Validation of a micro‐CT technique for measuring volumetric wear in retrieved acetabular liners

A. E. Bowden, S. M. Kurtz, A. A. Edidin

Publisher page

Abstract

Abstract In this study, a novel micro‐CT–based technique for evaluating wear in retrieved acetabular liners was introduced and validated. Six UHMWPE acetabular components ranging in implantation time from 2.7 to 14.4 years were collected and evaluated with the use of a high‐resolution micro‐CT scanner. The components were scanned with a uniform volumetric resolution of 74 microns (16‐bit precision) with the use of a 1024 × 1024 in‐plane image matrix. Manual rigid 3D image registration of the interior hemispherical portion of the acetabular cup with geometric primitives by trained observers allowed for isolation, visualization, and measurement of the wear volume. Results for these six components indicated an average wear rate of 65 mm3/year. Overall scanner error was quantified gravimetrically and associated with a maximum uncertainty of 0.6%. Intra‐ and interobserver uncertainty analysis showed the method to be both accurate and repeatable. © 2005 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2005
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Hip disorders and treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (28)

Show source-provided references
1. 10.1097/00003086-197510000-0002110.1097/00003086-197510000-00021 ↗
2. Comparison of two and three‐dimensional computerized polyethylene wear analysis after total hip arthroplasty10.2106/00004623-200306000-00020 ↗
3. 10.1520/STP1445-EB10.1520/STP1445-EB ↗
4. 10.1054/arth.2001.2658910.1054/arth.2001.26589 ↗
5. A clinically practical method of manually assessing polyethylene liner thickness10.2106/00004623-200112000-00006 ↗
6. 10.1016/j.orthres.2003.12.00110.1016/j.orthres.2003.12.001 ↗
7. 10.1002/jbm.1047110.1002/jbm.10471 ↗
8. 10.1016/S0736-0266(02)00240-110.1016/S0736-0266(02)00240-1 ↗
9. 10.1016/S0142-9612(02)00238-710.1016/S0142-9612(02)00238-7 ↗
10. 10.1302/0301-620X.84B3.1134410.1302/0301-620X.84B3.11344 ↗
11. 10.1016/j.orthres.2003.09.00410.1016/j.orthres.2003.09.004 ↗
12. Sterilization and polyethylene wear: clinical studies to support laboratory data10.2106/00004623-200405000-00020 ↗
13. 10.2106/00004623-200406000-0002110.2106/00004623-200406000-00021 ↗
14. 10.1515/bmte.1999.44.9.24710.1515/bmte.1999.44.9.247 ↗
15. 10.1243/095441100153560610.1243/0954411001535606 ↗
16. 10.1007/s00264-003-0477-210.1007/s00264-003-0477-2 ↗
17. Comparison of the in vivo wear rates of 43 surgically retrieved direct compression molded and ram extruded ultra high molecular weight polyethylene acetabular components
18. 10.2106/00004623-199703000-0000510.2106/00004623-199703000-00005 ↗
19. 10.1002/jor.110017051110.1002/jor.1100170511 ↗
20. 10.1097/00003086-199908000-0001610.1097/00003086-199908000-00016 ↗
21. 10.1243/PIME_PROC_1995_209_350_0210.1243/PIME_PROC_1995_209_350_02 ↗
22. Evaluation of contemporary software methods used to quantify polyethylene wear after total hip arthroplasty10.2106/00004623-200312000-00020 ↗
23. Measurement of polyethylene wear in metal‐backed acetabular cups. II. Clinical application
24. 10.2106/00004623-200303000-0001710.2106/00004623-200303000-00017 ↗
25. 10.1243/095441199153508610.1243/0954411991535086 ↗
26. Wright TM · 2001 · 224
27. Short‐term in vivo wear of cross‐linked polyethylene10.2106/00004623-200404000-00012 ↗
28. 10.1016/S0883-5403(03)00341-310.1016/S0883-5403(03)00341-3 ↗