journal article·2021-11-24Open access

Wear Resistance of Ti–6Al–4V Alloy Ball Heads for Use in Implants

Svetlana Skvortsova, Aleksei Orlov, Georgii Valyano, Victor Spektor, Natalia Mamontova

Abstract

The effect of thermohydrogen treatment and vacuum ion–plasma nitriding on the determination of the volume and surface structure of ball heads made of Ti–6Al–4V alloy was studied. It was found that the submicrocrystalline structure formed in the head during thermohydrogen treatment makes it possible to achieve hardness values of 39–41 units HRC and a surface roughness of 0.02 μm. It was shown that the creation of a modified layer consisting of ε (TiN) and δ (Ti2N) titanium nitrides on the surface of a ball head and the solid interstitial solution of nitrogen in α-titanium makes it possible to completely eliminate material wear when testing for friction on ultra-high-molecular-weight polyethylene. The equivalent analysis was also conducted with a ball head that had been implanted in a human body for 12 years. It was found that the change in the color of the head, from slightly golden after nitriding to metallic, is due to the formation of an oxynitride nanoscale layer on the surface. It was shown that in contrast with films made of titanium oxide, the film developed in this study has high wear resistance.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Bone Tissue Engineering Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Moscow Aviation Institute, National Research University, 4, Volokolamskoe Highway, 125993 Moscow, Russia

Joint Institute for High Temperatures of Russian Academy of Sciences (JIHT), 13 Bd.2, Izhorskaya Str., 125412 Moscow, Russia

References (33)

Show source-provided references
1. Metallic Biomaterials of Knee and Hip—A Review
2. Biomaterials in Orthopaedics10.1098/rsif.2008.0151 ↗
3. Narayan, R. (2009). Biomedical Materials, Springer.10.1007/978-0-387-84872-3 ↗
4. Ti Based Biomaterials, the Ultimate Choice for Orthopaedic Implants—A Review10.1016/j.pmatsci.2008.06.004 ↗
5. Hwang, Y.-J., Choi, Y.-S., Hwang, Y.-H., Cho, H.-W., and Lee, D.-G. (2020). Biocompatibility and Biological Corrosion Resistance of Ti-39Nb-6Zr+0.45Al Implant Alloy. J. Funct. Biomater., 12.10.3390/jfb12010002 ↗
6. The Mechanism-Based Approach of Understanding Run-in and Steady State: A Gross-Slip Fretting Experiment to Fathom Tribocorrosion of Total Hip Taper Junctions10.1016/j.biotri.2021.100165 ↗
7. Abdel, M.P., and Della Valle, C.J. (2017). Corrosion at the Modular Head–Neck Junction. Complications after Primary Total Hip Arthroplasty: A Comprehensive Clinical Guide, Springer International Publishing.
8. Fretting Corrosion of CoCrMo and Ti6Al4V Interfaces10.1016/j.biomaterials.2012.04.015 ↗
9. Fretting-Corrosion Behavior in Hip Implant Modular Junctions: The Influence of Friction Energy and PH Variation10.1016/j.jmbbm.2016.05.024 ↗
10. Jamari, J., Ammarullah, M.I., Saad, A.P.M., Syahrom, A., Uddin, M., van der Heide, E., and Basri, H. (2021). The Effect of Bottom Profile Dimples on the Femoral Head on Wear in Metal-on-Metal Total Hip Arthroplasty. J. Funct. Biomater., 12.10.3390/jfb12020038 ↗
11. Design, Microstructure and Mechanical Characterization of Ti6Al4V Reinforcing Elements for Cement Composites with Fractal Architecture10.1016/j.matdes.2019.107758 ↗
12. A Study of the Effect of Conventional Drilling and Helical Milling in Surface Quality in Titanium Ti-6Al-4V and Ti-6AL-7Nb Alloys for Medical Applications10.1007/s13369-020-05047-8 ↗
13. Mechanical, Chemical and Biological Damage Modes within Head-Neck Tapers of CoCrMo and Ti6Al4V Contemporary Hip Replacements10.1002/jbm.b.33972 ↗
14. Surface and Subsurface Changes as a Result of Tribocorrosion at the Stem-Neck Interface of Bi-Modular Prosthesis10.1016/j.biotri.2017.02.002 ↗
15. The Evolution of Subsurface Micro-Structure and Tribo-Chemical Processes in Cocrmo-Ti6al4v Fretting-Corrosion Contacts: What Lies at and below the Surface?10.1016/j.wear.2019.203095 ↗
16. Fretting of CoCrMo and Ti6Al4V Alloys in Modular Prostheses10.1179/1751584X15Y.0000000014 ↗
17. Why Does Titanium Alloy Wear Cobalt Chrome Alloy despite Lower Bulk Hardness: A Nanoindentation Study?10.1016/j.tsf.2013.06.020 ↗
18. Formation of Functional Coatings Based on Interstitial Compounds on Titanium under the Conditions of Thermodiffusion Saturation10.1007/s11003-006-0083-8 ↗
19. Arzamasov, B.N., Bratukhin, A.G., Eliseev, Y.S., and Panayoti, T.A. (1999). Ionic Chemical Heat Treatment of Alloys, Publishing House of the Bauman Moscow State Technical University.
20. The Effect of Nitriding on the Mechanical Properties and Wear-Resistance of Titanium Alloys
21. Structural Aspects of Ionic Nitriding of Titanium Alloys
22. Ion Implantation of Surgical Ti-6Al-4V for Improved Resistance to Wear-Accelerated Corrosion10.1002/jbm.820210308 ↗
23. Low-Temperature Ion Nitriding of Implants Made of VT20 Ti Alloy in Different Structural States
24. Quantitative Assessment of Ensuring the Performance of Titanium Alloys by Vacuum Ion-Plasma Treatment
25. Paulitsch-Fuchs, A.H., Wolrab, L., Eck, N., Dyer, N.P., Bödendorfer, B., and Lohberger, B. (2021). TiAl6V4 Alloy Surface Modifications and Their Impact on Biofilm Development of S. Aureus and S. Epidermidis. J. Funct. Biomater., 12.10.3390/jfb12020036 ↗
26. Principles of Construction of Complex Technological Processes for the Production of Implants from Titanium Alloys, Includ-Ing Vacuum Ion-Plasma Nanotechnology
27. Substantiation of Technological Parameters of Mechanical Processing of Ball Heads of Endoprostheses Made of Titanium Al-Loy Ti-6Al-4V
28. Ilyin, A.A., Kolachev, B.A., Nosov, V.K., and Mamonov, A.M. (2002). Hydrogen Technology of Titanium Alloys, MISiS.
29. Controlling the Phase Composition, Structure, and Complex of Properties of the High-Modulus Titanium Alloy by Thermohydrogen Processing10.3103/S1067821218020062 ↗
30. Ilyin, A.A., Balberkin, A.V., Mamonov, A.M., Karpov, V.N., Nadezhin, A.M., Ovchinikov, A.V., Shavyrin, A.B., and Kolondaev, A.F. (2008, January 18–21). Materials Science and Technological Features of Designing Products from Titanium Alloys for Oncological Orthopedics. Proceedings of the International Conference Ti–2008 in the CIS countries, Saint Petersburg, Russia.
31. Kolesnik, A.I., Goneev, S.V., Goncharov, N.N., Sokolenko, N.V., Gavryushenko, N.S., Androsov, V.V., Dokalin, A.Y., Bulgakov, V.G., and Fomin, L.V. (2014, January 15). Method for Express Modeling of the Wear of a Polyethylene Insertion of a Metal Cup or a Polyethylene Cup of the Experimental Module of Hip Joint Endoprosis under Dynamic Conditions. Available online: https://i.moscow/patents/RU2547694C1_20150410.
32. Theoretical and Experimental Analysis of Standardized Methods and Results of Technical Tests of Friction Pairs of Endoprothesis from Titanium Alloy
33. Zwicker, U. (1974). Titan und Titanlegierungen, Springer.10.1007/978-3-642-80587-5 ↗