journal article·2017-03-01

A biomimetic mesh for treating female stress urinary incontinence

Kuilin Lai, Jiazhen Zhang, Guoshuai Wang, Xin Luo, Man Liu, Xinzhi Zhang, Yuyu Yuan, Bin Liu, Tao Xu

Biofabrication· Volume 9· Issue 1
Publisher page

Abstract

Abstract For patients with medium to severe incontinence, sub-urethral support surgery has a high cure rate, but using synthetic meshes leads to some complications such as mesh erosion/exposure and thigh pain. Autologous or acellular extracellular matrix grafts present few complications but have a high recurrence rate. Regensling ™ is a new sling product made of a synthetic material with a biomimetic structure, aiming to provide long-term mechanical support with a lower complication rate. To assess the safety and effectiveness of Regensling ™ , both in vitro and in vivo experiments were performed. The mesh was implanted in the subcutaneous, intramuscular and sub-urethral regions of rabbits. At 4, 12, and 26 weeks post-implantation, the animals were executed and the implants were studied for their mechanical and biocompatible properties. Compared to the control material, the Regensling ™ was covered by a thin layer of fibrous tissue with good compliance, and had a milder inflammatory response. During the period of the experiment, Regensling ™ showed stable strength with an increasing trend over time.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Pelvic floor disorders treatments · Hernia repair and management · Anorectal Disease Treatments and Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

South China University of Technology

Center for Medical Device Evaluation, China Food and Drug Administration, Beijing,

The First Affiliated Hospital of Jinan University

Tsinghua University

Medprin Biotech GmbH, Frankfurt,

University of Texas at El Paso

Funding

China Shenzhen Peacock Plan Project · Guangdong Innovative Research Team Program, China · The National High Technology Research and Development Program of China

References (15)

Show source-provided references
1. Tension-free vaginal tape for the treatment of urodynamic stress incontinence: efficacy and adverse effects at 10-year follow-up10.1016/j.eururo.2012.01.038 ↗
2. Thüroff · EAU Guidelines on Urinary Incontinence. Actas Urológicas Españolas (English Edition) · 2011 · 35 · 37310.1016/j.acuroe.2011.03.002 ↗
3. 2012 Update: guidelines for adult urinary incontinence collaborative consensus document for the Canadian urological association10.5489/cuaj.12248 ↗
4. How common are tape erosions? A comparison of two versions of the transobturator tension-free vaginal tape procedure10.1111/j.1464-410X.2006.06348.x ↗
5. Analysis of risk factors associated with vaginal erosion after synthetic sling procedures for stress urinary incontinence10.1007/s00192-007-0400-2 ↗
6. The collagenic architecture of human dura mater10.3171/2010.12.JNS101732 ↗
7. Extracellular matrix biomaterials for soft tissue repair10.1016/j.cpm.2009.08.001 ↗
8. Microarchitecture of three-dimensional scaffolds influences cell migration behavior via junction interactions10.1529/biophysj.107.122598 ↗
9. Use of modified 3D scaffolds to improve cell adhesion and drive desired cell responses10.3303/CET1227070 ↗
10. Autologous, cadaveric, and synthetic materials used in sling surgery: comparative biomechanical analysis10.1016/S0090-4295(01)01205-5 ↗
11. Osteogenic cells on bio-inspired materials for bone tissue engineering10.33549/physiolres.931776 ↗
12. On the relationship between common amplitude surface roughness parameters and surface area: Implications for the study of cell–material interactions10.1016/j.ibiod.2006.09.011 ↗
13. Molecular interactions in adherens-type contacts10.1242/jcs.1987.Supplement_8.14 ↗
14. Protein packing in adsorbed layers studied by excitation energy transfer10.1016/0021-9797(90)90111-Z ↗
15. Fibroblasts cell proliferation of charged hydroxyethyl methacrylate copolymers10.1016/0021-9797(85)90010-4 ↗