Edge loading does not increase wear rates of ceramic‐on‐ceramic and metal‐on‐polyethylene articulations
Source abstract
Abstract
Abstract The mal‐positioning of total hip arthroplasty components can result in edge loading conditions. Purpose of this study was to determine if the wear rate of ceramic‐on‐ceramic and metal‐on‐polyethylene increases under edge loading conditions. The literature was reviewed to determine which of the commonly used hip bearings is the most forgiving to implant mal‐orientation. Two 28‐mm ceramic‐on‐ceramic articulations were tested in vitro: pure alumina (PAL) ceramic versus the new alumina‐toughened zirconia (ATZ). Two 28‐mm metal‐on‐polyethylene articulations were tested in vitro: conventional ultrahigh molecular weight polyethylene (UHMWPE) versus highly crosslinked polyethylene (HXLPE) stabilized with vitamin E. All bearings were tested at standard and at highest possible inclination angles. Hip simulator tests were run for five million cycles based on N = 3 tests per condition. The average wear rate of ATZ‐on‐ATZ is 0.024 mm3/Mcycles at 45° and 0.018 mm3/Mcycles at 65°. Wear rate of PAL‐on‐PAL is between 0.02 and 0.03 mm3/Mcycles at 45°, as well as 65°. The wear rate of UHMWPE was 31 ± 1 mm3/Mcycles at an inclination angle of 45° and 26 ± 1 mm3/Mcycles at 80°. The wear rate of vitamin E stabilized HXLPE was 5.9 ± 0.2 mm3/Mcycles at 45° and 5.8 ± 0.2 mm3/Mcycles at 80°. Edge loading does not increase the wear rate of ceramic‐on‐ceramic and metal‐on‐polyethylene articulations. The newest biomaterials showed markedly lower wear rates compared with their conventional counterparts. ATZ‐on‐ATZ showed the lowest wear rate of all tested pairings, but the vitamin E stabilized HXLPE seems to be the most forgiving material when it comes to implant mal‐orientation. © 2014 Wiley Periodicals, Inc. J Biomed Mater Res Part B, 102B: 1627–1638, 2014.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Orthopedics Clinical Orthopedic Research Center (CORC‐mN), Diakonessenhuis Utrecht/Zeist, Bosboomstraat 1 3582 KE Utrecht The Netherlands
Complexo Hospitalario Universitario de Santiago de Compostela Spain
Innovation Group, Mathys Ltd Bettlach Bettlach Switzerland
Mathys Orthopädie GmbH Mörsdorf Germany
Funding
Mathys Orthopaedics, Bettlach Ltd, Switzerland
References (68)
Show source-provided references
1. An international comparison of the estimated effect of the aging of the population on the major cause of disablement, musculoskeletal disorders
2. 10.2106/JBJS.D.0244110.2106/JBJS.D.02441 ↗
3. 10.2106/00004623-200704000-0001210.2106/00004623-200704000-00012 ↗
4. Australian National Joint Replacement Registry Annual Report 2011. Available at:https://Aoanjrr.dmac.adelaide.edu.au/nl/annual‐reports‐2011. Accessed 2012 June.
5. 10.1016/j.medengphy.2011.05.01110.1016/j.medengphy.2011.05.011 ↗
6. 10.1302/0301-620X.79B1.684210.1302/0301-620X.79B1.6842 ↗
7. 10.2106/00004623-200100022-0000510.2106/00004623-200100022-00005 ↗
8. 10.1186/1475-925X-4-3210.1186/1475-925X-4-32 ↗
9. 10.1016/j.jbiomech.2011.10.03310.1016/j.jbiomech.2011.10.033 ↗
10. 10.1007/s11999-010-1632-x10.1007/s11999-010-1632-x ↗
11. 10.1302/0301-620X.90B9.1989110.1302/0301-620X.90B9.19891 ↗
12. 10.2106/JBJS.H.0048510.2106/JBJS.H.00485 ↗
13. 10.1302/0301-620X.90B7.2021310.1302/0301-620X.90B7.20213 ↗
14. 10.2106/JBJS.A.02039pp10.2106/JBJS.A.02039pp ↗
15. 10.1302/0301-620X.92B1.2277010.1302/0301-620X.92B1.22770 ↗
16. 10.2106/JBJS.H.0062110.2106/JBJS.H.00621 ↗
17. 10.1302/0301-620X.93B3.2554510.1302/0301-620X.93B3.25545 ↗
18. 10.1007/s11999-009-1187-x10.1007/s11999-009-1187-x ↗
19. 10.2106/JBJS.J.0187710.2106/JBJS.J.01877 ↗
20. 10.1302/0301-620X.93B8.2604010.1302/0301-620X.93B8.26040 ↗
21. 10.1302/0301-620X.93B8.2650510.1302/0301-620X.93B8.26505 ↗
22. 10.1016/j.arth.2011.04.03110.1016/j.arth.2011.04.031 ↗
23. 10.1002/jbm.b.3046210.1002/jbm.b.30462 ↗
24. 10.1097/00003086-200107000-0002010.1097/00003086-200107000-00020 ↗
25. Polyethylene wear and acetabular component orientation10.2106/00004623-200300004-00007 ↗
26. 10.1007/s11999-009-0845-310.1007/s11999-009-0845-3 ↗
27. 10.1016/S0883-5403(98)80003-X10.1016/S0883-5403(98)80003-X ↗
28. The effect of abduction differences between cup and contralateral acetabular angle on polyethylene component wear10.3928/0147-7447-20020101-18 ↗
29. 10.1302/0301-620X.87B11.1610110.1302/0301-620X.87B11.16101 ↗
30. 10.1302/0301-620X.94B3.2746010.1302/0301-620X.94B3.27460 ↗
31. 10.1016/j.biomaterials.2010.01.07610.1016/j.biomaterials.2010.01.076 ↗
32. 10.1007/s00264-010-1161-y10.1007/s00264-010-1161-y ↗
33. 10.1007/s11999-010-1717-610.1007/s11999-010-1717-6 ↗
34. 10.1243/09544119JEIM51810.1243/09544119JEIM518 ↗
35. Metal‐on‐metal bearings: The problem is edge‐loading wear
36. 10.1007/s11999-009-0830-x10.1007/s11999-009-0830-x ↗
37. 10.1023/A:100897002730610.1023/A:1008970027306 ↗
38. 10.1002/jbm.b.3139810.1002/jbm.b.31398 ↗
39. 10.1002/jbm.b.3170810.1002/jbm.b.31708 ↗
40. 10.1016/j.arth.2005.04.02610.1016/j.arth.2005.04.026 ↗
41. 10.1007/s11999-009-1154-610.1007/s11999-009-1154-6 ↗
42. 10.1023/A:101280230863610.1023/A:1012802308636 ↗
43. 10.1002/jbm.b.1003510.1002/jbm.b.10035 ↗
44. 10.1016/S0021-9290(00)00239-610.1016/S0021-9290(00)00239-6 ↗
45. 10.1243/0954411036057936710.1243/09544110360579367 ↗
46. 10.1007/s11999-012-2330-710.1007/s11999-012-2330-7 ↗
47. A novel method of cross‐linking ultra‐high‐molecular‐weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties
48. 10.1002/jbm.a.1010810.1002/jbm.a.10108 ↗
49. 10.1097/01.blo.0000238856.61862.7d10.1097/01.blo.0000238856.61862.7d ↗
50. Multi‐lab evaluation of hip simulator test conditions on relative wear rates of crosslinked polyethylenes
51. 10.1002/jor.110017020310.1002/jor.1100170203 ↗
52. 10.1016/j.arth.2005.07.00910.1016/j.arth.2005.07.009 ↗
53. 10.1002/jbm.a.3329710.1002/jbm.a.33297 ↗
54. 10.1016/j.biomaterials.2008.05.00410.1016/j.biomaterials.2008.05.004 ↗
55. 10.1002/jbm.a.3212210.1002/jbm.a.32122 ↗
56. 10.1007/978-3-642-27361-2_310.1007/978-3-642-27361-2_3 ↗
57. 10.1177/095441191247461310.1177/0954411912474613 ↗
58. 10.1054/arth.2000.810010.1054/arth.2000.8100 ↗
59. 10.1002/jbm.b.2003310.1002/jbm.b.20033 ↗
60. Comparison of the wear rates of twenty‐eight and thirty‐two‐millimeter femoral heads on cross‐linked polyethylene acetabular cups in a wear simulator10.2106/00004623-200312000-00009 ↗
61. 10.1016/j.arth.2008.02.00810.1016/j.arth.2008.02.008 ↗
62. 10.1002/jbm.b.3069110.1002/jbm.b.30691 ↗
63. 10.1002/jbm.b.1007710.1002/jbm.b.10077 ↗
64. 10.1080/00016470031739335810.1080/000164700317393358 ↗
65. Ceramic liner fracture after cementless alumina‐on‐alumina total hip arthroplasty10.1097/BLO.0b013e3180303e87 ↗
66. 10.2106/JBJS.E.0061810.2106/JBJS.E.00618 ↗
67. 10.2106/JBJS.H.0086710.2106/JBJS.H.00867 ↗
68. 10.1054/arth.2002.3366410.1054/arth.2002.33664 ↗