Effect of Implantoplasty on Roughness, Fatigue and Corrosion Behavior of Narrow Diameter Dental Implants
Source abstract
Abstract
Implantoplasty (IP) is used in dental implants with peri-implantitis and aims to remove threads and polish rough surfaces in order to prevent bacterial colonization. As a result of this procedure, implant strength might be compromised. We tested 20 tapered screw-shaped Ti6Al4V dental implants with a simulated bone loss of 50%. Ten implants underwent IP and 10 served as controls. Surface topography (Sa, Sz, Ssk, and Sdr) was analyzed with a confocal optical microscope. Subsequently, a minimum of four series of cyclic loads were applied with a servo-hydraulic mechanical testing machine (5 × 106 cycles at 15 Hz, between a maximal nominal value–starting at 529 N in the IP group and 735 N in the control group–and 10% of that force). We recorded the number of cycles until failure and the type of failure. Implant failure was analyzed by visual inspection and scanning electron microscopy. Open circuit potential and potenctiodynamic tests were carried out with high precision potentiostat using Hank’s solution at 37 °C to evaluate the effect of the implantoplasty on the corrosion resistance. Implantoplasty significantly reduced the surface topography values (median) and interquartile range (IQR); Sa from 1.76 (IQR = 0.11) to 0.49 (IQR = 0.16), Sz from 20.98 (IQR = 8.14) to 8.19 (IQR = 4.16), Ssk from 0.01 (IQR = 0.34) to −0.74 (IQR = 0.53) and Sdr from 18.20 (IQR = 2.26) to 2.67 (IQR = 0.87). The fatigue limits of the control and implantoplasty groups were 551 N and 529 N, respectively. The scanning electron micrographs showed fatigue striations indicating fatigue failure. Besides, the fractographic analysis revealed a typical brittle intergranular fracture mechanism. The infinite life range of the dental implants evaluated was largely above the threshold of usual chewing forces. Implantoplasty seems to render a fairly smooth surface and has a limited impact on fatigue resistance. In addition, implantoplasty produces a decrease in the corrosion resistance of the implant. Corrosion current density from 0.019 μA/cm2 for as-received to 0.069 μA/cm2 in the interface smooth-roughened dental implant. These places between the machining and the rough area of the implant are the most susceptible, with the appearance of pitting.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Orthopaedic implants and arthroplasty · Dental materials and restorations
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Oral Surgery and Implantology, Faculty of Medicine and Health Sciences, University of Barcelona, Campus de Bellvitg, C/Feixa Llarga, s/n, Pavelló Govern, L’Hospitalet de Llobregat, 08907 Barcelona, Spain
Bellvitge Biomedical Research Institute, Oral Surgery and Implantology, (IDIBELL), 08907 Barcelona, Spain
Bioengineering Institute of Technology, Facultad de Medicina y Ciencia de la Salud, Universitat Internacional de Catalunya, Sant Cugat del Vallés, 08907 Barcelona, Spain
References (66)
Show source-provided references
1. A systematic review of the survival and complication rates of implant-supported fixed dental prostheses (FDPs) after a mean observation period of at least 5 years10.1111/j.1600-0501.2012.02546.x ↗
2. Prevalences of peri-implantitis and peri-implant mucositis: Systematic review and meta-analysis10.1016/j.jdent.2017.04.011 ↗
3. Systematic review of the survival rate and the incidence of biological, technical, and aesthetic complications of single crowns on implants reported in longitudinal studies with a mean follow-up of 5 years10.1111/j.1600-0501.2012.02547.x ↗
4. Peri-implant diseases and conditions: Consensus report of workgroup 4 of the 2017 World Workshop on the Classification of Periodontal and Peri-Implant Diseases and Conditions10.1002/JPER.17-0739 ↗
5. Peri-implantitis10.1002/JPER.16-0350 ↗
6. Microbiologic results after non-surgical erbium-doped:yttrium, aluminum, and garnet laser or air-abrasive treatment of peri-implantitis: A randomized clinical trial10.1902/jop.2011.100660 ↗
7. Topical minocycline microspheres versus topical chlorhexidine gel as an adjunct to mechanical debridement of incipient peri-implant infections: A randomized clinical trial10.1111/j.1600-051X.2006.00919.x ↗
8. Mechanical non-surgical treatment of peri-implantitis: A single-blinded randomized longitudinal clinical study. II. Microbiological results10.1111/j.1600-051X.2010.01561.x ↗
9. Non-surgical therapeutic outcomes of peri-implantitis: 12-month results10.1007/s00784-019-02943-8 ↗
10. How do implant surface characteristics influence peri-implant disease?10.1111/j.1600-051X.2010.01661.x ↗
11. Management of peri-implant mucositis and peri-implantitis10.1111/prd.12049 ↗
12. The influence of implantoplasty on the diameter, chemical surface composition, and biocompatibility of titanium implants10.1007/s00784-016-2030-x ↗
13. Therapy of peri-implantitis with resective surgery10.1111/j.1600-0501.2004.01084.x ↗
14. Therapy of peri-implantitis with resective surgery. A 3-year clinical trial on rough screw-shaped oral implants. Part II: Radiographic outcome10.1111/j.1600-0501.2006.01318.x ↗
15. Impact of implantoplasty on strength of the implant-abutment complex10.11607/jomi.3227 ↗
16. Analysis of implant strength after implantoplasty in three implant-abutment connection designs: An in vitro study10.11607/jomi.4399 ↗
17. Influence of implantoplasty on stress distribution of exposed implants at different bone insertion levels10.1590/1807-3107bor-2017.vol31.0096 ↗
18. Does implantoplasty affect the failure strength of narrow and regular diameter implants? A laboratory study10.1007/s00784-020-03534-8 ↗
19. Effect of implantoplasty on fracture resistance and surface roughness of standard diameter dental implants10.1111/clr.13037 ↗
20. Corrosion behaviour of commercially pure tianium shot blasted with different materials and sizes of shot particles for dental implant applications10.1016/S0142-9612(02)00314-9 ↗
21. Toledano-Serrabona, J., Sánchez-Garcés, M.A., Gay-Escoda, C., Valmaseda-Castellon, E., Camps-Font, O., Verdeguer, P., Molmeneu, M., and Gil, F.J. (2021). Mechanical properties and corrosión behavior of Ti6Al4V particles obtained by Implatoplasty. An in vivo study. Part II. Materials, 14.10.3390/ma14216519 ↗
22. (2014). Standard Reference Test Method for Making Potentiostatic and Potentiodynamic Anodic Polarization Measurements, ASTM International. Technical Report no. ASTM G5-14e1.
23. (2009). Biological Evaluation of Medical Devices. Part 5: Tests for In Vitro Cytotoxicity (Standard No. ISO 10993-5:2009).
24. (2010). Standard Practice for Calculation of Corrosion Rates and Related Information from Electrochemical Measurements (Standard No. ASTM G-102-89).
25. Biocompatible carbohydrate-functionalized stainless steel surfaces: A new method for passivating biomedical implants10.1021/am200158y ↗
26. Effect of passivation and dry heat-sterilization on surface energy and topography of unalloyed titanium implants10.1016/S0927-7757(97)00237-9 ↗
27. Corrosion resistance measurements of dental alloys, are they correlated?10.1016/j.dental.2006.06.008 ↗
28. Titanium Corrosion Mechanisms in the Oral Environment: A Retrieval Study10.3390/ma6115258 ↗
29. Implant-abutment interface design affects fatigue and fracture strength of implants10.1111/j.1600-0501.2008.01581.x ↗
30. In vivo vertical forces on implants
31. Oral implant surfaces: Part 1—Review focusing on topographic and chemical properties of different surfaces and in vivo responses to them
32. Surface roughness of dental implants and treatment time using six different implantoplasty procedures10.1111/clr.12682 ↗
33. The influence of implantoplasty on surface roughness, biofilm formation, and biocompatibility of titanium implants: A systematic review10.11607/jomi.8785 ↗
34. Beheshti Maal, M., Aanerød Ellingsen, S., Reseland, J.E., and Verket, A. (2020). Experimental implantoplasty outcomes correlate with fibroblast growth in vitro. BMC Oral Health, 20.10.1186/s12903-020-1012-1 ↗
35. In vitro evaluation of peri-implantitis treatment modalities on Saos-2osteoblasts10.1111/clr.12686 ↗
36. Toma, S., Behets, C., Brecx, M.C., and Lasserre, J.F. (2018). In vitro comparison of the efficacy of peri-implantitis treatments on the removal and recolonization of streptococcus gordonii biofilm on titanium disks. Materials, 11.10.3390/ma11122484 ↗
37. Biofilm analysis of retrieved dental implants after different peri-implantitis treatments
38. Decontamination of the infected implant surface: A scanning electron microscope study10.11607/prd.4568 ↗
39. Lack of clinical benefit of implantoplasty to improve implant survival rate10.1177/0022034520944158 ↗
40. Resolution of peri-implantitis by means of implantoplasty as adjunct to surgical therapy: A retrospective study10.1002/JPER.21-0103 ↗
41. Fracture resistance after implantoplasty in three implant-abutment connection designs
42. Leitão-Almeida, B., Font, O.C., Correia, A., Mari, J.M., Figueiredo, R., and Castellón, E.V. (2020). Effect of crown to implant ratio and implantoplasty on the fracture resistance of narrow dental implants with marginal bone loss: An in vitro study. BMC Oral Health, 20.10.1186/s12903-020-01323-z ↗
43. Effect of bone loss on the fracture resistance of narrow dental implants after implantoplasty. An in vitro study
44. Reduced fracture load of dental implants after implantoplasty with different instrumentation sequences. An in vitro study10.1111/clr.13754 ↗
45. Implantoplasty: Carbide burs vs. diamond sonic tips. An in vitro study10.1111/clr.13702 ↗
46. Titanium alloys for dental implants: A review10.3390/prosthesis2020011 ↗
47. Implant-abutment connections: Influence of the design on the microgap and their fatigue and fracture behavior of dental implants10.1007/s10856-014-5211-7 ↗
48. Fatigue characteristics of five types of implant-abutment joint designs10.3365/met.mat.2008.04.133 ↗
49. Effect of dental implant diameter on fatigue performance. Part II: Failure analysis10.1111/j.1708-8208.2012.00476.x ↗
50. Effect of implantoplasty on the elastic limit of dental implants of different diameters10.1186/s40729-021-00363-6 ↗
51. Effect of crown-to-implant ratio and crown height space on marginal bone stress: A finite element analysis10.1186/s40729-021-00368-1 ↗
52. Evaluation of test protocol variables for dental implant fatigue research10.1016/j.dental.2009.07.003 ↗
53. Effect of dental implant diameter on fatigue performance. Part I: Mechanical behavior10.1111/j.1708-8208.2012.00477.x ↗
54. Shemtov-Yona, K., and Rittel, D. (2016). Fatigue of dental implants: Facts and fallacies. Dent. J., 4.10.3390/dj4020016 ↗
55. Identification of failure mechanisms in retrieved fractured dental implants10.1016/j.engfailanal.2014.01.002 ↗
56. Shah, S.D. (2019). Implant Strength after Implantoplasty. [Ph.D. Thesis, The Ohio State University].
57. Fractographic analysis of fractured dental implant components10.1016/j.jds.2012.09.006 ↗
58. Effect of material characteristics and/or surface topography on biofilm development10.1111/j.1600-0501.2006.01353.x ↗
59. The surface charge of anti-bacterial coatings alters motility and biofilm architecture10.1039/c3bm00197k ↗
60. The effect of surface charge property on Escherichia coli initial adhesion and subsequent biofilm formation10.1002/bit.24429 ↗
61. Schubert, A., Wassmann, T., Holtappels, M., Kurbad, O., Krohn, S., and Bürgers, R. (2019). Predictability of microbial adhesion to dental materials by roughness parameters. Coatings, 9.10.3390/coatings9070456 ↗
62. The influence of abutment surface roughness on plaque accumulation and peri-implant mucositis10.1034/j.1600-0501.1996.070302.x ↗
63. The influence of titanium abutment surface roughness on plaque accumulation and gingivitis: Short-term observations
64. Maintenance therapy in patients following the surgical treatment of peri-implantitis: A 5-year follow-up study10.1111/clr.12418 ↗
65. Lozano, P., Peña, M., Herrero-Climent, M., Rios-Santos, J.V., Rios-Carrasco, B., Brizuela, A., and Gil, J. (2022). Corrosion Behavior of Titanium Dental Implants with Implantoplasty. Materials, 15.10.3390/ma15041563 ↗
66. Commercial oral hygiene products and implant collar surfaces: Scanning electron microscopy observations