Towards Additive Manufacture of Functional, Spline-Based Morphometric Models of Healthy and Diseased Coronary Arteries: In Vitro Proof-of-Concept Using a Porcine Template
Source abstract
Abstract
The aim of this study is to assess the additive manufacture of morphometric models of healthy and diseased coronary arteries. Using a dissected porcine coronary artery, a model was developed with the use of computer aided engineering, with splines used to design arteries in health and disease. The model was altered to demonstrate four cases of stenosis displaying varying severity, based on published morphometric data available. Both an Objet Eden 250 printer and a Solidscape 3Z Pro printer were used in this analysis. A wax printed model was set into a flexible thermoplastic and was valuable for experimental testing with helical flow patterns observed in healthy models, dominating the distal LAD (left anterior descending) and left circumflex arteries. Recirculation zones were detected in all models, but were visibly larger in the stenosed cases. Resin models provide useful analytical tools for understanding the spatial relationships of blood vessels, and could be applied to preoperative planning techniques, but were not suitable for physical testing. In conclusion, it is feasible to develop blood vessel models enabling experimental work; further, through additive manufacture of bio-compatible materials, there is the possibility of manufacturing customized replacement arteries.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Anatomy and Medical Technology · Additive Manufacturing and 3D Printing Technologies · Manufacturing Process and Optimization
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Mechanical Engineering, University of Birmingham, Birmingham B15 2TT, UK
References (84)
Show source-provided references
1. Mackay, J., and Mensah, G. (2004). The Atlas of Heart Disease and Stroke, World Health Organisation.
2. Pathophysiology of atherosclerosis plaque progression10.1016/j.hlc.2013.03.001 ↗
3. Nichols, W., O’Rourke, M., and Vlachopoulos, C. (1990). Mcdonald’s Blood Flow in Arteries, CRC Press. [3rd ed.].
4. Steady flow in a model of the human carotid bifurcation. Part I—flow visualization10.1016/0021-9290(82)90057-4 ↗
5. Textbook of surgery: The biological basis of modern surgical practice
6. Anatomy of the heart10.1016/j.mpmed.2010.04.005 ↗
7. Gray’s anatomy for students10.1308/rcsann.2006.88.5.513b ↗
8. Anatomy, histology, and pathology of coronary arteries: A review relevant to new interventional and imaging techniques—Part I10.1002/clc.4960150613 ↗
9. Evolution and progression of atherosclerotic lesions in coronary arteries of children and young adults
10. A new imaging technique to study 3-D plaque and shear stress distribution in human coronary artery bifurcations in vivo10.1016/j.jbiomech.2006.12.007 ↗
11. Blood flow in arteries10.1146/annurev.fluid.29.1.399 ↗
12. Development of physical model of the coronary artery circulation10.1016/j.apcbee.2012.11.032 ↗
13. Computational fluid dynamics analysis of the effect of plaques in the left coronary artery10.1155/2012/504367 ↗
14. Rapid prototyping: A new tool in understanding and treating structural heart disease10.1161/CIRCULATIONAHA.107.740977 ↗
15. Comparison of coronary artery dynamics pre- and poststenting10.1016/S0021-9290(02)00447-5 ↗
16. Current progress in 3D printing for cardiovascular tissue engineering10.1088/1748-6041/10/3/034002 ↗
17. 3D printing of intracranial artery stenosis based on the source images of magnetic resonance angiograph
18. Rapid protyping technology in medicine—basics and applications10.1016/S0895-6111(99)00025-7 ↗
19. Blood flow patterns in an anatomically realistic coronary vessel: Influence of three different reconstruction methods10.1016/S0021-9290(02)00179-3 ↗
20. Determination of in vivo velocity and endothelial shear stress patterns with phasic flow in human coronary arteries: A methodology to predict progression of coronary atherosclerosis10.1067/mhj.2002.123118 ↗
21. Unsteady blood flow and mass transfer of a human left coronary artery bifurcation: FSI vs. CFD10.1016/j.icheatmasstransfer.2012.04.009 ↗
22. Computational simulations and experimental studies of 3D phase-contrast imaging of fluid flow in carotid bifurcation geometries10.1002/jmri.22096 ↗
23. Validated computation of physiologic flow in a realistic coronary artery branch10.1016/S0021-9290(97)00118-8 ↗
24. Helical velocity patterns in a human coronary artery showing the relation with local wall thickness
25. Analysis of the flow in stenosed carotid artery bifurcation models—hydrogen-bubble visualisation10.1016/0021-9290(94)90067-1 ↗
26. Coronary pressure measurement to assess the hemodynamic significance of serial stenoses within one coronary artery: Validation in humans10.1161/01.CIR.102.19.2371 ↗
27. Relation between myocardial blood flow and the severity of coronary artery stenosis10.1056/NEJM199406233302503 ↗
28. Pressure drop across artificially induced stenoses in the femoral arteries of dogs10.1161/01.RES.36.6.735 ↗
29. Flow disturbance measurements through a constricted tube at moderate Reynolds numbers10.1016/0021-9290(83)90096-9 ↗
30. Effect of flow partition on wall shear in a cast of a human coronary artery10.1093/cvr/22.5.340 ↗
31. Disorder distal to modeled stenoses in steady and pulsatile flow10.1016/0021-9290(78)90056-8 ↗
32. Characterization and evolution of poststenotic flow disturbances10.1016/0021-9290(81)90038-5 ↗
33. Laser Doppler anemometer measurements of pulsatile flow in a model carotid bifurcation10.1016/0021-9290(87)90048-0 ↗
34. Pulsatile non-newtonian blood flow in three-dimensional carotid bifurcation models: A numerical study of flow phenomena under different bifurcation angles10.1016/0141-5425(91)90100-L ↗
35. Flow characteristics in models of arterial stenosis—I10.1016/0021-9290(73)90099-7 ↗
36. Biomedical microfluidic devices by using low-cost fabrication techniques: A review10.1016/j.jbiomech.2015.11.031 ↗
37. Microfluidic tools toward industrial biotechnology10.1002/btpr.2350 ↗
38. Wall expansion assessment of an intracranial aneurysm model by a 3D Digital Image Correlation System10.1016/j.measurement.2016.03.045 ↗
39. An FDA viewpoint on unique considerations for medical-device clinical trials10.1056/NEJMra1512592 ↗
40. Lumbar model generator: A tool for the automated generation of a parametric scalable model of the lumbar spine10.1098/rsif.2017.0829 ↗
41. Modeling blood flow through intracranial aneurysms: A comparison of Newtonian and non-Newtonian viscosity10.1007/s40846-016-0142-z ↗
42. Bouteloup, H., de Oliveira Marinho, J.G., Chatpun, S., and Espino, D.M. (2018). Combining magnetic resonance angiography and computational analysis to predict the effect of pre-bifurcation stenosis on the hemodynamics of the internal and external carotid arteries. J. Mech. Med. Biol., submitted.
43. Hemodynamics through the congenitally bicuspid aortic valve: A computational fluid dynamics comparison of opening orifice area and leaflet orientation10.1177/0267659116656775 ↗
44. Evaluation of a transient, simultaneous, arbitrary Lagrange–Euler based multi-physics method for simulating the mitral heart valve10.1080/10255842.2012.688818 ↗
45. Van Der Horst, A., Boogaard, F.L., Van’T Veer, M., Rutten, M.C.M., Pijls, N.H.J., and van de Vosse, F.N. (2013). Towards patient-specific modeling of coronary hemodynamics in healthy and diseased state. Comput. Math. Methods Med., 393792.10.1155/2013/393792 ↗
46. Poly(dimethylsiloxane) thin films as biocompatible coatings for microfluidic devices: Cell culture and flow studies with glial cells10.1002/jbm.a.30166 ↗
47. Surface finish has a critical influence on biofilm formation and mammalian cell attachment to additively manufactured prosthetics10.1021/acsbiomaterials.7b00336 ↗
48. Analysis of errors in medical rapid prototyping models10.1054/ijom.2000.0135 ↗
49. Rapid prototyping models for surgical planning with hard and soft tissue representation10.1016/j.ics.2004.03.139 ↗
50. Latex vessels with customized compliance for use in arterial flow models10.1088/0967-3334/20/3/305 ↗
51. Oscillatory flow in arteries: The constrained elastic tube as a model of arterial flow and pulse transmission10.1088/0031-9155/2/2/305 ↗
52. Reuderink, P.J. (1991). Analysis of the Flow in a 3D Distensible Model of the Carotid Artery Bifurcation. [Master’s Thesis, Eindhoven Technical University].10.1007/978-3-642-52338-0_41 ↗
53. The physical properties of elastic tissue10.1016/B978-0-12-363707-9.50011-3 ↗
54. Duck, F. (1991). Physical Properties of Tissue. A Comprehensive Reference Book. Med. Phys.
55. Viscoelastic properties of mitral valve leaflets: An analysis of regional variation and frequency-dependency10.1177/0954411917719741 ↗
56. Viscoelastic characterisation of chordae tendineae of the mitral valve: Requirements for future replacement materials10.1016/S0140-6736(13)60481-X ↗
57. Frequency and diameter dependent viscoelastic properties of mitral valve chordae tendineae10.1016/j.jmbbm.2013.11.013 ↗
58. Dynamic viscoelasticity and surface properties of porcine left anterior descending coronary arteries10.1007/s13239-016-0288-4 ↗
59. Influence of curvature dynamics on pulsatile coronary artery flow in a realistic bifurcation model10.1016/j.jbiomech.2004.01.021 ↗
60. Non-Newtonian blood flow in human right coronary arteries: Steady state simulations10.1016/j.jbiomech.2003.09.016 ↗
61. Pressure-flow relations in coronary circulation10.1152/physrev.1990.70.2.331 ↗
62. Flow patterns and spatial distribution of atherosclerotic lesions in human coronary arteries10.1161/01.RES.66.4.1045 ↗
63. Computation of hemodynamics in the left coronary artery with variable angulations10.1016/j.jbiomech.2011.04.033 ↗
64. Nonquasi-steady character of pulsatile flow in human coronary arteries10.1115/1.3138514 ↗
65. Effect of geometrical assumptions on numerical modeling of coronary blood flow under normal and disease conditions10.1115/1.4001033 ↗
66. Olson, D. (1972). Fluid Mechanics Relevant to Respiration: Flow Within Curved or Elliptical Tubes and Bifurcating Systems. [Ph.D. Thesis, University of London].
67. Hall, J.E., and Guyton, A.C. (2011). Textbook of Medical Physiology, Saunders/Elsevier. [12th ed.].
68. Steady flow in a model of the human carotid bifurcation. Part II—Laser-Doppler anemometer measurements10.1016/0021-9290(82)90058-6 ↗
69. Effect of exercise on blood flow through the aortic valve: A combined clinical and numerical study10.1080/10255842.2013.771179 ↗
70. Combining numerical and clinical methods to assess aortic valve hemodynamics during exercise10.1177/0267659114521103 ↗
71. Estimation of maximum intraventricular pressure: A three-dimensional fluid–structure interaction model
72. Fluid-Structure Interaction modeling of aortic valve stenosis at different heart rates
73. Effects of freezing, fixation and dehydration on surface roughness properties of porcine left anterior descending coronary arteries10.1016/j.micron.2017.06.009 ↗
74. Investigation of techniques for the measurement of articular cartilage surface roughness10.1016/j.micron.2012.06.007 ↗
75. Thomas-Seale, L.E.J., Kirkman-Brown, J.C., Attallah, M.M., Espino, D.M., and Shepherd, D.E.T. (2018). The barriers to the progression of additive manufacture: An industrial review. Int. J. Prod. Econ., submitted.10.1016/j.ijpe.2018.02.003 ↗
76. The mechanical properties of the rubber elastic polymer polydimethylsiloxane for sensor applications10.1088/0960-1317/7/3/017 ↗
77. The effect of arterial curvature on blood flow in arterio-venous fistulae: Realistic geometries and pulsatile flow10.1007/s13239-017-0321-2 ↗
78. Lumen diameter of normal human coronary arteries. Influence of age, sex, anatomic variation, and left ventricular hypertrophy or dilation10.1161/01.CIR.86.1.232 ↗
79. Shear stress and advanced atherosclerosis in human coronary arteries10.1016/j.jbiomech.2012.11.006 ↗
80. Intravascular ultrasound classification of plaque distribution in left main coronary artery bifurcations where is the plaque really located?10.1161/CIRCINTERVENTIONS.109.906016 ↗
81. Objet Eden 250 (2017, April 01). The 16-Micron-layer 3D Printing System Data Sheet. Available online: http://inventionstudio.gatech.edu/wp-content/uploads/sites/147/2016/01/Eden_250.pdf.
82. SolidScape (2017, April 01). 3Z Pro Data Sheet. Available online: https://www.solidscape.com/wp-content/uploads/2015/03/3Z-PRO-Data-Sheet-Solidscape-3-2-2015.pdf.
83. Simaan, M., Boston, J.R., Miller, P.J., and Antaki, J.F. (1998, January 15). Estimation of blood pump parameters for cardiovascular system identification. Proceedings of the 1998 IEEE International Conference on Acoustics, Speech and Signal Processing, Seattle, WA, USA.
84. Studies in vitro of the relationship between ultrasound and laser Doppler velocimetry and applicability to the simplified Bernoulli relationship10.1161/01.CIR.73.2.300 ↗