The Influence of Insertion Torque on Stress Distribution in Peri-Implant Bones Around Ultra-Short Implants: An FEA Study
Source abstract
Abstract
Using ultra-short dental implants is a promising alternative to extensive bone grafting procedures for patients with atrophic posterior mandibles and vertical bone loss. However, the amount of insertion torque (IT) applied during implant placement significantly influences stress distribution in the peri-implant bone, which affects implant stability and long-term success. Materials and Methods: This study used finite element analysis (FEA) to examine how different insertion torques (35 N·cm and 75 N·cm) affect stress distribution in cortical and trabecular bone types D2 and D4 surrounding ultra-short implants. Von Mises equivalent stress values were compared with ultimate bone strength thresholds to evaluate the potential for microdamage during insertion. Results: The findings demonstrate that increasing IT from 35 N·cm to 75 N·cm led to a significant increase in peri-implant bone stress. Specifically, cortical bone stress in D4 bone increased from approximately 79 MPa to 142 MPa with higher IT, exceeding physiological limits and elevating the risk of microfractures and bone necrosis. In contrast, lower IT values kept stress within safe limits, ensuring optimal primary stability without damaging the bone. These results underscore the need to strike a balance between achieving sufficient implant stability and avoiding mechanical trauma to the surrounding bone. Conclusions: Accurate control of insertion torque during the placement of ultra-short dental implants is crucial to minimize bone damage and promote optimal osseointegration. Excessive torque, especially in low-density bone, can compromise implant success by inducing excessive stress, thereby increasing the risk of early failure.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials · Bone health and osteoporosis research
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Mechanics, Mathematics and Management, Polytechnic University of Bari, 70125 Bari, Italy
Independent Researcher, 00131 Rome, Italy
Independent Researcher, 31020 San Vendemiano, Italy
Independent Researcher, 92019 Sciacca, Italy
Faculty of Dentistry, University of Toronto, Toronto, ON M5G 1G6, Canada
Department of Medical, Oral and Biotechnological Sciences, University of Chieti-Pescara, 66100 Chieti, Italy
References (100+)
Show source-provided references
1. Periodontitis, edentulism, and risk of mortality: A systematic review with meta-analyses10.1177/0022034520952401 ↗
2. Effect of total edentulism on the internal bone structure of mandibular condyle: A preliminary study10.1007/s11282-020-00444-z ↗
3. Stanbouly, D., Zeng, Q., Jou, Y.T., and Chuang, S.K. (2024). Edentulism (missing teeth) and brain central nervous system (CNS) deafferentation: A narrative review. Front. Oral Maxillofac. Med., 6.10.21037/fomm-21-117 ↗
4. Emfietzoglou, R., and Dereka, X. (2024). Survival rates of short dental implants (≤6 mm) used as an alternative to longer (>6 mm) implants for the rehabilitation of posterior partial edentulism: A systematic review of RCTs. Dent. J., 12.10.3390/dj12060185 ↗
5. Vertical augmentation of the atrofic posterior mandibular ridges with onlay grafts: Intraoral blocks vs. guided bone regeneration. Systematic review10.4317/jced.60294 ↗
6. Severely atrophic mandibles restored with fiber-reinforced composite prostheses supported by 5.0-mm ultra-short implants present high survival rates up to eight years10.1016/j.joms.2021.09.018 ↗
7. Implant survival rate and marginal bone loss with the all-on-4 immediate-loading strategy: A clinical retrospective study with 1 to 4 years of follow-up10.1016/j.prosdent.2021.12.020 ↗
8. Zhang, Y., Gulati, K., Li, Z., Di, P., and Liu, Y. (2021). Dental implant nano-engineering: Advances, limitations and future directions. Nanomaterials, 11.10.3390/nano11102489 ↗
9. The Role of Guided Bone Regeneration in Enhancing Dental Implant Success: Guided Bone Regeneration in Dental Implant10.31661/gmj.v13iSP1.3681 ↗
10. Primary stability of immediate implants placed in fresh sockets in comparison with healed sites: A systematic review and meta-analysis
11. Abiram Bala, M.D., and Maloney, W.J. (2023). Cementless Acetabular Revision for Severe Acetabular Bone Loss without Pelvic Discontinuity (Type IIIA and B Defects). Treatment of Acetabular Bone Loss and Chronic Pelvic Discontinuity-E-Book, Elsevier.
12. A Review on Bone Tumor Management: Cutting-Edge Strategies in Bone Grafting, Bone Graft Substitute, and Growth Factors for Defect Reconstruction
13. Laubach, M., Hildebrand, F., Suresh, S., Wagels, M., Kobbe, P., Gilbert, F., Kneser, U., Holzapfel, B.M., and Hutmacher, D.W. (2023). The concept of scaffold-guided bone regeneration for the treatment of long bone defects: Current clinical application and future perspective. J. Funct. Biomater., 14.10.3390/jfb14070341 ↗
14. Ferraz, M.P. (2023). Bone grafts in dental medicine: An overview of autografts, allografts and synthetic materials. Materials, 16.10.3390/ma16114117 ↗
15. Inchingolo, F., Hazballa, D., Inchingolo, A.D., Malcangi, G., Marinelli, G., Mancini, A., Maggiore, M.E., Bordea, I.R., Scarano, A., and Farronato, M. (2022). Innovative concepts and recent breakthrough for engineered graft and constructs for bone regeneration: A literature systematic review. Materials, 15.10.3390/ma15031120 ↗
16. Tomasetti, B.J., and Ewers, R. (2020). Short Implants, Springer. No. 180175.10.1007/978-3-030-44199-9 ↗
17. Short implants versus bone augmentation for placing longer implants in atrophic maxillae: One-year post-loading results of a pilot randomised controlled trial
18. Berzaghi, A., Testori, T., Scaini, R., and Bortolini, S. (2025). Occlusion and Biomechanical Risk Factors in Implant-Supported Full-Arch Fixed Dental Prostheses—Narrative Review. J. Pers. Med., 15.10.3390/jpm15020065 ↗
19. Occlusal load considerations in implant-supported fixed restorations10.3390/prosthesis2040023 ↗
20. Early loading of implants in the atrophic posterior maxilla: Lateral sinus lift with autogenous bone and Bio-Oss versus crestal mini sinus lift and 8-mm hydroxyapatite-coated implants. A randomised controlled clinical trial
21. Bone augmentation versus 5-mm dental implants in posterior atrophic jaws. Four-month post-loading results from a randomised controlled clinical trial
22. Short implants versus bone augmentation in combination with standard-length implants in posterior atrophic partially edentulous mandibles: Systematic review and meta-analysis with the Bayesian approach10.1016/j.ijom.2018.05.009 ↗
23. Short implants versus longer implants in vertically augmented posterior mandibles: A randomised controlled trial with 5-year after loading follow-up
24. Dental implant length and diameter: A proposed classification scheme10.1111/jopr.12517 ↗
25. Lombardo, G., Signoriello, A., Marincola, M., Liboni, P., Faccioni, P., Zangani, A., D’agostino, A., and Nocini, P.F. (2022). Short and ultra-short implants, in association with simultaneous internal sinus lift in the atrophic posterior maxilla: A five-year retrospective study. Materials, 15.10.3390/ma15227995 ↗
26. Short Implants Versus Bone Augmentation and Longer Implants in Atrophic Maxillae. Five Year Post Loading Results of a Randomised Controlled Trial10.36130/CTD.03.2020.04 ↗
27. Load transfer along the bone–dental implant interface10.1016/j.jbiomech.2010.02.017 ↗
28. Lee, C.H., Mukundan, A., Chang, S.C., Wang, Y.L., Lu, S.H., Huang, Y.C., and Wang, H.C. (2021). Comparative analysis of stress and deformation between one-fenced and three-fenced dental implants using finite element analysis. J. Clin. Med., 10.10.3390/jcm10173986 ↗
29. A review article on factors affecting bone loss in dental implants10.1016/j.matpr.2020.07.428 ↗
30. Biomechanical analysis of alveolar bone stress around implants with different thread designs and pitches in the mandibular molar area10.1007/s00784-011-0517-z ↗
31. Three interfaces of the dental implant system and their clinical effects on hard and soft tissues10.1039/D1MH01621K ↗
32. Finite element analysis in implant dentistry: State of the art and future directions10.1016/j.dental.2023.04.002 ↗
33. Anitua, E., Larrazabal Saez de Ibarra, N., Morales Martín, I., and Saracho Rotaeche, L. (2021). Influence of dental implant diameter and bone quality on the biomechanics of single-crown restoration. A finite element analysis. Dent. J., 9.10.3390/dj9090103 ↗
34. Physiological functions of podosomes: From structure and function to therapy implications in osteoclast biology of bone resorption10.1016/j.arr.2023.101842 ↗
35. Effect of insertion load on insertion torque value10.1016/j.jds.2025.03.029 ↗
36. Bone healing at non-submerged implants installed with different insertion torques: A split-mouth histomorphometric randomized controlled trial10.1186/s40729-019-0194-2 ↗
37. Primary stability, insertion torque and bone density of cylindric implant ad modum Branemark: Is there a relationship? An in vitro study10.1111/j.1600-0501.2010.02036.x ↗
38. The clinical effects of insertion torque for implants placed in healed ridges: A two-year randomized controlled clinical trial
39. High insertion torque versus regular insertion torque: Early crestal bone changes on dental implants in relation to primary stability—A retrospective clinical study10.1186/s40729-024-00540-3 ↗
40. Correlation of insertion torques with bone mineral density from dental quantitative CT in the mandible10.1034/j.1600-0501.2003.00932.x ↗
41. Clinical effect of the high insertion torque on dental implants: A systematic review and meta-analysis10.1016/j.prosdent.2020.06.012 ↗
42. Application of the finite element method in dental implant research10.1080/10255840600837074 ↗
43. Finite-element analysis of stress on dental implant prosthesis10.1111/j.1708-8208.2008.00142.x ↗
44. Callea, C., Ceddia, M., Piattelli, A., Specchiulli, A., and Trentadue, B. (2023). Finite element analysis (FEA) for a different type of cono-in dental implant. Appl. Sci., 13.10.3390/app13095313 ↗
45. The role of the finite element model in dental implants10.1563/1548-1336(2000)026<0077:TROTFE>2.3.CO;2 ↗
46. Static, dynamic and fatigue behaviors of dental implant using finite element method10.1016/j.advengsoft.2006.02.004 ↗
47. Evaluating Parameters of osseointegrated dental implants using finite element analysis a two-dimensional comparative study examining the effects of implant diameter, implant shape, and load direction10.1563/1548-1336(1998)024<0080:EPOODI>2.3.CO;2 ↗
48. A comprehensive biomechanical evaluation of length and diameter of dental implants using finite element analyses: A systematic review10.1016/j.heliyon.2024.e26876 ↗
49. Biomechanical evaluations of the long-term stability of dental implant using finite element modeling method: a systematic review10.4047/jap.2022.14.3.182 ↗
50. Review on finite element analysis of dental implants10.54527/jdir.2022.41.3.50 ↗
51. Biomechanical finite element analysis of small diameter and short dental implant10.1515/bmt.2010.049 ↗
52. Biomechanical behavior analysis of four types of short implants with different placement depths using the finite element method10.1016/j.prosdent.2023.01.005 ↗
53. Alqahtani, A.R., Desai, S.R., Patel, J.R., Alqhtani, N.R., Alqahtani, A.S., Heboyan, A., Fernandes, G.V.O., Mustafa, M., and Karobari, M.I. (2023). Investigating the impact of diameters and thread designs on the Biomechanics of short implants placed in D4 bone: A 3D finite element analysis. BMC Oral Health, 23.10.1186/s12903-023-03370-8 ↗
54. Evaluation of design parameters of eight dental implant designs: A two-dimensional finite element analysis10.4103/1119-3077.97308 ↗
55. Bone density: A key determinant for treatment planning
56. Evaluation of postoperative dental implant primary stability using 3D finite element analysis10.1080/10255842.2018.1552682 ↗
57. Alemayehu, D.B., and Jeng, Y.R. (2021). Three-dimensional finite element investigation into effects of implant thread design and loading rate on stress distribution in dental implants and anisotropic bone. Materials, 14.10.20944/preprints202109.0221.v1 ↗
58. Mapping of trabecular bone anisotropy and volume fraction in 3D using μCT images of the human calcaneus10.1002/ajpa.24474 ↗
59. A comparative finite element simulation of stress in dental implant–bone interface using isotropic and orthotropic material models in three mastication cycles10.1007/s40430-018-1409-9 ↗
60. Gasik, M., Lambert, F., and Bacevic, M. (2021). Biomechanical properties of bone and mucosa for design and application of dental implants. Materials, 14.10.3390/ma14112845 ↗
61. On the relation of bone mineral density and the elastic modulus in healthy and pathologic bone10.1007/s11914-018-0449-5 ↗
62. Mathematical relationships between bone density and mechanical properties: A literature review10.1016/j.clinbiomech.2007.08.024 ↗
63. Ceddia, M., Lamberti, L., and Trentadue, B. (2024). FEA comparison of the mechanical behavior of three dental crown materials: Enamel, ceramic, and zirconia. Materials, 17.10.3390/ma17030673 ↗
64. Nikam, N., Shenoy, B.S., KN, C., Keni, L.G., Shetty, S., and Bhat, N.S. (2025). Advancements in Surface Coatings for Enhancing Longevity in Hip Implants: A Review. Prosthesis, 7.10.3390/prosthesis7010021 ↗
65. Ceddia, M., Solarino, G., Pulcrano, A., Benedetto, A., and Trentadue, B. (2025). Finite Element Analysis of a 3D-Printed Acetabular Prosthesis for an Acetabular Defect According to the Paprosky Classification. Materials, 18.10.3390/ma18061295 ↗
66. Effects of material and coefficient of friction on taper joint dental implants10.1016/j.jpor.2019.10.003 ↗
67. Trabecular bone structure and strength-remodelling and repair
68. Post-yield and failure properties of cortical bone10.1038/bonekey.2016.60 ↗
69. Elias, C.N. (2011). Factors affecting the success of dental implants. Implant Dentistry: A Rapidly Evolving Practice, InTech.
70. Factors influencing the fracture of dental implants10.1111/cid.12572 ↗
71. The influence of implant geometry on the stress distribution around dental implants10.1002/jbm.820160406 ↗
72. A comprehensive review of factors affecting fatigue life of dental implants10.1016/j.matpr.2020.08.414 ↗
73. Some factors affecting dental implant design10.1002/jbm.820090428 ↗
74. Stress distribution around dental implants placed at different depths10.1097/SCS.0000000000002119 ↗
75. Albanchez-Gonzalez, M.I., Brinkmann, J.C.B., Pelaez-Rico, J., Lopez-Suarez, C., Rodriguez-Alonso, V., and Suarez-Garcia, M.J. (2022). Accuracy of digital dental implants impression taking with intraoral scanners compared with conventional impression techniques: A systematic review of in vitro studies. Int. J. Environ. Res. Public Health, 19.10.3390/ijerph19042026 ↗
76. Accuracy and efficiency of robotic dental implant surgery with different human-robot interactions: An in vitro study10.1016/j.jdent.2023.104642 ↗
77. Storelli, S., Caputo, A., Palandrani, G., Peditto, M., Del Fabbro, M., Romeo, E., and Oteri, G. (2021). Use of Narrow-Diameter Implants in Completely Edentulous Patients as a Prosthetic Option: A Systematic Review of the Literature. BioMed Res. Int., 2021.10.1155/2021/5571793 ↗
78. Implementation of Ultra-Short Implants into Dental Practice: A Systematic Review
79. Evaluation of short and ultra-short dental implants in challenging clinical situations of resorbed ridges: A narrative review
80. Ultra-short versus standard-length dental implants in conjunction with osteotome-mediated sinus floor elevation: A randomized controlled clinical trial10.1111/cid.12995 ↗
81. Wang, M., Liu, F., Ulm, C., Shen, H., and Rausch-Fan, X. (2022). Short Implants versus Longer Implants with Sinus Floor Elevation: A Systemic Review and Meta-Analysis of Randomized Controlled Trials with a Post-Loading Follow-Up Duration of 5 Years. Materials, 15.10.3390/ma15134722 ↗
82. Short versus standard implants at sinus augmented sites: A systematic review and meta-analysis10.1007/s00784-022-04628-1 ↗
83. Yuan, X., Liu, Y., Yang, Y., Ren, M., Luo, L., Zheng, L., and Liu, Y. (2023). Effect of short implant crown-to-implant ratio on stress distribution in anisotropic bone with different osseointegration rates. BMC Oral Health, 23.10.1186/s12903-023-03379-z ↗
84. Clinical performance of short and extrashort dental implants with wide diameter: A systematic review with meta-analysis10.1016/j.prosdent.2023.01.004 ↗
85. Comuzzi, L., Tumedei, M., Romasco, T., Petrini, M., Afrashtehfar, K.I., Inchingolo, F., Piattelli, A., and Di Pietro, N. (2022). Insertion torque, removal torque, and resonance frequency analysis values of ultrashort, short, and standard dental implants: An in vitro study on polyurethane foam sheets. J. Funct. Biomater., 14.10.3390/jfb14010010 ↗
86. Ceddia, M., Romasco, T., Marchioli, G., Comuzzi, L., Cipollina, A., Piattelli, A., Lamberti, L., Di Pietro, N., and Trentadue, B. (2025). Finite Element Analysis of Implant Stability Quotient (ISQ) and Bone Stresses for Implant Inclinations of 0°, 15°, and 20°. Materials, 18.10.3390/ma18071625 ↗
87. Comuzzi, L., Romasco, T., Mourão, C.F., Marchioli, G., Piattelli, A., and Di Pietro, N. (2025). Biomechanical Analysis of Truncated Cone Implants for Maxillary Sinus Lift: An In Vitro Study on Polyurethane Laminas. Bioengineering, 12.10.3390/bioengineering12010053 ↗
88. Evaluation of rotational stability and stress shielding of a stem optimized for hip replacements—A finite element study10.3390/prosthesis5030048 ↗
89. Ficarella, E., Minooei, M., Santoro, L., Toma, E., Trentadue, B., De Spirito, M., Papi, M., Pruncu, C.I., and Lamberti, L. (2021). Visco-hyperelastic characterization of the equine immature zona pellucida. Materials, 14.10.3390/ma14051223 ↗
90. Ceddia, M., Solarino, G., Tucci, M., Lamberti, L., and Trentadue, B. (2024). Stress analysis of tibial bone using three different materials for bone fixation plates. J. Compos. Sci., 8.10.3390/jcs8090334 ↗
91. Ceddia, M., Solarino, G., Giannini, G., De Giosa, G., Tucci, M., and Trentadue, B. (2024). A Finite Element Analysis Study of Influence of Femoral Stem Material in Stress Shielding in a Model of Uncemented Total Hip Arthroplasty: Ti-6Al-4V versus Carbon Fibre-Reinforced PEEK Composite. J. Compos. Sci., 8.10.3390/jcs8070254 ↗
92. Bone stress and damage distributions during dental implant insertion: A novel dynamic FEM analysis10.1080/10255842.2021.2012765 ↗
93. Stepwise analysis of the dental implant insertion process using the finite element technique10.1111/j.1600-0501.2007.01427.x ↗
94. An effect of various parameters on insertion torque to improve the success rate of dental implantation: A review10.1016/j.matpr.2020.07.222 ↗
95. Maximum Insertion Torque, Stress, and Strain in Bone during Insertion of Three Dental Implants with Different Macroscopic Designs: A Dynamic Finite-Element Analysis10.1615/JLongTermEffMedImplants.2021037917 ↗
96. The influence of truncated-conical implant length on primary stability in maxillary and mandibular regions: An in vitro study using polyurethane blocks10.1007/s00784-023-05444-x ↗
97. Prediction of insertion torque and stiffness of a dental implant in bovine trabecular bone using explicit micro-finite element analysis10.1016/j.jmbbm.2019.06.024 ↗
98. Insertion torque of immediate wide-diameter implants: A finite element analysis
99. Prediction of dental implant torque with a fast and automatic finite element analysis: A pilot study10.1016/j.tripleo.2009.11.010 ↗
100. Implant insertion torque and marginal bone loss: A systematic review and meta-analysis