A Comparison of Wear Patterns on Retrieved and Simulator-Tested Total Knee Replacements

Rebecca H. Dammer, Carmen Zietz, Rainer Bader

Abstract

Aseptic implant loosening is the most common reason for revision surgery after total knee replacement. This is associated with adverse biological reactions to wear debris from the articulating implant components. To predict the amount of wear debris generated in situ, standard wear testing of total knee replacement (TKR) is carried out before its clinical use. However, wear data reported on retrievals of total knee replacement (TKR) revealed significant discrepancies compared with standard wear simulator studies. Therefore, the aim of the present study was to compare the wear patterns on identical posterior-cruciate-retaining TKR designs by analyzing retrieved and experimentally tested implants. The identification and classification of wear patterns were performed using 21 retrieved ultra-high-molecular-weight-polyethylene (UHMW-PE) inserts and four sets of inserts of identical design and material tested in a knee wear simulator. These four sets had undergone different worst-case conditions and a standard test in a wear simulator according to ISO 14243-1. Macroscopic and microscopic examinations of the polyethylene inserts were performed, including the determination of seven modes of wear that correspond to specific wear patterns, the calculation of wear areas, and the classification of the damage over the whole articulating area. Retrieved and standard wear simulator-tested UHMW-PE inserts showed significant differences in wear area and patterns. The total wear areas and the damage score were significantly larger on the retrievals (52.3% versus 23.9%, 32.7 versus 22.7). Furthermore, the range of wear patterns found on the retrievals was not reproducible in the simulator-tested inserts. However, good correspondence was found with the simulator-tested polyethylene inserts under worst-case conditions (third body wear), i.e., deep wear areas could be replicated according to the in vivo situation compared with other wear test scenarios. Based on the findings presented here, standard simulator testing can be used to directly compare different TKR designs but is limited in the prediction of their in situ wear. Preclinical wear testing may be adjusted by worst-case conditions to improve the prediction of in situ performance of total knee implants in the future.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Biomechanics and Implant Technology Research Laboratory, Department of Orthopaedics, Rostock University Medical Center, Doberaner Strasse 142, 18057 Rostock, Germany

References (47)

Show source-provided references
1. The German Arthroplasty Registry (EPRD) (2021). Annual Report 2021, EPRD, German Arthroplasty Registry gGmbH.
2. Swedish Knee Arthroplasty Register (2020). Annual Report 2020, Swedish Knee Arthroplasty Register: Lund University, Department of Clinical Sciences, Orthopedics, Skåne University Hospital.
3. The Survival of Total Knee Arthroplasty: Current Data from Registries on Tribology10.1007/s11420-016-9513-9 ↗
4. Trieb, K., Heller, K.-D., and Wirtz, D.C. (2011). Septic Revision Knee Arthroplasty–U.S. Experience. Revisionsendoprothetik des Kniegelenks, Springer.10.1007/978-3-642-16123-0 ↗
5. Particle disease: Biologic mechanisms of periprosthetic osteolysis in total hip arthroplasty10.1177/1753425912451779 ↗
6. Swedish Arthroplasty Register (2021). Annual Report 2021, Swedish Arthroplasty Register.
7. Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prostheses10.1002/jbm.820170510 ↗
8. Biological reaction to debris in relation to joint prostheses10.1243/0954411971534304 ↗
9. A Striated Pattern of Wear in Ultrahigh-molecular-weight Polyethylene Components of Miller-Galante Total Knee Arthroplasty Material10.1016/S0883-5403(98)90069-9 ↗
10. Klassifikation bei Prothesen—Insuffizienz und Partikelbestimmung10.1007/s00292-008-1070-7 ↗
11. A novel method for the prediction of functional biological activity of polyethylene wear debris10.1243/0954411011533599 ↗
12. (2012). Implants for Surgery—Wear of Total Hip-Joint Prostheses—Part 1: Loading and Displacement Parameters for Wear-Testing Machines and Corresponding Environmental Conditions for Test (Standard No. ISO 14242-1:2012).
13. (2009). Implants for Surgery—Wear of Total Hip-Joint Prostheses—Part 3: Loading and Displacement Parameters for Orbital Bearing Type Wear Testing Machines and Corresponding Environmental Conditions for Test (Standard No. ISO 14242-3:2009).
14. (2009). Implants for Surgery—Wear of Total Knee-Joint Prostheses—Part 1: Loading and Displacement Parameters for Wear-Testing Machines with Load Control and Corresponding Environmental Conditions for Test (Standard No. ISO 14243-1:2009).
15. (2004). Implants for Surgery—Wear of Total Knee-Joint Prostheses—Part 3: Loading and Displacement Parameters for Wear-Testing Machines with Displacement Control and Corresponding Environmental Conditions for Test (Standard No. ISO 14243-3:2004).
16. Australian Orthopaedic Association, and National Joint Replacement Registry (AOANJRR) (2021). Hip, Knee & Shoulder Arthroplasty: 2021 Annual Report, Australian Orthopaedic Association.
17. Wear testing of total hip replacements under severe conditions10.1586/17434440.2015.1050378 ↗
18. The Characteristics of acetabular cups worn in the human body10.1302/0301-620X.60B3.681413 ↗
19. Polyethylene insert damage in unicondylar knee replacement: A comparison of in vivo function and in vitro simulation10.1243/09544119JEIM699 ↗
20. Retrieval, Experimental, and Computational Assessment of the Performance of Total Knee Replacements10.1002/jor.20181 ↗
21. Comparison of Polyethylene Tibial Insert Damage from In Vivo Function and In Vitro Wear Simulation10.1002/jor.20743 ↗
22. Experimental testing of total knee replacements with UHMW-PE inserts: Impact of severe wear test conditions10.1007/s10856-015-5470-y ↗
23. Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach
24. Fabry, C., Zietz, C., Dammer, R., and Bader, R. (2015). Patterns of Wear in Total Knee Replacement. The Unhappy Total Knee Replacement, Springer International Publishing.10.1007/978-3-319-08099-4_13 ↗
25. Wear Patterns on Retrieved Polyethylene Tibial Inserts and Their Relationship to Technical Considerations During Total Knee Arthroplasty10.1097/00003086-199402000-00005 ↗
26. Micro-wear patterns on UHMWPE tibial inserts in total knee joint simulation10.1002/jbm.10027 ↗
27. Manescu, V., Antoniac, I., Antoniac, A., Paltanea, G., Miculescu, M., Bita, A.-I., Laptoiu, S., Niculescu, M., Stere, A., and Paun, C. (2022). Failure Analysis of Ultra-High Molecular Weight Polyethylene Tibial Insert in Total Knee Arthroplasty. Materials, 15.10.3390/ma15207102 ↗
28. Deformation, yielding, fracture and fatigue behavior of conventional and highly cross-linked ultra high molecular weight polyethylene10.1016/j.biomaterials.2004.03.022 ↗
29. Third-body abrasive wear of tibial polyethylene inserts combined with metallic and ceramic femoral components in a knee simulator study10.5301/ijao.5000189 ↗
30. Zietz, C., Dempwolf, H., Dammer, R., Mick, E., and Bader, R. (2022, November 18). Wear Simulator Study of Total Knee Replacement with Increased Kinematic Test Conditions. ESB-Congress, Prague, Poland 2015. Available online: https://www.czech-in.org/cm/ESB/CM.NET.WebUI/CM.NET.WEBUI.scpr/SCPRfunctiondetail.aspx?confID=05000000-0000-0000-0000-000000000056&sesID=05000000-0000-0000-0000-000000003560&absID=07000000-0000-0000-0000-000000023249.
31. The 2012 Frank Stinchfield Award10.1007/s11999-012-2497-y ↗
32. Wright, T.M., and Goodman, S. (2000, January 21–23). Material and design considerations—What are the wear mechanisms and what controls them?. Proceedings of the Implant Wear in Total Joint Replacement: Symposium, Oakbrook, IL, USA.
33. Wear of Polyethylene in Total Joint Replacements—Observations from Retrieved PCA Knee Implants10.1097/00003086-199203000-00016 ↗
34. A Quantitative Technique for Reporting Surface Degradation Patterns of UHMWPE Components of Retrieved Total Knee Replacements10.1002/jab.770060103 ↗
35. Midterm Assessment of Causes and Results of Revision Total Knee Arthroplasty10.1007/s11999-009-1204-0 ↗
36. Revision Total Knee Arthroplasty with Modular Cemented Stems10.1016/j.arth.2007.03.025 ↗
37. Long-term Survival Rate of Implants and Modes of Failure After Revision Total Knee Arthroplasty by a Single Surgeon10.1016/j.arth.2012.08.021 ↗
38. Duration and frequency of every day activities in total hip patients10.1016/S0021-9290(01)00035-5 ↗
39. Keller, R., and Amis, A. (2015). Anatomy and Biomechanics of the Natural Knee and After TKR. The Unhappy Total Knee Replacement, Springer International Publishing.10.1007/978-3-319-08099-4_1 ↗
40. Joint Line Elevation and Tibial Slope are associated with increased Polyethylene Wear in Cruciate Retaining Total Knee Replacement10.1002/jor.24710 ↗
41. The relationship between alignment, function and loading in total knee replacement: In-vivo analysis of a unique patient population10.1016/j.jbiomech.2020.110042 ↗
42. Relative Contributions of Design, Alignment, and Loading Variability in Knee Replacement Mechanics10.1002/jor.22169 ↗
43. Backside Wear Analysis of Retrieved Acetabular Liners with a Press-Fit Locking Mechanism in Comparison to Wear Simulation In Vitro
44. In Vivo Oxidation Contributes to Delamination but not Pitting in Polyethylene Components for Total Knee Arthroplasty10.1016/j.arth.2010.07.010 ↗
45. “Thicker” Polyethylene Bearings Are Associated with Higher Failure Rates in Primary Total Knee Arthroplasty10.1016/j.arth.2010.04.031 ↗
46. Wear of ultra-high molecular weight polyethylene (UHMWPE) in total knee prostheses: A review of key influences10.1243/0954411001535390 ↗
47. Clinical Performance of Contemporary Tibial Polyethylene Components10.1016/j.arth.2005.10.012 ↗