Biomechanical Effects of a New Macrogeometry Design of Dental Implants: An In Vitro Experimental Analysis

Sergio Alexandre Gehrke, Leticia Pérez-Díaz, Patricia Mazón, Piedad N. De Aza

Abstract

The purpose of the present study was to measure and compare the insertion torque, removal torque, and the implant stability quotient by resonance frequency analysis in different polyurethane block densities of two implant macrogeometries. Four different polyurethane synthetic bone blocks were used with three cortical thickness: Bone 1 with a cortical thickness of 1 mm, Bone 2 with a cortical thickness of 2 mm, Bone 3 with a cortical thickness of 3 mm, and Bone 4, which was totally cortical. Four groups were created in accordance with the implant macrogeometry (n = 10 per group) and surface treatment: G1—regular implant design without surface treatment; G2—regular implant design with surface treatment; G3—new implant design without surface treatment; G4—new implant design with surface treatment. All implants used were 4 mm in diameter and 10 mm in length and manufactured in commercially pure titanium (grade IV) by Implacil De Bortoli (São Paulo, Brazil). The implants were installed using a computed torque machine, and following installation of the implant, the stability quotient (implant stability quotient, ISQ) values were measured in two directions using Osstell devices. The data were analyzed by considering the 5% level of significance. All implant groups showed similar mean ISQ values without statistical differences (p > 0.05), for the same synthetic bone block: for Bone 1, the value was 57.7 ± 3.0; for Bone 2, it was 58.6 ± 2.2; for Bone 3, it was 60.6 ± 2.3; and for Bone 4, it was 68.5 ± 2.8. However, the insertion torque showed similar higher values for the regular macrogeometry (G1 and G2 groups) in comparison with the new implant macrogeometry (G3 and G4 groups). The analysis of the results found that primary stability does not simply depend on the insertion torque but also on the bone quality. In comparison with the regular implant macrogeometry, the new implant macrogeometry decreased the insertion torque without affecting the implant stability quotient values.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Research, Biotecnos, Cuareim 1483, Montevideo CP 11100, Uruguay

Department of Bioengineering, Instituto de Bioingenieria, Universidad Miguel Hernández, Avda. Ferrocarril s/n, 03202 Elche (Alicante), Spain

Laboratorio de Interacciones Molecular, Facultad de Ciencias, Universidad de la Republica, Calle Iguá 4225, Montevideo 11400, Uruguay

Departamento de Materiales, Óptica Tecnologia Electrónica, Universidad Miguel Hernández, Avda. Ferrocarril s/n, 03202 Elche (Alicante), Spain

References (57)

Show source-provided references
1. Development of a new implant primary stability parameter: Insertion torque revisited10.1111/j.1708-8208.2011.00392.x ↗
2. Matys, J., Świder, K., Grzech-Leśniak, K., Dominiak, M., and Romeo, U. (2019). Photobiomodulation by a 635nm Diode Laser on Peri-Implant Bone: Primary and Secondary Stability and Bone Density Analysis-A Randomized Clinical Trial. Biomed. Res. Int., 2019.10.1155/2019/2785302 ↗
3. Assessment of the primary stability of root analog zirconia implants designed using cone beam computed tomography software by means of the Periotest® device: An ex vivo study. A preliminary report10.17219/acem/65069 ↗
4. Required minimum primary stability and torque values for immediate loading of mini dental implants: An experimental study in nonviable bovine femoral bone10.1016/j.tripleo.2007.10.003 ↗
5. Changes in implant stability using different site preparation techniques: Twist drills versus piezosurgery: A single-blinded, randomized, controlled clinical trial10.1111/j.1708-8208.2011.00341.x ↗
6. Er:YAG laser, piezosurgery, and surgical drill for bone decortication during orthodontic mini-implant insertion: Primary stability analysis-an animal study10.1007/s10103-017-2381-9 ↗
7. Predicting osseointegration by means of im¬plant primary stability10.1111/j.1600-0501.2004.01059.x ↗
8. Relationship between the bone density estimated by cone-beam computed tomography and the primary stability of dental implants10.1111/j.1600-0501.2011.02203.x ↗
9. Influence of surgical technique, implant shape and diameter on the primary stability in cancellous bone10.1111/j.1365-2842.2010.02117.x ↗
10. Evaluation of primary and secondary stability of titanium implants using different surgical techniques10.1111/clr.12180 ↗
11. The combined effects of undersized drilling and implant macrogeometry on bone healing around dental implants: An experimental study10.1016/j.ijom.2014.03.017 ↗
12. The role of primary stability for successful immediate loading of dental implants. A literature review10.1016/j.jdent.2010.05.013 ↗
13. Makary, C., Menhall, A., Zammarie, C., Lombardi, T., Lee, S.Y., Stacchi, C., and Park, K.B. (2019). Primary Stability Optimization by Using Fixtures with Different Thread Depth According to Bone Density: A Clinical Prospective Study on Early Loaded Implants. Materials, 12.10.3390/ma12152398 ↗
14. Biological factors contributing to failures of osseointegrated oral implants. (II). Etiopathogenesis10.1046/j.0909-8836..t01-6-.x ↗
15. Methods used to assess implant stability: Current status
16. Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test10.5051/jpis.2013.43.1.30 ↗
17. A comparison of maxillary and mandibular bone mineral densities10.1016/S0022-3913(98)70245-8 ↗
18. Tsai, W.C., Chen, P.Q., Lu, T.W., Wu, S.S., Shih, K.S., and Lin, S.C. (2009). Comparison and prediction of pullout strength of conical and cylindrical pedicle screws within synthetic bone. BMC Musculoskelet. Disord., 10.10.1186/1471-2474-10-44 ↗
19. Insertion torque and pullout strength of vertebral screws with cylindrical and conic core10.1590/S0102-36162008001000005 ↗
20. Mechanical validation of whole bone composite femur models10.1016/0021-9290(95)00084-4 ↗
21. Mechanical validation of whole bone composite tibia models10.1016/S0021-9290(99)00186-4 ↗
22. Is there a need for standardization of tightening force used to connect the transducer for resonance frequency analysis in determining implant stability?10.11607/jomi.7361 ↗
23. Measurements comparing the initial stability of five designs of dental implants: A human cadaver study
24. Immediate functional loading of Brånemark System implants with enhanced initial stability: A prospective study 1 to 2-year clinical and radiographic study10.1111/j.1708-8208.2003.tb00011.x ↗
25. Correlation of insertion torques with bone mineral density from dental quantitative CT in the mandible10.1034/j.1600-0501.2003.00932.x ↗
26. Primary stability of a conical implant and a hybrid, cylindric screw-type implant in vitro
27. Effect of drilling dimension on implant placement torque and early osseointegration stages: An experimental study in dogs10.1016/j.joms.2011.08.006 ↗
28. Differences in implant stability associated with various methods of preparation of the implant bed: An in vitro study10.1016/S0022-3913(12)60092-4 ↗
29. ASTM F1839-97 (2001). Standard Specification for Rigid Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments, ASTM International.
30. Bone density: A key determinant for clinical success
31. Comuzzi, L., Iezzi, G., Piattelli, A., and Tumedei, M. (2019). An In Vitro Evaluation, on Polyurethane Foam Sheets, of the Insertion Torque (IT) Values, Pull-Out Torque Values, and Resonance Frequency Analysis (RFA) of NanoShort Dental Implants. Polymers, 11.10.3390/polym11061020 ↗
32. Gehrke, S.A., Aramburú Júnior, J., Pérez-Díaz, L., Eirles Treichel, T.L., Dedavid, B.A., De Aza, P.N., and Prados-Frutos, J.C. (2019). New Implant Macrogeometry to Improve and Accelerate the Osseointegration: An In Vivo Experimental Study. Appl. Sci., 9.10.3390/app9153181 ↗
33. Mechanical basis of bone strength: Influence of bone material, bone structure and muscle action
34. The influence of three different apical implant designs at stability and osseointegration process: Experimental study in rabbits10.1111/clr.12807 ↗
35. Does the time of osseointegration in the maxilla and mandible differ?10.1097/SCS.0000000000001067 ↗
36. Does Implant Design Affect Implant Primary Stability? A Resonance Frequency Analysis-Based Randomized Split-Mouth Clinical Trial10.1563/aaid-joi-D-13-00294 ↗
37. The comparison between implant stability quotient and bone-implant contact revisited: An experiment in Beagle dog10.1111/clr.12256 ↗
38. Optimal implant stabilization in low density bone10.1034/j.1600-0501.2001.120501.x ↗
39. Resonance frequency measurements of implant stability in vivo. A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla10.1034/j.1600-0501.1997.080309.x ↗
40. Chatvaratthana, K., Thaworanunta, S., Seriwatanachai, D., and Wongsirichat, N. (2017). Correlation between the thickness of the crestal and buccolingual cortical bone at varying depths and implant stability quotients. PLoS ONE, 12.10.1371/journal.pone.0190293 ↗
41. Correlation between resonance frequency, insertion torque and bone-implant contact in self-cutting threaded implants10.1007/s10266-016-0265-2 ↗
42. Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: A systematic review10.1111/cid.12565 ↗
43. Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: A laboratory study10.1111/j.1600-0501.2009.01823.x ↗
44. Effect of biofunctionalized implant surface on osseointegration—A histomorphometric study in dogs10.1590/S0103-64402009000200001 ↗
45. Comparison of implant stability after different implant surface treatments in dog bone10.1590/S1678-77572010000400016 ↗
46. Factors influencing primary dental implant stability remain unclear10.1016/S1532-3382(12)70033-7 ↗
47. In vivo comparison between laser-treated and grit blasted/acid etched titanium10.1111/clr.12109 ↗
48. A comparison between insertion torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading
49. In vivo study of the effect of treatment on bone on growth on press-fit titanium alloy implants10.1016/j.biomaterials.2004.09.039 ↗
50. Clinical study on the primary stability of two dental implant systems with resonance frequency analysis10.1007/s00784-007-0115-2 ↗
51. An evaluation of resonance frequency analysis for the determination of the primary stability of orthodontic palatal implants. A study in human cadavers10.1111/j.1600-0501.2005.01134.x ↗
52. Correlating implant stability to bone structure10.1111/j.1600-0501.2009.01745.x ↗
53. The effects of superficial roughness and design on the primary stability of dental implants10.1111/j.1708-8208.2009.00202.x ↗
54. Comparison of the primary stabilities of conical and cylindrical endosseous dental implants: An in-vitro study
55. Influence of macro-geometry in the primary stability of implants
56. Clinical study on the insertion torque and wireless resonance frequency analysis in the assessment of torque capacity and stability of self-tapping dental implants10.1111/j.1365-2842.2009.01990.x ↗
57. Implant stability measurements using resonance frequency analysis: Biological and biomechanical aspects and clinical implications10.1111/j.1600-0757.2008.00267.x ↗