Evaluation of the silver iontophoretic catheter in an animal model

Ray Y. Hachem, Kenneth C. Wright, Alfonso Zermeno, Gerald P. Bodey, Issam I. Raad

Biomaterials· Volume 24· Issue 20
Publisher page

Abstract

Silver iontophoretic catheters (SIC) were shown to be highly efficacious in preventing catheter infections in vitro and in a rabbit model (J. Infect. Dis. 173 (1996) 495). Furthermore, we sought to determine the safety and durability of SIC prior to use in humans. A total of 30 New Zealand white rabbits (3-4 kg) were randomly assigned to one of three groups whereby SIC or Arrow Guard (AG) catheters were tunneled and inserted in the jugular vein. All animals were followed for 2-12 weeks after catheter implantation. Blood was collected from each rabbit for assessment of toxicity and determination of silver levels. The electrical current generated by each SIC was measured once daily. At the end of the follow-up period, tissue samples were collected from the skin surrounding the catheter, the lungs, spleen, and liver. Microscopically, none of the tissues examined from any of the animals showed evidence of silver deposits, silver toxicity, thermal or electrical injury. The silver levels in the animals that received the SIC ranged from 0.1 to 2.23 microg/l with a mean of 0.62 (+/-0.44 SD). In conclusion SIC were safe with normal serum silver levels and were not associated with any local or systemic toxicity.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Central Venous Catheters and Hemodialysis · Electrical and Bioimpedance Tomography · Microfluidic and Bio-sensing Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (10)

Show source-provided references
1. In vitro antimicrobial efficacy of silver iontophoretic catheter10.1016/0142-9612(96)85905-9 ↗
2. Silver iontophoretic catheter10.1093/infdis/173.2.495 ↗
3. Prevention of central venous catheter-related bloodstream infection by use of an antiseptic-impregnated catheter. A randomized controlled trial10.7326/0003-4819-127-4-199708150-00001 ↗
4. A comparison of two antimicrobial-impregnated central venous catheters10.1056/NEJM199901073400101 ↗
5. Central venous catheters coated with minocycline and rifampin for the prevention of catheter-related colonization and bloodstream infections10.7326/0003-4819-127-4-199708150-00002 ↗
6. Antimicrobial durability and rare ultrastructural colonization of indwelling central catheters coated with minocycline and rifampin10.1097/00003246-199802000-00015 ↗
7. Mechanism of electrical enhancement of efficacy of antibiotics in killing biofilm bacteria10.1128/AAC.38.12.2803 ↗
8. Electrically generated silver ions10.1128/AAC.9.2.357 ↗
9. The effects of electric current on bacteria colonising intravenous catheters10.1016/0163-4453(93)92068-8 ↗
10. MICROMEDEX. Thomson Healthcare Series, vol. 112. 1974–2002.