Pyroelectric composite film for X-ray intensity detection

Walter Katsumi Sakamoto, Giuliano Pierre Estevam, Aparecido Augusto de Carvalho, Wesley Pontes, Mauro Henrique de Paula

Materials Research· Volume 15· Issue 2

Abstract

Composite material obtained with modified lead titanate (Pz34) ferroelectric ceramic and polyether-ether-ketone (PEEK) polymer matrix was used as sensitive component to measure X-ray intensity in a novel detection system. The sensing element works as a thermal transducer, converting a non-quantified thermal flux into an output measurable quantity of electrical voltage. The samples were obtained up to 60 vol.% of ceramic, by hot pressing the mixture of Pz34 and PEEK powders at 368 °C and applying 12 MPa pressure for 2.0 hours. The sensor response varies from 2.70 to 0.80 V in the energy fluence rate range of 6.30 to 37.20 W.m-2. The absorbed incident energy was analyzed as a function of the ionizing energy. Furthermore, by measuring the pyroelectric activity of the composite film it was observed that there is no degradation of the sensor after the irradiation.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Analytical Chemistry and Sensors · Polymer Nanocomposite Synthesis and Irradiation · Electrical and Thermal Properties of Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Universidade Estadual Paulista, Brazil

Faculdade de Tecnologia, Brazil

Universidade Federal de Mato Grosso do Sul, Brazil

References (16)

Show source-provided references
1. Use of Ferroelectrics for Gamma-Ray Dosimetry10.1109/TNS2.1964.4315485 ↗
2. Use of lithium niobate detector for measuring X-ray intensity in mammographic range10.1049/el:20045729 ↗
3. Transparent pyroelectric sensors and organic field-effect transistors with fluorinated polymers: steps towards organic infrared detectors10.1109/TDEI.2006.247836 ↗
4. Suitability of some common polymer films for MeV proton beam dosimetry10.1016/j.nimb.2005.03.278 ↗
5. A prospective study should be performed to test the hypothesis that an increase in background radiation to residents in the gulf states will increase their longevity10.1118/1.1489045 ↗
6. Uncertainty in radiation dosimetry: basic concepts and methods10.1093/rpd/ncl094 ↗
7. Metcalfe P · 1997
8. Polyaniline/poly acid acrylic thin film composites: a new gamma radiation detector10.1016/S1044-5803(03)00090-1 ↗
9. Composite detector for mixed radiations based on CsI(Tl) and dispersions of small ZnSe(Te) crystals10.1016/j.nima.2004.12.030 ↗
10. Microcontrolled pyro-electric instrument for measuring X-ray intensity in mammography10.1007/BF02430953 ↗
11. 1980
12. Charrier JM · 1991 · 144
13. Espectroscopia dielétrica e corrente de despolarização termo-estimulada em filmes de peek
14. Pyroelectric coefficient direct measurement technique and application to a nsec response time detector10.1080/00150197208235326 ↗
15. Franco S · 2002
16. Measurement of X-ray intensity in the medical diagnostic range by a ferroelectric detector10.1109/58.656620 ↗