journal article·2022-12-23Open access

Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets

Luca Comuzzi, Margherita Tumedei, Tea Romasco, Morena Petrini, Kelvin I. Afrashtehfar, Francesco Inchingolo, Adriano Piattelli, Natalia Di Pietro

Abstract

Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values of ultrashort (3 mm length), short (7 mm length), and standard implants (10 mm length) inserted in 1-, 2-, 3-, and 4-mm thickness polyurethane sheets with densities of 10, 20, and 30 pounds per cubic foot (PCF). Standard-length implants were the gold standard (control). Overall, short-length implant IT values were higher or similar to the control in most experimental conditions. Those inserted into a 3 mm/30 PCF lamina showed the highest IT values, whereas 5 mm diameter ultrashort-length implants inserted into 2 and 3 mm/20 PCF laminas were higher than other implants. RT values followed the same trend and RFA values were more appreciable in short- and standard-length implants in all the scenarios. However, ultrashort-length implants reached a primary stability comparable to that of standard implants in lower thicknesses. In conclusion, although further studies are needed to corroborate this in vitro model with preclinical and clinical studies, our data shed light on short- and ultrashort-length implants geometries to a potential application in critical atrophy of the posterior jaws.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Orthopaedic implants and arthroplasty

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Independent Researcher, San Vendemiano-Conegliano Veneto, 31020 Treviso, Italy

Department of Medical, Surgical, and Dental Sciences, University of Milan, 20122 Milan, Italy

Department of Medical, Oral and Biotechnological Sciences, “G. d’Annunzio” University of Chieti-Pescara, 66013 Chieti, Italy

Center for Advanced Studies and Technology-CAST, “G. d’Annunzio” University of Chieti-Pescara, 66013 Chieti, Italy

Clinical Sciences Department, College of Dentistry, Ajman University, Ajman Emirate P.O. Box 346, United Arab Emirates

Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, 3010 Berne, Switzerland

Department of Interdisciplinary Medicine, University of Bari “Aldo Moro”, 70124 Bari, Italy

School of Dentistry, Saint Camillus International University of Health and Medical Sciences, 00131 Rome, Italy

References (48)

Show source-provided references
1. Survival rates of short dental implants (≤6 mm) compared with implants longer than 6 mm in posterior jaw areas: A meta-analysis10.1111/clr.13289 ↗
2. Prosthetic Rehabilitation of the Posterior Atrophic Maxilla, Short (≤6 mm) or Long (≥10 mm) Dental Implants? A Systematic Review, Meta-analysis, and Trial Sequential Analysis: Naples Consensus Report Working Group A
3. Short Implants (5 to 8 mm) Versus Longer Implants (>8 mm) with Sinus Lifting in Atrophic Posterior Maxilla: A Meta-Analysis of RCTs10.1111/cid.12432 ↗
4. Fixed and Removable Full-Arch Restorations Supported by Short (≤8-mm) Dental Implants in the Mandible: A Systematic Review and Meta-Analysis10.11607/jomi.7241 ↗
5. A Meta-Analysis Indicating Extra-Short Implants (≤6 Mm) as an Alternative to Longer Implants (≥8 Mm) with Bone Augmentation10.1038/s41598-021-87507-1 ↗
6. Short implants (5–8 mm) vs. long implants in augmented bone and their impact on peri-implant bone in maxilla and/or mandible: Systematic review10.1016/j.jormas.2018.11.007 ↗
7. Short implants (5–8 mm) vs long implants (≥10 mm) with augmentation in atrophic posterior jaws: A meta-analysis of randomised controlled trials10.1111/joor.12860 ↗
8. Comparing short dental implant and standard dental implant in terms of marginal bone level changes: A systematic review and meta-analysis of randomized controlled trials10.1111/cid.12774 ↗
9. Meta-analysis of randomized clinical trials comparing clinical and patient-reported outcomes between extra-short (≤6 mm) and longer (≥10 mm) implants10.1111/jcpe.13026 ↗
10. Short implants (≤6 mm) versus longer implants with sinus floor elevation in atrophic posterior maxilla: A systematic review and meta-analysis
11. Do Implant Length and Width Matter for Short Dental Implants (<10 Mm)? A Meta-Analysis of Prospective Studies10.1902/jop.2013.120745 ↗
12. Short implants (<8 mm) versus longer implants (≥8 mm) with lateral sinus floor augmentation in posterior atrophic maxilla: A meta-analysis of RCT’s in humans
13. Lombardo, G., Signoriello, A., Marincola, M., and Nocini, P.F. (2020). Assessment of Peri-Implant Soft Tissues Conditions around Short and Ultra-Short Implant-Supported Single Crowns: A 3-Year Retrospective Study on Periodontally Healthy Patients and Patients with a History of Periodontal Disease. Int. J. Environ. Res. Public Health, 17.10.3390/ijerph17249354 ↗
14. Minimally Invasive Fixed Rehabilitation of an Extremely Atrophic Posterior Mandible Using 4-Mm Ultrashort Implants: A Case Report with a 7-Year Follow-Up10.11607/prd.4957 ↗
15. Five-Year Results from a Randomised Controlled Trial Comparing Prostheses Supported by 5-Mm Long Implants or by Longer Implants in Augmented Bone in Posterior Atrophic Edentulous Jaws
16. Comuzzi, L., Iezzi, G., Piattelli, A., and Tumedei, M. (2019). An In Vitro Evaluation, on Polyurethane Foam Sheets, of the Insertion Torque (IT) Values, Pull-Out Torque Values, and Resonance Frequency Analysis (RFA) of NanoShort Dental Implants. Polymers, 11.10.3390/polym11061020 ↗
17. Rehabilitation of the Atrophic Posterior Mandible with Short (4-Mm) Implants: A Case Report10.11607/prd.1733 ↗
18. Influence of the dental implant macrogeometry and threads design on primary stability: An in vitro simulation on artificial bone blocks10.1080/10255842.2021.1875219 ↗
19. Evaluation of Ultrashort and Longer Implants with Microrough Surfaces: Results of a 24- to 36-Month Prospective Study10.11607/jomi.4648 ↗
20. Observational Study on the Preparation of the Implant Site with Piezosurgery vs. Drill: Comparison between the Two Methods in terms of Postoperative Pain, Surgical Times, and Operational Advantages10.1155/2019/8483658 ↗
21. Rehabilitation of a Severe Man-dibular Atrophy with Four Millimeter Extra-Short Implant and Guided Bone Regeneration (GBR): Case Report with 7-Years Follow-Up
22. Evidence-Based Clinical Outcomes of Immediate and Early Loading of Short Endosseous Dental Implants: A Meta-Analysis10.11607/jomi.8541 ↗
23. Preliminary Results of Fixed, Fiber-Reinforced Resin Bridges on Four 4- × 5-Mm Ultrashort Implants in Compromised Bony Sites: A Pilot Study10.1016/j.joms.2014.10.031 ↗
24. Pierfelice, T.V., D’Amico, E., Iezzi, G., Piattelli, A., Di Pietro, N., D’Arcangelo, C., Comuzzi, L., and Petrini, M. (2022). Nanoporous Titanium Enriched with Calcium and Phosphorus Promotes Human Oral Osteoblast Bioactivity. Int. J. Environ. Res. Public Health, 19.10.3390/ijerph19106212 ↗
25. Davies, J.E. (2000). Bone response to dual acid-etched and machined titanium implant surfaces. Bone Engineering, EM Squared Incorporated.
26. Platelet interactions with titanium: Modulation of platelet activity by surface topography10.1016/S0142-9612(01)00009-6 ↗
27. Removal Torque and Histomorphometric Investigation of 4 Different Titanium Surfaces: An Experimental Study in the Rabbit Tibia
28. Early endosseous integration enhanced by dual acid etching of titanium: A torque removal study in the rabbit10.1034/j.1600-0501.2001.012004350.x ↗
29. Bone Response to Two Dental Implants with Different Sandblasted/Acid-Etched Implant Surfaces: A Histological and Histomorphometrical Study in Rabbits10.1155/2017/8724951 ↗
30. Surface Modifications and Their Effects on Titanium Dental Implants10.1155/2015/791725 ↗
31. The insertion torque-depth curve integral as a measure of implant primary stability: An in vitro study on polyurethane foam blocks10.1016/j.prosdent.2017.04.012 ↗
32. Comparison of Osteotome and Conventional Drilling Techniques for Primary Implant Stability: An In Vitro Study10.1563/aaid-joi-D-15-00176 ↗
33. García-Braz, S.H., Prados-Privado, M., Zanatta, L.C.S., Calvo-Guirado, J.L., Prados-Frutos, J.C., and Gehrke, S.A. (2019). A Finite Element Analysis to Compare Stress Distribution on Extra-Short Implants with Two Different Internal Connections. J. Clin. Med., 8.10.3390/jcm8081103 ↗
34. Romanos, G.E., Bastardi, D.J., Moore, R., Kakar, A., Herin, Y., and Delgado-Ruiz, R.A. (2019). In Vitro Effect of Drilling Speed on the Primary Stability of Narrow Diameter Implants with Varying Thread Designs Placed in Different Qualities of Simulated Bone. Materials, 12.10.3390/ma12081350 ↗
35. Gehrke, S.A., Tumedei, M., Aramburú Júnior, J., Treichel, T.L.E., Kolerman, R., Lepore, S., Piattelli, A., and Iezzi, G. (2020). Histological and Histomorphometrical Evaluation of a New Implant Macrogeometry. A Sheep Study. Int. J. Environ. Res. Public Health, 17.10.3390/ijerph17103477 ↗
36. A Comparative Evaluation between Aluminium and Titanium Dioxide Microparticles for Blasting the Surface Titanium Dental Implants: An Experimental Study in Rabbits10.1111/clr.12973 ↗
37. Retrospective Study of Short and Extra-Short Implants Placed in Posterior Regions: Influence of Crown-to-Implant Ratio on Marginal Bone Loss10.1111/cid.12073 ↗
38. Implants for Elderly Patients10.1111/prd.12166 ↗
39. The Survival of Ultrashort Locking-Taper Implants
40. Long-Term Short Implants Performance: Systematic Review and Meta-Analysis of the Essential Assessment Parameters10.1590/0103-6440201300265 ↗
41. In vitro evaluation of the primary stability of short and conventional implants10.1563/aaid-joi-D-16-00094 ↗
42. An in vitro evaluation on polyurethane foam sheets of the insertion torque, removal torque values, and resonance frequency analysis (RFA) of a self-tapping threads and round apex implant10.1177/0262489320971796 ↗
43. Influence of the implant diameter and bone quality on the primary stability of porous tantalum trabecular metal dental implants: An in vitro biomechanical study10.1111/clr.12792 ↗
44. Does Bone Mineral Density Influence the Primary Stability of Dental Implants? A Systematic Review10.1111/j.1600-0501.2011.02228.x ↗
45. A Correlation between Bone (B), Insertion Torque (IT), and Implant Stability (S): BITS Score10.1016/j.prosdent.2014.02.011 ↗
46. Effects on Primary Stability of Three Different Techniques for Implant Site Preparation in Synthetic Bone Models of Different Densities10.1016/j.bjoms.2016.07.004 ↗
47. Bone density: A key determinant for clinical success
48. Single versus splinted short implants at sinus augmented sites: A systematic review and meta-analysis10.1016/j.jormas.2020.08.013 ↗