Determination of the activation energies of and aggregate rates for exothermic physico-chemical changes in UHMWPE by isothermal heat-conduction microcalorimetry (IHCMC)
Source abstract
Abstract
Exothermic heat flow rates (Q=microW=microJ/s), as a function of elapsed time, were measured by isothermal heat-conduction microcalorimetry (IHCMC) in order to study the aggregate rate of physico-chemical change in specimens of unsterilized and sterilized ultra-high-molecular-weight polyethylene (UHMWPE). Standard protocols for performing the IHCMC tests were developed and are described. Use of the standard protocols yielded the desired results-data that were not significantly different among either replicate sets of unsterilized specimens or as a function of which calorimeter test well was used. Heat flow rates measured in air at 20 degrees C, 25 degrees C, 35 degrees C, and 45 degrees C yielded estimates of activation energies of 47, 11, and 41 kJ/mol for unsterilized, gamma-radiation sterilized, and ethylene oxide gas (EtO) sterilized polymer, respectively. These results support the ideas that (a). initial exothermic degradation takes place much more easily in the radiation-sterilized material, due to direct oxidation of readily available free radicals, and (b). the much slower degradation process in EtO-sterilized UHMWPE is not appreciably different than in unsterilized polymer. Comparison with other activation energy data suggests that the rate-limiting process in EtO- or un-sterilized polymer is oxygen diffusion into the polymer. For shelf storage in air, for periods up to 8 months, the mean exothermic heat flow in air, at 25 degrees C (Q(m)) [determined from the Q values averaged over the time period between 15 and 20 h after test start], from UHMWPE gamma-radiation sterilized in air was significantly higher than for unsterilized material (2.91+/-0.11 vs. 0.73+/-0.11 microW). The higher rate can be attributed to oxidation of radiation-induced free radicals in the polymer near its surface. For the gamma-irradiated polymer, the decline in Q(m) with shelf storage time suggests that, eventually, degradation might become oxygen diffusion limited in this case also. However, in vivo, surface wear of an UHMWPE articular component may continue to expose unoxidized free radicals, keeping the exothermic reaction rate high and, possibly, continuing to produce an oxidized UHMWPE surface prone to wear.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Polymer crystallization and properties · Orthopaedic implants and arthroplasty · Recycling and Waste Management Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (16)
Show source-provided references
1. Current sterilization and packaging methods for polyethylene10.1097/00003086-199912000-00006 ↗
2. Effects of the sterilization method on the wear of UHMWPE acetabular cups tested in a hip joint simulator10.1016/S0142-9612(01)00265-4 ↗
3. The impact of sterilization method on wear in knee arthroplasty10.1097/00003086-199811000-00024 ↗
4. Effects of microtoming on oxidation of ultra high molecular weight polyethylene (UHMWPE)10.1016/S0141-3910(01)00080-5 ↗
5. Thermomechanical behavior of virgin and highly crosslinked ultra-high molecular weight polyethylene used in total joint replacements10.1016/S0142-9612(02)00102-3 ↗
6. Detection of peroxy species in ultra-high-molecular-weight polyethylene by Raman spectroscopy10.1016/S0142-9612(96)00149-4 ↗
7. Radiation sterilization and subsequent oxidation of medical grade polyethylene10.1016/S0168-583X(99)00083-X ↗
8. Isothermal microcalorimetry
9. Charlebois SJ. Stability and biocompatibility of implantable materials using isothermal calorimetry. PhD dissertation, Department of Mechanical Engineering, The University of Memphis, Memphis, TN, USA, 1999.
10. Isothermal microcalorimetry10.1016/S0142-9612(02)00317-4 ↗
11. Isothermal heat-conduction microcalorimetry (IHCMC) as a versatile technique for the evaluation of synthetic biomaterials10.1002/jbm.b.10044 ↗
12. Gamma sterilization of UHMWPE articular implants10.1016/0142-9612(95)00349-5 ↗
13. Solubility of gases in polyethylene10.1002/pol.1961.1205015411 ↗
14. Study of the behavior of UHMWPE systems10.1002/(SICI)1097-4636(199805)40:2<214::AID-JBM6>3.0.CO;2-O ↗
15. Fried · 1995
16. Properties of crosslinked ultra-high-molecular-weight polyethylene10.1016/S0142-9612(00)00195-2 ↗