Osseointegration of Tantalum Trabecular Metal in Titanium Dental Implants: Histological and Micro-CT Study
Source abstract
Abstract
This study aimed to investigate the impact of the Tantalum Trabecular Metal dental implant design on implant stability and the process of osseointegration following its placement in the rabbit femoral condyle. The subjects for the experiment consisted of 10 New Zealand white rabbits. Twenty implants, comprising 10 Trabecular Metal (TM) and 10 Traditional Screw Vent (TSV) implants, were placed into the femoral condyles of these rabbits. The implant type was alternated based on a random sequence. Following a healing period of 8 weeks, the implants were retrieved for further analysis using micro-computed tomography (micro-CT), histological studies, and histomorphometry evaluations. The Bone-to-Implant Contact (BIC) ratio and the Bone Volume (BV) percentage in the region of interest were subsequently assessed. The BIC and BV values between TM and TSV implants were compared using the Student t-test. The TM implants exhibited significantly greater BIC and BV scores. In particular, the BIC percentage was recorded as 57.9 ± 6.5 for the TM implants, as opposed to 47.6 ± 8 for the TSV implants. Correspondingly, the BV percentage was 57 ± 7.3 for the TM implants and 46.4 ± 7.4 for the TSV implants. The bone volume percentage measured using micro-CT evaluation was 89.1 ± 8.7 for the TM implants and 79.1 ± 8.6 for the TSV implants. Given the observed results, it is plausible to suggest that the bone growth surrounding the tantalum mesh could have improved the integration of the bone and facilitated its ingrowth into the TM implant.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials · Dental Implant Techniques and Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Prosthetic Dental Science, College of Dentistry, King Saud University, P.O. Box 60169, Riyadh 11545, Saudi Arabia
Department of Dentistry, Oral Health Institute, Hamad Medical Corporation, Doha P.O. Box 3050, Qatar
Funding
Deanship of Scientific Research, Dental Implant and Osseointegration Chair (DIORC), College of Dentistry, King Saud University, Riyadh, Saudi Arabia
References (54)
Show source-provided references
1. Survival rates of teeth, implants, and double crown-retained removable dental prostheses: A systematic literature review
2. Outcomes assessment of patients treated with osseointegrated dental implants at the university of washington graduate prosthodontic program, 1988 to 2000
3. What are the longevities of teeth and oral implants?10.1111/j.1600-0501.2007.01434.x ↗
4. Concepts for prevention of complications in implant therapy10.1111/prd.12278 ↗
5. Bone ingrowth and initial stability of titanium and porous tantalum dental implants: A pilot canine study10.1097/ID.0b013e31829b17b5 ↗
6. Bone-to-implant apposition with machined and mtx microtextured implant surfaces in human sinus grafts
7. Biomechanical and histomorphometric analysis of endosteal implants placed by using the osseodensification technique in animal models: A systematic review and meta-analysis10.1016/j.prosdent.2020.07.004 ↗
8. Okazaki, Y., and Katsuda, S.I. (2021). Biological safety evaluation and surface modification of biocompatible ti-15zr-4nb alloy. Materials, 14.10.3390/ma14040731 ↗
9. Comparing initial wound healing and osteogenesis of porous tantalum trabecular metal and titanium alloy materials10.1563/aaid-joi-D-17-00258 ↗
10. Tantalum coating of porous carbon scaffold supplemented with autologous bone marrow stromal stem cells for bone regeneration in vitro and in vivo10.1177/1535370216629578 ↗
11. Development and applications of porous tantalum trabecular metal-enhanced titanium dental implants10.1111/cid.12059 ↗
12. Comparative study of bone height changes around immediately loaded porous tantalum parallel sided trabecular & screw shaped implants retaining mandibular implant- overdentures using cbct (rct)10.21608/edj.2019.75975 ↗
13. Transcriptomic profiling of tantalum metal implant osseointegration in osteopenic patients10.1038/s41405-018-0004-6 ↗
14. Carraro, F., and Bagno, A. (2023). Tantalum as trabecular metal for endosseous implantable applications. Biomimetics, 8.10.3390/biomimetics8010049 ↗
15. The physicochemical/biological properties of porous tantalum and the potential surface modification techniques to improve its clinical application in dental implantology10.1016/j.msec.2015.01.007 ↗
16. Tantalum and its derivatives in orthopedic and dental implants: Osteogenesis and antibacterial properties10.1016/j.colsurfb.2021.112055 ↗
17. Immediate placement of a porous-tantalum, trabecular metal-enhanced titanium dental implant with demineralized bone matrix into a socket with deficient buccal bone: A clinical report10.1016/j.prosdent.2014.09.022 ↗
18. Prospective, multicenter evaluation of trabecular metal-enhanced titanium dental implants placed in routine dental practices: 1-year interim report from the development period (2010 to 2011)10.1111/cid.12232 ↗
19. van der Schoor, P., Schlee, M., and Wen, H.-B. (2022). Prospective pilot study of immediately provisionalized restorations of trabecular metal-enhanced titanium dental implants: A 5-year follow-up report. Appl. Sci., 12.10.3390/app12030942 ↗
20. New bone formation and trabecular bone microarchitecture of highly porous tantalum compared to titanium implant threads: A pilot canine study10.1111/clr.13074 ↗
21. Evaluation of machining methods for trabecular metal implants in a rabbit intramedullary osseointegration model10.1002/jbm.b.30627 ↗
22. Recommendations for the health monitoring of rodent and rabbit colonies in breeding and experimental units10.1258/0023677021911740 ↗
23. Percie du Sert, N., Ahluwalia, A., Alam, S., Avey, M.T., Baker, M., Browne, W.J., Clark, A., Cuthill, I.C., Dirnagl, U., and Emerson, M. (2020). Reporting animal research: Explanation and elaboration for the arrive guidelines 2.0. PLoS Biol., 18.10.1371/journal.pbio.3000411 ↗
24. The influence of implant geometry and surface composition on bone response10.1111/clr.12190 ↗
25. Micro-computed tomography assessment of human alveolar bone: Bone density and three-dimensional micro-architecture10.1111/cid.12109 ↗
26. Combination of topological parameters and bone volume fraction better predicts the mechanical properties of trabecular bone10.1016/S0021-9290(02)00060-X ↗
27. Kreve, S., Ferreira, I., da Costa Valente, M.L., and Dos Reis, A.C. (2022). Relationship between dental implant macro-design and osseointegration: A systematic review. Oral. Maxillofac. Surg., 1–14.10.1007/s10006-022-01116-4 ↗
28. Factors affecting late fixture loss and marginal bone loss around teeth and dental implants10.1111/j.1708-8208.2000.tb00118.x ↗
29. Failure load and stress analysis of orthodontic miniscrews with different transmucosal collar diameter10.1016/j.jmbbm.2018.07.032 ↗
30. The influence of implant design on bone remodeling around surface-modified southern implants(r)10.1111/j.1708-8208.2010.00308.x ↗
31. Osseointegration of implants: A biological and clinical overview
32. Dental implant stability and its measurements to improve osseointegration at the bone-implant interface: A review10.1016/j.matpr.2020.08.243 ↗
33. Experimental and clinical performance of porous tantalum in orthopedic surgery10.1016/j.biomaterials.2006.04.041 ↗
34. Effect of pore size on bone ingrowth into porous titanium implants fabricated by additive manufacturing: An in vivo experiment10.1016/j.msec.2015.10.069 ↗
35. Chang, Y.Y., Huang, H.L., Chen, Y.C., Hsu, J.T., Shieh, T.M., and Tsai, M.T. (2014). Biological characteristics of the mg-63 human osteosarcoma cells on composite tantalum carbide/amorphous carbon films. PLoS ONE, 9.10.1371/journal.pone.0095590 ↗
36. Ta-coated titanium surface with superior bacteriostasis and osseointegration10.2147/IJN.S218640 ↗
37. Novel hydroxyapatite/tantalum surface coating for metallic dental implant10.1016/j.matlet.2006.07.013 ↗
38. A trabecular metal implant 4 months after placement: Clinical-histologic case report10.1097/ID.0000000000000020 ↗
39. Enhanced bone apposition to a chemically modified sla titanium surface10.1177/154405910408300704 ↗
40. Implant surface roughness alterations induced by different prophylactic procedures: An in vitro study10.1111/clr.12849 ↗
41. The relationship of surface roughness and cell response of chemical surface modification of titanium10.1007/s10856-012-4611-9 ↗
42. A multi-center study comparing dual acid-etched and machined-surfaced implants in various bone qualities10.1902/jop.2001.72.10.1384 ↗
43. Osseointegration of titanium implants with different rough surfaces: A histologic and histomorphometric study in an adult minipig model10.1097/ID.0000000000000560 ↗
44. Statistical demonstration of the relative effect of surface chemistry and roughness on human osteoblast short-term adhesion10.1007/s10856-006-8475-8 ↗
45. Effect of surface structure on biomechanical properties and osseointegration10.1016/j.msec.2008.03.015 ↗
46. Kligman, S., Ren, Z., Chung, C.H., Perillo, M.A., Chang, Y.C., Koo, H., Zheng, Z., and Li, C. (2021). The impact of dental implant surface modifications on osseointegration and biofilm formation. J. Clin. Med., 10.10.3390/jcm10081641 ↗
47. Schulze, C., Weinmann, M., Schweigel, C., Kessler, O., and Bader, R. (2018). Mechanical properties of a newly additive manufactured implant material based on ti-42nb. Materials, 11.10.3390/ma11010124 ↗
48. A role for surface topography in creating and maintaining bone at titanium endosseous implants10.1067/mpr.2000.111966 ↗
49. Osseointegration of oral implants after delayed placement in rabbits: A microcomputed tomography and histomorphometric study10.11607/jomi.3133 ↗
50. Effect of intermittent loading and surface roughness on peri-implant bone formation in a bone chamber model10.1111/j.1600-051X.2007.01135.x ↗
51. Evaluation of bone regeneration using the rat critical size calvarial defect10.1038/nprot.2012.113 ↗
52. Animal models for evaluation of bone implants and devices: Comparative bone structure and common model uses10.1177/0300985815593124 ↗
53. Bone healing performance of electrophoretically deposited apatite-wollastonite/chitosan coating on titanium implants in rabbit tibiae10.1002/term.186 ↗
54. Animal models of impaired long bone healing and tissue engineering- and cell-based in vivo interventions10.1002/jor.25277 ↗