and Hsu, J.T. (2020). Relationship between Cortical Bone Thickness and Cancellous Bone Density at Dental Implant Sites in the Jawbone. Diagnostics, 10.10.3390/diagnostics10090710 ↗
8. Aljabr, A.A., Almas, K., Aljofi, F.E., Aljabr, A.A., Alzaben, B., and Alqanas, S. (2023). A CBCT Study of Labial Alveolar Bone Thickness in the Maxillary Anterior Region in a Teaching Hospital Population in the Eastern Province of Saudi Arabia. Biomedicines, 11.10.3390/biomedicines11061571 ↗
9. Short implants versus longer implants in the posterior alveolar region after an observation period of at least five years: A systematic review and meta-analysis10.1016/j.jdent.2020.103386 ↗
10. The rise and fall of the alveolar process: Dependency of teeth and metabolic aspects10.1016/j.archoralbio.2018.09.016 ↗
11. Comparative finite element analysis of short implants and lateralization of the inferior alveolar nerve with different prosthesis heights10.1097/SCS.0000000000002083 ↗
12. Investigation of the effect of implant crown ratio and material type on the force transmitted to the implant in implant-supported restorations: A finite element analysis study
13. Sghaireen, M.G., Srivastava, K.C., Shrivastava, D., Ganji, K.K., Patil, S.R., Abuonq, A., Mousa, M.A., Dar-Odeh, N., Sghaireen, G.M., and Kamal, M.A. (2020). A CBCT Based Three-Dimensional Assessment of Mandibular Posterior Region for Evaluating the Possibility of Bypassing the Inferior Alveolar Nerve While Placing Dental Implants. Diagnostics, 10.10.3390/diagnostics10060406 ↗
14. Proximity of the Mandibular Canal to the Tooth Apex10.1016/j.joen.2010.11.030 ↗
15. Oral health-related quality of life in subjects with implant-supported prostheses: A systematic review10.1016/j.jdent.2017.08.003 ↗
16. Do, T.A., Le, H.S., Shen, Y.W., Huang, H.L., and Fuh, L.J. (2020). Risk Factors related to Late Failure of Dental Implant—A Systematic Review of Recent Studies. Int. J. Environ. Res. Public Health, 17.10.3390/ijerph17113931 ↗
17. Mechanical complications of implant-supported complete-arch restorations and impact on patient quality of life: A retrospective cohort study10.1016/j.prosdent.2020.01.006 ↗
18. The influence of bone quality on the biomechanical behavior of full-arch implant-supported fixed prostheses10.1016/j.msec.2014.01.013 ↗
19. (2021). Standard Specification for Rigid Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments (Standard No. ASTM F1839-08).
20. Bone density and its importance in orthodontics10.1016/j.jobcr.2013.01.001 ↗
21. (2022). Open Die Steel Forgings for General Engineering Purposes—Part 2: Non-Alloy Quality and Special Steels (Standard No. EN 10250-2:2022).
22. Słowiński, J.J., Roszak, M., Kazimierczak, M., Skrzypczak, G., and Stępczak, M. (2025). Enhanced Mechanical Performance of Resin-Infused 3D-Printed Polymer Lattices. Polymers, 17.10.3390/polym17081028 ↗
23. Romanowska, K., Pyka, D., Opieliński, K., Krawiec, K., Śliwiński, R., and Jamroziak, K. (2023). Preliminary Numerical Analysis of Mechanical Wave Propagation Due to Elastograph Measuring Head Application in Non-Invasive Liver Condition Assessment. Appl. Sci., 13.10.3390/app132111843 ↗
24. Artificial composite bone as a model of human trabecular bone: The implant–bone interface10.1016/j.jbiomech.2006.04.007 ↗
25. Słowiński, J., Roszak, M., Krawiec, K., Henklewski, R., and Jamroziak, K. (2024). Numerical Analysis of Stabilization of a Horse’s Third Metacarpal Bone Fracture for Prediction of the Possibility of Bone Union. Appl. Sci., 14.10.3390/app14177976 ↗
26. Ceddia, M., Montesani, L., Comuzzi, L., Cipollina, A., Deporter, D.A., Pietro, N.D., and Trentadue, B. (2025). The Influence of Insertion Torque on Stress Distribution in Peri-Implant Bones Around Ultra-Short Implants: An FEA Study. J. Funct. Biomater., 16.10.3390/jfb16070260 ↗
27. Versatility of Basal Cortical Screw Implants with Immediate Functional Loading10.1007/s12663-021-01638-6 ↗
28. (2021). Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (Standard No. ASTM F136-13(2021)e1).
29. Recommendations for standardised screw pull-out from polyurethane foam-The influence of density variations of the test foam and the insertion method10.1016/j.medengphy.2022.103750 ↗
30. Trends in Dental Implant Use in the U.S., 1999–2016, and Projections to 202610.1177/0022034518792567 ↗
31. Public perceptions of dental implants: A qualitative study10.1016/j.jdent.2015.04.012 ↗
32. Teaching and assessment of implant dentistry in undergraduate and postgraduate education: A European consensus10.1111/j.1600-0579.2008.00556.x ↗
33. The development and future of dental implants10.4012/dmj.2019-140 ↗
34. Improvements in Implant Dentistry over the Last Decade: Comparison of Survival and Complication Rates in Older and Newer Publications10.11607/jomi.2014suppl.g5.2 ↗
35. Clinical study on screw loosening in dental implant prostheses: A 6-year retrospective study10.5125/jkaoms.2020.46.2.133 ↗
36. Prosthetic failures in dental implant therapy10.1111/prd.12416 ↗
37. Prevention and treatment of peri-implant diseases—The EFP S3 level clinical practice guideline10.1111/jcpe.13823 ↗
38. Identification of failure mechanisms in retrieved fractured dental implants10.1016/j.engfailanal.2014.01.002 ↗
39. Morphological study of Osteoplate 2000-extension implants after bending10.1097/00008505-200009030-00012 ↗