Mechanical Behavior of Five Different Morse Taper Implants and Abutments with Different Conical Internal Connections and Angles: An In Vitro Experimental Study

Claudia Caballero, Fernando Rodriguez, Guillermo Castro Cortellari, Antonio Scarano, Juan Carlos Prados-Frutos, Piedad N. De Aza, Gustavo Vicentis Oliveira Fernandes, Sergio Alexandre Gehrke

Abstract

The present study evaluated the mechanical behavior of five designs of Morse taper (MT) connections with and without the application of loads. For this, the detorque of the fixing screw and the traction force required to disconnect the abutment from the implant were assessed. A total of 100 sets of implants/abutments (IAs) with MT-type connections were used, comprising five groups (n = 20/group): (1) Group Imp 11.5: IA sets with a cone angulation of 11.5°; (2) Group SIN 11.5: with a cone angulation of 11.5°; (3) Group SIN 16: with a cone angulation of 16°; (4) Group Neo 16: with a cone angulation of 16°; and (5) Group Str 15: with a cone angulation of 15°. All sets received the torque recommended by the manufacturer. After applying the torque, the counter torque of the fixing screws was measured in ten IA sets of each group without the application of cyclic loads (frequencies ≤ 2 Hz, 360,000 cycles, and force at 150 Ncm). The other ten sets of each group were subjected to cyclic loads, after which the detorque was measured. Afterwards, the force for disconnection between the implant and the abutment was measured by traction on all the samples. The untwisting of the abutment fixation screws showed a decrease in relation to the initial torque applied in all groups. In the unloaded samples, it was found to be −25.7% in Group 1, −30.4% in Group 2, −36.8% in Group 3, −29.6% in Group 4, and −25.7% in Group 5. After the applied loads, it was found to be −44% in Group 1, −43.5% in Group 2, −48.5% in Group 3, −47.2% in Group 4, and −49.8% in Group 5. The values for the IA sets were zero for SIN 16 (Group 3) and Neo16 (Group 4), both without and with loads. In the other three groups, without loads, the disconnection value was 56.3 ± 2.21 N (Group 1), 30.7 ± 2.00 N (Group 2), and 26.0 ± 2.52 N (Group 5). After applying loads, the values were 63.5 ± 3.06 N for Group 1, 34.2 ± 2.45 N in Group 2, and 23.1 ± 1.29 N in Group 5. It was concluded that in terms of the mechanical behavior of the five designs of MT IA sets, with and without the application of loads, the Imp 11.5, SIN 11.5, and Srt 15 groups showed better results compared to the SIN 16 and Neo 16 groups, showing that lower values of cone angulation increase the friction between the parts (IA), thus avoiding the need to maintain the torque of the fixing screw to maintain the union of the sets.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Orthodontics and Dentofacial Orthopedics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Implantology, Bioface/Postgrados en Odontología/Universidad Catolica de Murcia, 11100 Montevideo, Uruguay

Department of Innovative Technologies in Medicine & Dentistry, University of Chieti-Pescara, 66013 Chieti, Italy

Department of Medical Specialties and Public Health, IDIBO Group (High Performance Group in Research and Development of Biomaterials in Dentistry), Rey Juan Carlos University, 28300 Madrid, Spain

Department of Bioengineering, Universidad Miguel Hernandez de Elche, 03202 Alicante, Spain

Missouri School of Dentistry & Oral Health, A. T. Still University, St. Louis, MO 63104, USA

Department of Biotechnology, Universidad Católica de Murcia (UCAM), 30107 Murcia, Spain

Funding

Ministry of Science and Innovation

References (44)

Show source-provided references
1. Sartoretto, S.C., Shibli, J.A., Javid, K., Cotrim, K., Canabarro, A., Louro, R.S., Lowenstein, A., Mourão, C.F., and Moraschini, V. (2023). Comparing the Long-Term Success Rates of Tooth Preservation and Dental Implants: A Critical Review. J. Funct. Biomater., 14.10.3390/jfb14030142 ↗
2. Trends in Dental Implant Use in the U.S., 1999–2016, and Projections to 202610.1177/0022034518792567 ↗
3. Potential risk factors for early and late dental implant failure: A retrospective clinical study on 9080 implants10.1186/s40729-020-00276-w ↗
4. The Etiology and Management of Dental Implant Failure: A Review
5. Occlusal overload with dental implants: A review10.1186/s40729-019-0180-8 ↗
6. Implications of considering peri-implant bone loss a disease, a narrative review10.1111/cid.13102 ↗
7. Complications in implant dentistry10.4103/ejd.ejd_340_16 ↗
8. Technical complications of implant-causes and management: A comprehensive review10.4103/0975-5950.168233 ↗
9. Cervino, G., Germanà, A., Fiorillo, L., D’amico, C., Abbate, F., and Cicciù, M. (2022). Passant Connection Screw of Dental Implants: An In Vitro SEM Preliminary Study. BioMed Res. Int., 2022.10.1155/2022/9720488 ↗
10. Investigating the micromorphological differences of the implant–abutment junction and their clinical implications: A pilot study
11. The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling10.11607/jomi.2727 ↗
12. Influence of abutment materials on the implant-abutment joint stability in internal conical connection type implant systems10.4047/jap.2014.6.6.491 ↗
13. Mechanism of and factors associated with the loosening of the implant abutment screw: A review10.1111/jerd.12494 ↗
14. Implant-abutment screw removal torque values between customized titanium abutment, straight titanium abutment, and hybrid zirconia abutment after a million cyclic loading: An in vitro comparative study10.1186/s40729-021-00378-z ↗
15. Misfit of Three Different Implant-Abutment Connections Before and After Cyclic Load Application: An In Vitro Study10.11607/jomi.5629 ↗
16. Performance of conical abutment (Morse Taper) connection implants: A systematic review10.1002/jbm.a.34709 ↗
17. Goyeneche, V.P., Cortellari, G.C., Rodriguez, F., De Aza, P.N., da Costa, E.M., Scarano, A., Júnior, N.D.B., and Gehrke, S.A. (2024). Does the index in Morse taper connection affect the abutment stability? An in vitro experimental study. PLoS ONE, 19.10.1371/journal.pone.0298462 ↗
18. Influence of abutment-to-fixture design on reliability and failure mode of all-ceramic crown systems10.1016/j.dental.2014.01.005 ↗
19. Implant–abutment interface design affects fatigue and fracture strength of implants10.1111/j.1600-0501.2008.01581.x ↗
20. Comparison of the Effect of Three Abutment-implant Connections on Stress Distribution at the Internal Surface of Dental Implants: A Finite Element Analysis
21. Changes in the abutment-implant interface in morse taper implant connections after mechanical cycling: A pilot study10.11607/jomi.3113 ↗
22. Load fatigue performance of conical implant-abutment connection: Effect of torque level and interface junction
23. Nie, H., Tang, Y., Yang, Y., Wu, W., Zhou, W., and Liu, Z. (2023). Influence of a new abutment design concept on the biomechanics of peri-implant bone, implant components, and microgap formation: A finite element analysis. BMC Oral Health, 23.10.1186/s12903-023-02989-x ↗
24. Behavior of implant and abutment sets of three different connections during the non-axial load application: An in vitro experimental study using a radiographic method10.3233/BME-211221 ↗
25. (2015). Dentistry—Implants—Dynamic Fatigue Test for Endosseous Dental Implants (Standard No. ISO 14801:2007).
26. Peri-implant tissue healing at implants with different designs and placement protocols: An experimental study in dogs10.1111/clr.13339 ↗
27. Implantat-Abutment-Verbund10.1055/s-0031-1280211 ↗
28. Laser microgrooved vs. machined healing abutment disconnection/reconnection: A comparative clinical, radiographical and biochemical study with split-mouth design10.1186/s40729-021-00301-6 ↗
29. Implant-to-bone force transmission: A pilot study for in vivo strain gauge measurement technique10.1016/j.jmbbm.2018.10.014 ↗
30. A systematic review of the influence of the implant-abutment connection on the clinical outcomes of ceramic and metal implant abutments supporting fixed implant reconstructions10.1111/clr.13362 ↗
31. Abutment screw loosening in angulation-correcting implants: An in vitro study10.1016/j.prosdent.2018.03.005 ↗
32. Effect of unilateral misfit on preload of retention screws of implant-supported prostheses submitted to mechanical cycling10.1016/j.jpor.2010.05.002 ↗
33. Torque required to loosen single-tooth implant abutment screws before and after simulated function
34. Effect of repeated torque/mechanical loading cycles on two different abutment types in implants with internal tapered connections: An in vitro study10.1111/j.1600-0501.2008.01690.x ↗
35. The Effect of Crown/Implant Ratio and Crown Height Space on Stress Distribution in Unsplinted Implant Supporting Restorations10.1016/j.joms.2011.01.036 ↗
36. An In Vitro analysis of implant screw torque loss with external hex and internal connection implant systems10.1097/01.id.0000245440.09464.48 ↗
37. Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading10.1186/s40729-018-0149-z ↗
38. The effect of tissue entrapment on screw loosening at the implant/abutment interface of external- and internal-connection implants: An In Vitro study10.1111/jopr.12329 ↗
39. Influence of diamondlike carbon-coated screws on the implant-abutment interface
40. Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading10.4047/jap.2018.10.6.415 ↗
41. Effect of cyclic loading and retightening on reverse torque value in external and internal implants10.4047/jap.2015.7.4.288 ↗
42. Bagegni, A., Weihrauch, V., Vach, K., and Kohal, R. (2022). The Mechanical Behavior of a Screwless Morse Taper Implant–Abutment Connection: An In Vitro Study. Materials, 15.10.3390/ma15093381 ↗
43. Mechanical resistance of screwless morse taper and screw-retained implant-abutment connections10.1111/clr.12303 ↗
44. Comparison of the effect of four different abutment screw torques on screw loosening in single implant-supported prosthesis after the application of mechanical loading10.1155/2021/3595064 ↗