journal article·2022-01-12Open access

Hemolytic Performance in Two Generations of the Sputnik Left Ventricular Assist Device: A Combined Numerical and Experimental Study

Alexandra N. Romanova, Alexander A. Pugovkin, Maxim V. Denisov, Ivan A. Ephimov, Dmitry V. Gusev, Marian Walter, Thomas Groth, Olga L. Bockeria, Tatyana G. Le, Anna S. Satyukova, Sergey V. Selishchev, Dmitry V. Telyshev

Abstract

Background: Currently, left ventricular assist devices (LVADs) are a successful surgical treatment for patients with end-stage heart failure on the waiting list or with contraindicated heart transplantation. In Russia, Sputnik 1 LVAD was also successfully introduced into clinical practice as a bridge-to-transplant and a destination therapy device. Development of Sputnik 2 LVAD was aimed at miniaturization to reduce invasiveness, optimize hemocompatibility, and improve versatility for patients of various sizes. Methods: We compared hemolysis level in flow path of the Sputnik LVADs and investigated design aspects influencing other types of blood damage, using predictions of computational fluid dynamics (CFD) and experimental assessment. The investigated operating point was a flow rate of 5 L/min and a pressure head of 100 mm Hg at an impeller rotational speed of 9100 min−1. Results: Mean hemolysis indices predicted with CFD were 0.0090% in the Sputnik 1 and 0.0023% in the Sputnik 2. Averaged values of normalized index of hemolysis obtained experimentally for the Sputnik 1 and the Sputnik 2 were 0.011 ± 0.003 g/100 L and 0.004 ± 0.002 g/100 L, respectively. Conclusions: Obtained results indicate obvious improvements in hemocompatibility and sufficiently satisfy the determined miniaturization aim for the Sputnik 2 LVAD development.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Mechanical Circulatory Support Devices · Cardiac and Coronary Surgery Techniques · Transplantation: Methods and Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Institute of Biomedical Systems, National Research University of Electronic Technology, 124498 Zelenograd, Moscow, Russia

Center for Digital Biodesign and Personalized Healthcare, I. M. Sechenov First Moscow State Medical University, 119991 Moscow, Russia

Medical Information Technology, Helmholtz Institute of Biomedical Engineering, RWTH Aachen University, 52074 Aachen, Germany

Department Biomedical Materials, Institute of Pharmacy, Martin Luther University Halle-Wittenberg, 06120 Halle (Saale), Germany

Bakulev Scientific Center for Cardiovascular Surgery, 121552 Moscow, Russia

Funding

Russian Science Foundation · Deutsche Forschungsgemeinschaft

References (66)

Show source-provided references
1. The International Thoracic Organ Transplant Registry of the International Society for Heart and Lung Transplantation: 37th Adult Heart Transplantation Report—2020; Focus on Deceased Donor Characteristics10.1016/j.healun.2020.07.010 ↗
2. The Society of Thoracic Surgeons Intermacs 2020 Annual Report10.1016/j.athoracsur.2020.12.038 ↗
3. Fully magnetically levitated left ventricular assist system for treating advanced HF: A multicenter study10.1016/j.jacc.2015.09.083 ↗
4. A systematic review of the cost-effectiveness of long-term mechanical circulatory support10.1016/j.jval.2014.12.020 ↗
5. A systematic review of cost-effectiveness analyses of left ventricular assist devices: Issues and challenges10.1007/s40258-018-0439-x ↗
6. Model for a general mechanical blood damage prediction10.1111/j.1525-1594.1995.tb02385.x ↗
7. An estimation method of hemolysis within an axial flow blood pump by computational fluid dynamics analysis10.1046/j.1525-1594.2003.00034.x ↗
8. Computational fluid dynamics prediction of blood damage in a centrifugal pump10.1046/j.1525-1594.2003.00026.x ↗
9. Interactive blood damage analysis for ventricular assist devices10.1109/TVCG.2008.142 ↗
10. Evaluation of the impeller shroud performance of an axial flow ventricular assist device using computational fluid dynamics10.1111/j.1525-1594.2010.01099.x ↗
11. The use of computational fluid dynamics in the development of ventricular assist devices10.1016/j.medengphy.2010.10.014 ↗
12. A quantitative comparison of mechanical blood damage parameters in rotary ventricular assist devices: Shear stress, exposure time and hemolysis index10.1115/1.4007092 ↗
13. Thrombus formation patterns in the HeartMate II ventricular assist device: Clinical observations can be predicted by numerical simulations10.1097/MAT.0000000000000034 ↗
14. Numerical analysis of blood damage potential of the HeartMate II and HeartWare HVAD rotary blood pumps10.1111/aor.12542 ↗
15. A two-stage rotary blood pump design with potentially lower blood trauma: A computational study10.5301/ijao.5000482 ↗
16. Karimov, J.H., Fukamachi, K., and Starling, R.C. (2020). Computational fluid dynamics for mechanical circulatory support device development. Mechanical Support for Heart Failure, Springer.10.1007/978-3-030-47809-4 ↗
17. Fast three-dimensional numerical hemolysis approximation10.1111/j.1525-1594.2004.00026.x ↗
18. Long-term changes in kidney function after left ventricular assist device implant: An analysis of the STS Intermacs database10.1016/j.healun.2019.01.206 ↗
19. Hemolysis and free hemoglobin revisited: Exploring hemoglobin and hemin scavengers as a novel class of therapeutic proteins10.1182/blood-2012-11-451229 ↗
20. A cellular model of shear-induced hemolysis10.1111/aor.12832 ↗
21. Turbulence in blood damage modeling10.5301/ijao.5000476 ↗
22. A tensor-based measure for estimating blood damage10.1111/j.1525-1594.2004.00072.x ↗
23. Hemolysis estimation in a centrifugal blood pump using a tensor-based measure10.1111/j.1525-1594.2006.00256.x ↗
24. Strain-based blood damage estimation for computational design of ventricular assist devices10.5301/ijao.5000484 ↗
25. A fluid dynamic analysis of a rotary blood pump for design improvement10.1111/j.1525-1594.1993.tb00634.x ↗
26. A mathematical model for shear-induced hemolysis10.1111/j.1525-1594.1995.tb02384.x ↗
27. Hydrodynamic and hemolysis analysis on distance and clearance between impeller and diffuser of axial blood pump10.1142/S0219519416500147 ↗
28. Blood pump design variations and their influence on hydraulic performance and indicators of hemocompatibility10.1007/s10439-017-1951-0 ↗
29. Numerical investigation of the influence of blade radial gap flow on axial blood pump performance10.1097/MAT.0000000000000745 ↗
30. Numerical and experimental analysis of an axial flow left ventricular assist device: The influence of the diffuser on overall pump performance10.1111/j.1525-1594.2005.29095.x ↗
31. Numerical studies of an axial flow blood pump with different diffuser designs10.1142/S0219519413500292 ↗
32. The effect of rotor geometry on the H−Q curves of the sputnik implantable pediatric rotary blood pump10.1007/s10527-017-9669-8 ↗
33. Numerical investigation of the effects of the clearance gap between the inducer and impeller of an axial blood pump10.1111/j.1525-1594.2005.29046.x ↗
34. Optimization of axial pump characteristic dimensions and induced hemolysis for mechanical circulatory support devices10.1097/MAT.0000000000000719 ↗
35. Optimization of centrifugal pump characteristic dimensions for mechanical circulatory support devices10.1097/MAT.0000000000000393 ↗
36. Hemolysis analysis of axial blood pump with various structure impellers10.1142/S0219519413500541 ↗
37. Effect of blade curvature on the hemolytic and hydraulic characteristics of a centrifugal blood pump10.1177/0391398818785558 ↗
38. Parametric study of blade tip clearance, flow rate, and impeller speed on blood damage in rotary blood pump10.1111/j.1525-1594.2009.00754.x ↗
39. Proposal of hemodynamically improved design of an axial flow blood pump for LVAD10.1007/s11517-019-02097-5 ↗
40. Optimisation of the Sputnik-VAD design10.5301/ijao.5000518 ↗
41. Artificial heart in Russia: Past, present and future10.1111/aor.13860 ↗
42. Advances in hemodynamic analysis in cardiovascular diseases Investigation of energetic characteristics of adult and pediatric Sputnik left ventricular assist devices during mock circulation support10.1155/2019/4593174 ↗
43. Estimation of shear stress-related blood trauma in heart valve prostheses—In vitro comparison of 25 aortic valves10.1177/039139889001300507 ↗
44. A Couette viscometer for short time shearing of blood10.3233/BIR-1980-171-205 ↗
45. Evaluation of Eulerian and Lagrangian models for hemolysis estimation10.1097/MAT.0b013e318254833b ↗
46. Computational fluid dynamics analysis of an intra-cardiac axial flow pump10.1046/j.1525-1594.2003.07190.x ↗
47. Three-dimensional numerical prediction of stress loading of blood particles in a centrifugal pump10.1111/j.1525-1594.1995.tb02386.x ↗
48. ASTM International F-1841-97(2013) (2013). Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps. Annual Book of ASTM Standards, vol. 13.01. Medical and Surgical Materials and Devices (I): E667–F2477, ASTM International.
49. ASTM International F-1830-97(2013) (2013). Standard Practice for Selection of Blood for in vitro Evaluation of Blood Pumps. Annual Book of ASTM Standards, vol. 13.01. Medical and Surgical Materials and Devices (I): E667–F2477, ASTM International.
50. Plasma hemoglobin measurement techniques for the in vitro evaluation of blood damage caused by medical devices10.1111/j.1525-1594.1997.tb00486.x ↗
51. Design and development strategy for the rotary blood pump10.1046/j.1525-1594.1998.06098.x ↗
52. Evaluation of in vitro hemolysis and platelet activation of a newly developed maglev LVAD and two clinically used LVADs with human blood10.1111/aor.13471 ↗
53. Design rationale and preclinical evaluation of the HeartMate 3 left ventricular assist system for hemocompatibility10.1097/MAT.0000000000000388 ↗
54. Kuchinka, J., Willems, C., Telyshev, D.V., and Groth, T. (2021). Control of blood coagulation by hemocompatible material surfaces—A review. Bioengineering, 8.10.3390/bioengineering8120215 ↗
55. Bird, R.B., Armstrong, R.C., and Hassager, O. (1987). Dynamics of Polymeric Liquids, Volume 1: Fluid Mechanics, Wiley.
56. Kim, S. (2002). A Study of Non-Newtonian Viscosity and Yield Stress of Blood in a Scanning Capillary-Tube Rheometer. [Ph.D. Thesis, Drexel University].
57. Determination of rheological properties of whole blood with a scanning capillary-tube rheometer using constitutive models10.1007/s12206-009-0420-6 ↗
58. MRI measurement of time-resolved wall shear stress vectors in a carotid bifurcation model, and comparison with CFD predictions10.1002/jmri.10243 ↗
59. Reconstruction of blood flow patterns in a humancarotid bifurcation: A combined CFD and MRI study10.1002/(SICI)1522-2586(200003)11:3<299::AID-JMRI9>3.0.CO;2-M ↗
60. Noise reduction in three-dimensional phase contrast MR velocity measurements10.1002/jmri.1880030407 ↗
61. Numerical experiment for ultrasonic-measurement-integrated simulation of developed laminar pipe flow using axisymmetric model10.1299/jbse.3.101 ↗
62. Reconstruction of carotid bifurcation hemodynamics and wall thickness using computational fluid dynamics and MRI10.1002/mrm.10025 ↗
63. Feasibility of in vivo measurement of carotid wall shear rate using spiral Fourier velocity encoded MRI10.1002/mrm.22325 ↗
64. Effects of non-Newtonian fluid on centrifugal blood pump performance10.1016/j.icheatmasstransfer.2007.11.005 ↗
65. The rheology of blood flow in a branched arterial system10.1515/arh-2005-0020 ↗
66. A mathematical model to describe the change in the constitutive character of blood due to platelet activation10.1016/S1631-0721(02)01501-2 ↗