journal article·2023-06-13Open access

Osteoblastic and Bacterial Response of Hybrid Dental Implants

Daniel Robles, Aritza Brizuela, Manuel Fernández-Domínguez, Javier Gil

Abstract

Bacterial infections in dental implants generate peri-implantitis disease that causes bone loss and the mobility of the dental implant. It is well known that specific ranges of roughness favor the proliferation of bacteria, and it is for this reason that new dental implants called hybrids have appeared. These implants have a smooth area in the coronal part and a rough surface in the apical part. The objective of this research is the physico-chemical characterization of the surface and the osteoblastic and microbiological behavior. One-hundred and eighty discs of titanium grade 3 with three different surfaces (smooth, smooth–rough, and completely rough) were studied. The roughness was determined by white light interferometry, and the wettability and surface energy by the sessile drop technique and the application of Owens and Wendt equations. Human osteoblast SaOS-2 was cultured to determine cell adhesion, proliferation, and differentiation. Microbiological studies were performed with two common bacterial strains in oral infection, E. faecalis and S. gordonii, at different times of culture. The roughness obtained for the smooth surface was Sa = 0.23 and for the rough surface it was 1.98 μm. The contact angles were more hydrophilic for the smooth surface (61.2°) than for the rough surface (76.1°). However, the surface energy was lower for the rough surface (22.70 mJ/m2) in both its dispersive and polar components than the smooth surface (41.77 mJ/m2). Cellular activity in adhesion, proliferation, and differentiation was much higher on rough surfaces than on smooth surfaces. After 6 h of incubation, the osteoblast number in rough surfaces was more than 32% higher in relation to the smooth surface. The cell area in smooth surfaces was higher than rough surfaces. The proliferation increased and the alkaline phosphatase presented a maximum after 14 days, with the mineral content of the cells being higher in rough surfaces. In addition, the rough surfaces showed greater bacterial proliferation at the times studied and in the two strains used. Hybrid implants sacrifice the good osteoblast behavior of the coronal part of the implant in order to obstruct bacterial adhesion. The following fact should be considered by clinicians: there is a possible loss of bone fixation when preventing peri-implantitis.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Translational Medicine, CEU San Pablo University, Urbanización Montepríncipe, 28925 Madrid, Spain

Facultad de Odontología, Universidad Europea Miguel de Cervantes, C/del Padre Julio Chevalier 2, 47012 Valladolid, Spain

Department of Oral and Maxillofacial Surgery, University Hospital Monteprincipe, University CEU San Pablo, Av. de Montepríncipe, s/n, 28668 Madrid, Spain

Bioengineering Institute of Technology, Facultad de Medicina y Ciencias de la Salud, Universidad Internacional de Catalunya, Josep Trueta s/n. Sant Cugat del Vallés, 08195 Barcelona, Spain

References (54)

Show source-provided references
1. Implant treatment in the edentulous mandible: A prospective study on Branemark system implants over more than 20 years
2. A systematic review of the survival and complication rates of implant-supported fixed dental prostheses (FDPs) after a mean observation period of at least 5 years10.1111/j.1600-0501.2012.02546.x ↗
3. Intraosseous anchorage of dental prostheses. I. Experimental studies
4. The Effect of Titanium Surface Modifications on Dental Implant Osseointegration10.1159/000381694 ↗
5. Peri-implantitis–onset and pattern of progression10.1111/jcpe.12535 ↗
6. Effectiveness of implant therapy analyzed in a Swedish population: Early and late implant loss10.1177/0022034514563077 ↗
7. Are peri-implantitis lesions different from periodontitis lesions?10.1111/j.1600-051X.2010.01672.x ↗
8. (1983). Kasemo B: Biocompatibility of titanium implants: Surface science aspects. J. Prosthet. Dent., 49, 832–837.10.1016/0022-3913(83)90359-1 ↗
9. Bambini, F., Orilisi, G., Quaranta, A., and Memè, L. (2021). Biological Oriented Immediate Loading: A New Mathematical Implant Vertical Insertion Protocol, Five-Year Follow-Up Study. Materials, 14.10.3390/ma14020387 ↗
10. Effects of implant surface coatings and composition on bone integration: A systematic review10.1111/j.1600-0501.2009.01777.x ↗
11. Survival of hydroxypatite-coated implants: A metaanalytic review10.1053/joms.2000.18269 ↗
12. Optimum surface properties of oxidized implants for reinforcement of osseointegration: Surface chemistry, oxide thickness, porosity, roughness, and crystal structure
13. Oral implant surfaces: Part 1—Review focusing on topographic and chemical properties of different surfaces and in vivo responses to them
14. A prospective, multicenter, randomized controlled 5-year study of hybrid and fully etched implants for the incidence of periimplantitis10.1902/jop.2009.090492 ↗
15. Effects of titanium surface topography on bone integration: A systematic review10.1111/j.1600-0501.2009.01775.x ↗
16. A review on the wettability of dental implant surfaces I: Theoretical and experimental aspects10.1016/j.actbio.2014.02.040 ↗
17. Surface characteristics of dental implants: A review10.1016/j.dental.2017.09.007 ↗
18. Nonlinear Dependence of the Contact Angle of Nanodroplets on Contact Line Curvature10.1103/PhysRevLett.91.186101 ↗
19. A review on the Wettability of Dental Surfaces II: Biological and Clinical Aspects10.1016/j.actbio.2014.03.032 ↗
20. Clinical characteristics at implants with a history of progressive bone loss10.1111/j.1600-0501.2007.01448.x ↗
21. Jacobs, J.J., Andersson, G.B.J., Bell, J.E., Weinstein, S.L., Dormans, J.P., Gnatz, S.M., Lane, N., Puzas, J.E., St. Clair, E.W., and Yelin, E.H. (2008). United States Bone and Joint Decade: The Burden of Musculoskeletal Diseases in the United States, American Academy of Orthopaedic Surgeons. [1st ed.].
22. On the influence of flow conditions and wettability on blood material interactions10.1016/S0142-9612(03)00380-6 ↗
23. Role of trapped air in the formation of cell-and-protein micropatterns on superhydrophobic/superhydrophilic microtemplated surfaces10.1016/j.biomaterials.2012.08.017 ↗
24. Analysis for determining surface free energy uncertainty by the Owen–Wendt method10.1016/j.ijadhadh.2008.09.008 ↗
25. Velasco-Ortega, E., Ortiz-García, I., Jiménez-Guerra, A., Monsalve-Guil, L., Muñoz-Guzón, F., Perez, R.A., and Gil, F.J. (2019). Comparison between Sandblasted Acid-Etched and Oxidized Titanium Dental Implants: In Vivo Study. Int J. Mol. Sci., 20.10.3390/ijms20133267 ↗
26. Surface treatments of titanium dental implants for rapid osseointegration10.1016/j.dental.2006.06.025 ↗
27. Wettability and osteoblastic cell adhesion on ultrapolished commercially pure titanium surfaces: The role of the oxidation and pollution states10.1080/01694243.2014.893815 ↗
28. The influence of blasting and sterilization on static and time-related wettability and surface-energy properties of titanium surfaces10.1016/j.surfcoat.2007.12.021 ↗
29. Effect of roughness, wettability and morphology of engineered titanium surfaces on os-teoblast-like cell adhesion10.1016/j.colsurfa.2009.12.017 ↗
30. Controls in wettability of TiOx films for biomedical applications10.1016/j.surfcoat.2010.08.052 ↗
31. Surface energy characterization of unalloyed titanium implants10.1002/jbm.820281206 ↗
32. Wettability and in-vitro study of titanium surface profiling prepared by electrolytic plasma processing10.1016/j.surfcoat.2021.127119 ↗
33. Spatial Organization of Osteoblast Fibron-ectin-Matrix on Titanium Surface—Effects of Roughness, Chemical Heterogeneity, and Surface Free Energy10.1016/j.actbio.2009.07.030 ↗
34. Importance of the roughness and residual stresses of dental implants on fatigue and osseointegration behavior. In vivo study in rabbits10.1563/aaid-joi-D-16-00088 ↗
35. The effect of shot blasting and heat treatment on the fatigue behavior of titanium for dental implant applications10.1016/j.dental.2006.03.003 ↗
36. Antibacterial coatings on titanium surfaces: A comparison study between in vitro single-species and multispecies biofilm
37. Corrosion behaviour of commercially pure tianium shot blasted with different materials and sizes of shot particles for dental implant applications10.1016/S0142-9612(02)00314-9 ↗
38. The Developmental Control of Osteoblast-Specific Gene Expression: Role of Specific Transcription Factors and the Extracellular Matrix Environment10.1177/10454411990100010201 ↗
39. Influence of surface wettability on competitive protein adsorption and initial attachment of osteoblasts10.1088/1748-6041/4/4/045002 ↗
40. Mechanisms Involved in Osteoblast Response to Implant Surface Morphology10.1146/annurev.matsci.31.1.357 ↗
41. Surface roughness mediates its effects on osteoblasts via protein kinase A and phospholipase A210.1016/S0142-9612(99)00159-3 ↗
42. Thermal treatment to increase titanium wettability induces selective proteins adsorption from blood serum thus affecting osteoblasts adhesion10.1016/j.msec.2019.110250 ↗
43. Enhanced osteoblastic adhesion through improved wettability on polarized hydroxyapatite10.2109/jcersj2.118.474 ↗
44. Hydrophobicity, surface tension, and zeta potential measurements of glass-reinforced hydroxyapatite composites10.1002/(SICI)1097-4636(19990615)45:4<370::AID-JBM12>3.0.CO;2-0 ↗
45. Influence of titanium surfaces on attachment of osteoblast-like cells in vitro10.1016/S0142-9612(03)00571-4 ↗
46. Mimicking bone extracellular matrix: Integrin-binding peptidomimetics enhance osteoblast-like cells adhesion, proliferation, and differentiation on titanium10.1016/j.colsurfb.2014.12.057 ↗
47. Peri-implant diseases and conditions: Consensus report of workgroup 4 of the 2017 World Workshop on the Classification of Periodontal and Peri-Implant Diseases and Conditions10.1002/JPER.17-0739 ↗
48. How frequent does peri-implantitis occur? A systematic review and meta-analysis10.1007/s00784-017-2276-y ↗
49. Buxadera-Palomero, J., Godoy-Gallardo, M., Molmeneu, M., Punset, M., and Gil, F.J. (2020). Antibacterial Properties of Triethoxysilylpropyl Succinic Anhydride Silane (TESPSA) on Titanium Dental Implants. Polymers, 12.10.3390/polym12040773 ↗
50. Anhydride-functional silane immobilized onto titanium surfaces induces osteoblast cell differentiation and reduces bacterial adhesion and biofilm formation10.1016/j.msec.2015.10.051 ↗
51. In vitro evaluation of a multispecies oral biofilm over antibacterial coated titanium surfaces10.1007/s10856-018-6168-8 ↗
52. Punset, M., Vilarrasa, J., Nart, J., Manero, J.M., Bosch, B., Padrós, R., Perez, R.A., and Gil, J. (2021). Citric Acid Passivation of Titanium Dental Implants for Minimizing Bacterial Colonization Impact. Coatings, 11.10.3390/coatings11020214 ↗
53. Efficacy of hybrid implants in oral and maxillofacial surgery: A clinical prospective study10.4103/njms.NJMS_61_17 ↗
54. Robles, D., Brizuela, A., Fernández-Domínguez, M., and Gil, J. (2023). Corrosion Resistance and Titanium Ion Release of Hybrid Dental Implants. Materials, 16.10.3390/ma16103650 ↗