journal article·2026-06Open access

Flexible IMU Sensor Array For 3D Colonoscope Shape Reconstruction and AI‐Based Loop Detection

Tuukka Panula, Anni Halkilahti, Andrei Ivanov, Matti Kaisti

Advanced Science· Volume 13· Issue 34

Abstract

ABSTRACT Looping during colonoscopy procedures is a major cause of patient discomfort and increases the risk of complications. With the rising incidence of colorectal cancer, there is an urgent need for safer and more efficient colonoscopy techniques. We developed a proof‐of‐concept platform for 3D visualization of the colonoscope combined with artificial intelligence (AI)‐based loop detection. The system consists of an array of 15 inertial measurement units (IMUs) mounted on a flexible printed circuit board that can be retrofitted into the instrument channel of conventional colonoscopes. Using sensor fusion algorithms, the position and orientation of individual IMUs are estimated to reconstruct the 3D shape of the colonoscope in real time. The system successfully reconstructed the colonoscope shape in a silicone colon phantom, and the AI model achieved high loop‐detection performance with an area under the receiver operating characteristic curve of 0.95 on test data. These results demonstrate feasibility in an artificial environment, while further validation in clinical settings is required.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Colorectal Cancer Screening and Detection · Soft Robotics and Applications · Robotics and Sensor-Based Localization

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of computing University of Turku Turku Finland

References (28)

Show source-provided references
1. World Health Organization “Colorectal Cancer ”World Health Organization(2023). Available:https://www.who.int/news‐room/fact‐sheets/detail/colorectal‐cancer.
2. Colorectal Cancer Incidence Trends in Younger Versus Older Adults: An Analysis of Population‐Based Cancer Registry Data
3. 10.3748/wjg.v23.i20.363210.3748/wjg.v23.i20.3632 ↗
4. 10.1097/01.mcg.0000225595.01201.d210.1097/01.mcg.0000225595.01201.d2 ↗
5. 10.1155/2013/35376010.1155/2013/353760 ↗
6. 10.1111/j.1440-1746.2008.05739.x10.1111/j.1440-1746.2008.05739.x ↗
7. 10.1111/den.1346510.1111/den.13465 ↗
8. 10.2174/187518140130501000110.2174/1875181401305010001 ↗
9. 10.1109/TMECH.2025.354009910.1109/TMECH.2025.3540099 ↗
10. 10.4161/jig.2373810.4161/jig.23738 ↗
11. 10.1097/SLE.000000000000123610.1097/SLE.0000000000001236 ↗
12. 10.1109/TIM.2020.301560710.1109/TIM.2020.3015607 ↗
13. 10.1109/TIM.2023.324421910.1109/TIM.2023.3244219 ↗
14. Enhancing Human‐AI Collaboration: The Case of Colonoscopy
15. 10.1038/s42256-020-00231-910.1038/s42256-020-00231-9 ↗
16. M.Oda H. R.Roth T.Kitasaka et al. “Colonoscope Tracking Method Based on Shape Estimation Network ” inMedical Imaging 2019: Image‐Guided Procedures Robotic Interventions and Modeling10951(SPIE 2019) 183–188.10.1117/12.2512729 ↗
17. M.Oda H. R.Roth T.Kitasaka et al. “Colon Shape Estimation Method for Colonoscope Tracking Using Recurrent Neural Networks ” inInternational Conference on Medical Image Computing and Computer‐Assisted Intervention(Springer 2018) 176–184.10.1007/978-3-030-00937-3_21 ↗
18. Z.HuangandC.Hu “Real‐Time Attitude Tracking of Capsule Endoscope Based on MEMS IMU and Error Analysis ” in2021 IEEE International Conference on Real‐Time Computing and Robotics (RCAR)(IEEE 2021) 968–973.10.1109/RCAR52367.2021.9517523 ↗
19. 10.1109/TIM.2022.320410310.1109/TIM.2022.3204103 ↗
20. J.Lee H.Ishii andA.Takanishi “Shape Visualization Method of Flexible Colonoscopy Using Non‐Visual Sensor Network for Monitoring of Operation ” in2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society(IEEE 2011) 524–527.10.1109/IEMBS.2011.6090095 ↗
21. 10.1109/TMRB.2025.357305310.1109/TMRB.2025.3573053 ↗
22. R. E.Schapire “Explaining AdaBoost ” inEmpirical Inference: Festschrift in Honor of Vladimir N. Vapnik(Springer 2013) 37–52.10.1007/978-3-642-41136-6_5 ↗
23. O. J.Woodman “An Introduction to Inertial Navigation ”Technical Report UCAM‐CL‐TR‐696(University of Cambridge Computer Laboratory 2007).
24. M.Oda H.Kondo T.Kitasaka et al. “Tracking Accuracy Evaluation of Electromagnetic Sensor‐Based Colonoscope Tracking Method ” inInternational Workshop on Computer‐Assisted and Robotic Endoscopy(Springer 2015) 101–108.10.1007/978-3-319-29965-5_10 ↗
25. 10.1016/j.compmedimag.2013.01.01010.1016/j.compmedimag.2013.01.010 ↗
26. 10.1038/s41746-024-01412-110.1038/s41746-024-01412-1 ↗
27. P.Jackson R.Simon andC.Linte “Surgical Tracking Registration and Navigation Characterization for Image‐Guided Renal Interventions ” in2020 42nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)(IEEE 2020) 5081–5084.10.1109/EMBC44109.2020.9175270 ↗
28. 10.1186/s13014-021-01804-710.1186/s13014-021-01804-7 ↗