Healing Abutment Distortion in Implant Prostheses: An In Vitro Study
Source abstract
Abstract
Little has been studied regarding the repeated usage of healing abutments and their effects on the distortion of dental implant-healing abutment hex surfaces. Additionally, implant manufacturers do not provide specific guidelines on how many times a healing abutment can be used before discarding. Therefore, we evaluated the effect of repeated screwing-unscrewing of implant-healing abutment on the healing abutment hex surface and screw head. A total of 12 Biomate implants with 4 mm diameter and 13 mm length were inserted into a synthetic bone block. The standard healing abutments of 3 mm diameter and 4 mm length were screwed onto each implant using a torque ratchet at a final torque of 30 Ncm. Immediately, the abutments were unscrewed at 30 Ncm. Then, screwing-unscrewing was repeated for 4, 8, 16, 24, 32, 40, 80, 160, 320, and 400 times and the healing abutments were scanned under the scanning electron microscope for any distortion. Distortion was graded as 0, 1, 2, and 3. Data were analyzed using SPSS 24.0. Descriptive statistics were calculated. One-way ANOVA with post hoc using Tukey’s HSD test was performed to analyze the difference in distortion at different screwing-unscrewing times. A significant level was selected at p-value = 0.05. It was found that distortion healing abutments screws were seen after 32 times screwing and unscrewing. There was a significant difference in the distortion (p-value < 0.05) after 24 times of repeated usage of healing abutment and at 160, 320, and 400 times. No surface distortions were observed at the healing abutment screw head at 4, 8, 16, 24, 32, 40, 80, 160, 320, and 400 cycles of screwing-unscrewing. It can be concluded that repeated screwing and unscrewing of the implant-healing abutments causes damage to the healing abutment hex surface. The distortion of healing abutments screws was seen after 32 times screwing and unscrewing. No surface distortions were observed on the healing abutment screw head until 400 times of screwing and unscrewing. Hence, the clinician should be cautious while using the healing abutments repeatedly.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Dental Implant Techniques and Outcomes · Dental materials and restorations
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Clinical Dentistry, International College of Dentistry, Walailak University, Bangkok 10400, Thailand
Department of Oral Biosciences, International College of Dentistry, Walailak University, Bangkok 10400, Thailand
Funding
Walailak University International College of Dentistry Research Fund
References (40)
Show source-provided references
1. Management of obstructive sleep apnea with implant retained mandibular advancement device10.5005/jp-journals-10015-1285 ↗
2. Consensus statements and recommended clinical procedures regarding loading protocols for endosseous dental implants
3. Suphangul, S., Rokaya, D., Kanchanasobhana, C., Rungsiyakull, P., and Chaijareenont, P. (2022). Peek biomaterial in long-term provisional implant restorations: A review. J. Funct. Biomater., 13.10.3390/jfb13020033 ↗
4. Clinical and esthetic outcomes of implants placed in postextraction sites
5. Is it safe to reuse dental implant healing abutments sterilized and serviced by dealers of dental implant manufacturers? An in vitro sterility analysis
6. In-vitro study of the contamination remaining on used healing abutments after cleaning and sterilizing in dental practice10.1111/cid.12385 ↗
7. Reuse of implant healing abutments: Comparative evaluation of the efficacy of two cleaning procedures10.11607/ijp.5552 ↗
8. Fatigue characteristics of five types of implant-abutment joint designs10.3365/met.mat.2008.04.133 ↗
9. The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling10.11607/jomi.2727 ↗
10. The effect of implant/abutment hexagonal misfit on screw joint stability
11. Physicochemical and microscopic characterization of implant-abutment joints10.4103/ejd.ejd_3_17 ↗
12. The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws10.1067/mpr.2002.127604 ↗
13. Reusing titanium healing abutments: Comparison of two decontamination methods10.11607/ijp.5881 ↗
14. Investigation of different electrochemical cleaning methods on contaminated healing abutments in vitro: An approach for metal surface decontamination10.1186/s40729-020-00265-z ↗
15. Peri-implantitis update: Risk indicators, diagnosis, and treatment10.1055/s-0040-1715779 ↗
16. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws
17. Torque removal evaluation of prosthetic screws after tightening and loosening cycles: An in vitro study10.1111/j.1600-0501.2011.02165.x ↗
18. Titanium and its alloys in dental implantology
19. Titanium alloys for dental implants: A review10.3390/prosthesis2020011 ↗
20. Binary titanium alloys as dental implant materials—A review10.1093/rb/rbx027 ↗
21. In vitro osteogenic properties of two dental implant surfaces10.1155/2012/181024 ↗
22. Dental pulp stem cells on implant surface: An in vitro study10.1155/2021/3582342 ↗
23. Effect of repeated screw joint closing and opening cycles and cyclic loading on abutment screw removal torque and screw thread morphology: Scanning electron microscopy evaluation10.11607/jomi.5476 ↗
24. Mlikota, M., and Schmauder, S. (2018). On the critical resolved shear stress and its importance in the fatigue performance of steels and other metals with different crystallographic structures. Metals, 8.10.3390/met8110883 ↗
25. Effect of repeated closures on opening torque values in seven abutment-implant systems10.1067/mpr.2000.108069 ↗
26. Effect of repeated torque/mechanical loading cycles on two different abutment types in implants with internal tapered connections: An in vitro study10.1111/j.1600-0501.2008.01690.x ↗
27. Kim, K.S., and Lim, Y.J. (2020). Axial displacements and removal torque changes of five different implant-abutment connections under static vertical loading. Materials, 13.10.3390/ma13030699 ↗
28. Preloads generated with repeated tightening in three types of screws used in dental implant assemblies10.1111/j.1532-849X.2006.00096.x ↗
29. A methodology and new criteria to quantify the adhesive and abrasive wear damage on a die radius using white light profilometry10.1016/j.triboint.2012.02.018 ↗
30. Study of 2d and 3d methods for worn surface analysis of tool materials10.14513/actatechjaur.v8.n2.373 ↗
31. Comparative evaluation of surface changes in four ni-ti instruments with successive uses—An sem study10.4103/0972-0707.85817 ↗
32. A scanning electron microscopic study of wear surfaces10.1007/BF03355668 ↗
33. A comparison of the precision of three-dimensional images acquired by 2 digital intraoral scanners: Effects of tooth irregularity and scanning direction10.4041/kjod.2016.46.1.3 ↗
34. Characterization of 3d geometric distortion of magnetic resonance imaging scanners commissioned for radiation therapy planning10.1016/j.mri.2016.01.001 ↗
35. Characterization of surface topography of wear particles by sem stereoscopy10.1016/S0043-1648(96)07324-3 ↗
36. Stress distribution in implant retained finger prosthesis: A finite element study
37. Preload evaluation of different screws in external hexagon joint10.1097/ID.0b013e31823fcbce ↗
38. Preload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review
39. Loss of preload in screwed implant joints as a function of time and tightening/untightening sequences10.11607/jomi.3344 ↗
40. Endoscopic removal of an aspirated healing abutment and screwdriver under conscious sedation10.1097/ID.0000000000000100 ↗