journal article·2014-01

A comparison of the stress corrosion cracking susceptibility of commercially pure titanium grade 4 in Ringer's solution and in distilled water: A fracture mechanics approach

Michael D. Roach, R. Scott Williamson, Joseph A. Thomas, Jason A. Griggs, Lyle D. Zardiackas

Publisher page

Abstract

Abstract From the results of laboratory investigations reported in the literature, it has been suggested that stress corrosion cracking (SCC) mechanisms may contribute to early failures in titanium alloys that have elevated oxygen concentrations. However, the susceptibility of titanium alloys to SCC in physiological environments remains unclear. In this study, a fracture mechanics approach was used to examine the SCC susceptibility of CP titanium grade 4 in Ringer's solution and distilled de‐ionized (DI) water, at 37°C. The study duration was 26 weeks, simulating the non‐union declaration of a plated fracture. Four wedge loads were used corresponding to 86–95% of the alloy's ligament yield load. The longest cracks were measured to be 0.18 mm and 0.10 mm in Ringer's solution and DI water, respectively. SEM analysis revealed no evidence of extensive fluting and quasi‐cleavage fracture features which, in literature reports, were attributed to SCC. We thus postulate that the Ringer's solution accelerated the wedge‐loaded crack growth without producing the critical stresses needed to change the fracture mechanism. Regression analysis of the crack length results led to a significant best‐fit relationship between crack growth velocity (independent variable) and test electrolyte, initial wedge load, and time of immersion of specimen in electrolyte (dependent variables). © 2013 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 102B: 73–79, 2014.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Hydrogen embrittlement and corrosion behaviors in metals

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Biomedical Materials Science, University of Mississippi Medical Center 2500 North State Street Jackson Mississippi 39216

References (36)

Show source-provided references
1. 10.5435/00124635-200910000-0000710.5435/00124635-200910000-00007 ↗
2. 10.1097/00005131-200409000-0000610.1097/00005131-200409000-00006 ↗
3. Zardiackas LD · 1995 · 123
4. Metallurgical evaluation of retrieved implants and correlation of failures to patient record data
5. 10.1520/STP37556S10.1520/STP37556S ↗
6. Pohler O · 1985 · 670
7. Pohler OEM · 1983 · 158
8. Boyer R · 1994 · 236
9. 10.1002/jbm.82017030710.1002/jbm.820170307 ↗
10. Relationship between composition, microstructure, and stress corrosion cracking (in salt solution) in titanium alloys
11. Desai VH · 1984 · 1
12. 10.1016/0010-938X(93)90102-M10.1016/0010-938X(93)90102-M ↗
13. Judy R · 1992 · 2073
14. Meyn DA · 1969 · 1253
15. Schutz R · 1992 · 265
16. Judy RW · 1988 · 1747
17. Scully JC · 1979 · 237
18. 10.1520/STP33243S10.1520/STP33243S ↗
19. Desai VH · 1984 · 125
20. Effects of the environment on materials
21. Williams DF · 1977 · 11
22. Williams DF · 1981 · 9
23. 10.1179/00070597579832078310.1179/000705975798320783 ↗
24. 10.1016/0010-938X(95)00054-N10.1016/0010-938X(95)00054-N ↗
25. Bundy KJ · 2006 · 853
26. ASTM F67 standard specification for unalloyed titanium for surgical implant applications 2006 ASTM International West Conshohocken PA
27. 10.1520/STP33421S10.1520/STP33421S ↗
28. Schutz R · 1998 · 669
29. 10.1179/00070597379832192810.1179/000705973798321928 ↗
30. Seagle SR · 1968 · 170
31. 10.1520/STP37557S10.1520/STP37557S ↗
32. 10.1520/STP37555S10.1520/STP37555S ↗
33. McEvily AJ · 1990 · 433
34. ASTM E1681 standard test method for determining threshold stress intensity factor for enviornment‐assisted cracking of metallic materials 2008 ASTM International West Conshohocken PA. 13
35. Definition and classification of fracture non‐unions
36. ASTM E647 standard test method for measurement of fatigue crack growth rates 2008 ASTM International West Conshohocken PA. 45