Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions
Source abstract
Abstract
The polyethylene (PE) for hip implants presents serious clinical problems; the production of debris may induce adverse tissue reactions that may lead to extensive bone loss around the implant and consequently osteolysis and implant loosening. Several attempts have been made to improve the wear properties of ultra-high molecular weight polyethylene (UHMWPE). More recently the attention of various researchers has been focused on cross-linked polyethylene (XLPE), due to its improved wear resistance with respect to conventional UHMWPE. This study was aimed at comparing the wear performances of clinically available acetabular liners (Zimmer Inc.) made of electron beam XLPE and conventional UHMWPE. To evaluate the influence of the material properties on wear, conventional UHMWPE and XLPE acetabular cups were tested against deliberately scratched CoCrMo femoral heads (Ra = 0.12-0.14 microm) in a hip joint wear simulator run for 3 million cycles with bovine calf serum as lubricant. Gravimetric measurements revealed significant differences between the wear behaviours of the two sets of acetabular cups: XLPE exhibited a wear rate about 40 times lower than conventional UHMWPE. Raman spectroscopy coupled to partial least-squares analysis was used to evaluate the possible crystallinity changes induced by mechanical stress (and thus the material wear resistance): only the UHMWPE cup which showed the highest weight loss displayed significant crystallinity changes. These results were correlated to the thickness of the plasticity-induced damage layer. The wear debris produced during the tests were isolated according to a validated protocol and imaged by scanning electron microscopy . The wear particles produced by XLPE were smaller than those produced by UHMWPE; the latter were observed as fibrillar and agglomerated particles. The mean equivalent circle diameter was 0.71 and 0.26 microm for UHMWPE and XLPE, respectively.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (59)
Show source-provided references
1. Salute Md. Interrogazione per intervento principale. In: Salute Md, editor. www.ministerosalute.it; 2004.
2. Eggers EA. Hip replacement and current concepts. In: Orthopedic L, editor. http://www.louisvilleorthopedic.com/library/html/hip_replacement_and_current_concepts.html; 2004.
3. Property changes of UHMW polyethylene hip cup endoprostheses during implantation10.1002/jbm.820180915 ↗
4. Characterization of UHMWPE hip cups run on joint simulators10.1002/jbm.820210302 ↗
5. In vivo wear of polyethylene acetabular components10.1302/0301-620X.75B2.8444946 ↗
6. The origin of submicron polyethylene wear debris in total hip arthroplasty
7. Clinical review10.1002/jab.770040310 ↗
8. Howie DW. Tissue response in relation to type of wear particles around failed hip arthroplasties. J Arthroplasty 1990; 5(4): 337–48.10.1016/S0883-5403(08)80093-9 ↗
9. Wear of polyethylene in total joint replacement; observations from retrieved PCA knee implants10.1097/00003086-199203000-00016 ↗
10. An analysis of Hylamer and polyethylene bearings from retrieved acetabular components10.3928/0147-7447-19980801-09 ↗
11. Advances in the processing sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty10.1016/S0142-9612(99)00053-8 ↗
12. Low-friction arthroplasty of the hip using alumina ceramic and cross-linked polyethylene. A ten-year follow-up report10.1302/0301-620X.81B1.0810054 ↗
13. Oonishi H. Long term clinical results of THR: clinical results of THR of an alumina head with a cross-linked UHMWPE cup. Orthop Surg Traumatol 1995: 1255–63.
14. Jasty M, Bragdon CR, O’Connor DO. Marked improvement in the wear resistance of a new form of UHMWPE in a physiologic hip simulator. Transactions of the 43rd Orthopedic Research Society, vol.43. 1997.p. 785.
15. A novel method of crosslinking ultra-high molecular weight polyethylene to improve wear, reduce oxidation and retain mechanical properties10.1054/arth.2001.20540 ↗
16. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups10.1302/0301-620X.78B2.0780280 ↗
17. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements10.1002/jor.1100170203 ↗
18. Wear of gamma radiation crosslinked PE acetabular cups after aging and against roughened femoral heads
19. Wear of gamma cross-linked polyethylene acetabular cups against roughened femoral balls10.1097/00003086-199912000-00008 ↗
20. A comparison of the wear and physical properties of silane cross-linked polyethylene and ultra high molecular weight polyethylene10.1054/arth.2001.27234 ↗
21. Wear of conventional and cross-linked ultra-high-molecular-weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulator10.1002/jbm.10471 ↗
22. Effect of counterface roughness on the wear of conventional and crosslinked ultrahigh molecular weight polyethylene studied with a multidirectional motion pin-on-disk device10.1002/1097-4636(20011215)57:4<506::AID-JBM1196>3.0.CO;2-H ↗
23. Micro-Raman spectroscopy for the crystallinity characterization of UHMWPE hip cups run on joint simulators10.1016/S0022-2860(99)00427-5 ↗
24. The performance of gamma-and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 210.1016/S0142-9612(03)00264-3 ↗
25. Vibrational spectroscopy of ultra-high molecular weight polyethylene hip prostheses10.1016/S0022-2860(02)00141-2 ↗
26. Effects of the sterilisation method on the wear of the UHMWPE acetabular cups tested in a hip joint simulator10.1016/S0142-9612(01)00265-4 ↗
27. Isolation snd morphological characterisation of UHMWPE wear debris generated in vitro10.1016/S0142-9612(00)00403-8 ↗
28. An in vitro investigation of diamond-like carbon as a femoral head coating10.1002/(SICI)1097-4636(2000)53:3<221::AID-JBM6>3.0.CO;2-Z ↗
29. The effects of sterilisation methods on the wear of polyethylene in a hip simulator
30. A hip joint simulator study using new and physiologically scratched femoral heads with ultra-high molecular weight polyethylene acetabular cups10.1243/0954411001535598 ↗
31. A five-station hip joint simulator for wear rate studies10.1243/PIME_PROC_1992_206_291_02 ↗
32. A tribological study of UHMWPE acetabular cups and polyurethane compliant layer acetabular cups10.1002/1097-4636(2000)53:6<710::AID-JBM14>3.0.CO;2-R ↗
33. Saikko V, Ahlroos T, Calonius O, Keränen. Wear simulation of total hip prostheses with polyethylene against CoCr, alumina and diamond-like carbon. Biomaterials 2001; 22: 1507–14.10.1016/S0142-9612(00)00306-9 ↗
34. Measurement of the migration of a cemented hip prosthesis in an in vitro test10.1016/S0268-0033(00)00116-9 ↗
35. The effect of femoral head diameter upon lubrication and wear of metal–metal total hip replacements10.1243/0954411011533724 ↗
36. Partial least-squares regression10.1016/0003-2670(86)80028-9 ↗
37. Estimation of crystallinity in polyethylene by Raman spectroscopy10.1002/jrs.1250240813 ↗
38. Use of micro-Raman spectroscopy for the quantitative determination of polyethylene density using partial least-squares calibration10.1002/jrs.1250260606 ↗
39. Painter PC, Coleman MM, Koenig JL. The theory of vibrational spectroscopy and its application to polymeric materials. Wiley-Interscience, New York, 1982.
40. Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations10.2106/00004623-199703000-00005 ↗
41. Etiology of osteolysis around porous-coated cementless total hip arthroplasties10.1097/00003086-199411000-00018 ↗
42. Plasticity-induced damage layer is a precursor to wear in radiation-cross-linked UHMWPE acetabular components for total hip replacement10.1016/S0883-5403(99)90086-4 ↗
43. Rimnac · 2001 · vol. 49
44. Muratoglu OK, Bragdon CR, Jasty M. The effect of radiation damage on the wear rate of UHMWPE components. Symposium on characterization and properties of ultra-high molecular weight polyethylene, ASTM conference, New Orleans, 1995.
45. Effect of cross-linkage by gamma radiation in heavy doses to low-wear polyethylene in total hip prostheses
46. Irradiation of chemically crosslinked ultra high molecular weight polyethylene10.1002/(SICI)1099-0488(19960430)34:6<1063::AID-POLB4>3.0.CO;2-Z ↗
47. Bajaria SH, Bellare A. deformation, morphology, and wear behavior of polyethylene used in orthopedic implants. Med Plastics Biomater, 1998; March: 40–5.
48. The performance of gamma-and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 110.1016/S0142-9612(02)00238-7 ↗
49. Deformation, morphology, and wear behavior of polyethylene used in orthopedic implants.
50. Influence of packaging conditions on the properties of gamma-irradiated UHMWPE following accelerated ageing and shelf ageing10.1016/S0142-9612(98)00237-3 ↗
51. A study on solid-state drawn fibers of polyethylene by confocal Raman microspectrometry10.1016/S0032-3861(00)00913-7 ↗
52. Ultra-high molecular weight polyethylene wear debris generated in vivo and in laboratory yests; the influence of counterface roughness10.1243/PIME_PROC_1996_210_385_02 ↗
53. The influence of molecular weight, crosslinking and counterface roughness on TNF-alpha production by macrophages in response to ultra high molecular weight polyethylene particles10.1016/j.biomaterials.2003.10.054 ↗
54. Debris-mediated osteolysis—a cascade phenomenon involving motion, wear, particulates, macrophage induction, and bone lysis
55. Histopathological effects of ultra-high-molecular-weight polyethylene and metal wear debris in porous and cemented surface replacements
56. I Prodotti D’usura Nella Artroprotesi D’anca
57. Biological reactions to wear debris in total joint replacement10.1243/0954411001535219 ↗
58. Comparative wear and wear debris under three different counterface conditions of crosslinked and non-crosslinked ultra high molecular weight polyethylene
59. Do current wear particle isolation procedure underestimate the number of particles generated by prosthetic bearing components?10.1016/S0043-1648(01)00762-1 ↗