Tibial Baseplate Microstructure Governs High Cycle Fatigue Fracture In Vivo
Source abstract
Abstract
ABSTRACT Previous studies report rare occurrences of tibial baseplate fractures following primary total knee arthroplasty (TKA). However, at a microstructural scale, it remains unclear how fatigue models developed in vitro apply to fractures in vivo. In this study, we asked: (1) do any clinical factors differentiate fracture patients from a broader revision sample; and (2) in vivo, how does microstructure influence fatigue crack propagation? We identified three fractured tibial baseplates from an institutional review board exempt implant retrieval program. Then, for comparison, we collated clinical data from the same database for n = 2120 revision TKA patients with tibial trays. To identify mechanisms, we characterized fracture features using scanning electron and digital optical microscopy. Additionally, we performed cross sectional analysis using focused ion beam milling. The fracture cohort consisted of moderately to very active patients with increased implantation time (15.6 years) compared to the larger revision patient sample (5.1 years, p = 0.009). We did not find a significant difference in patient weight between the two groups (p = 0.98). Macroscopic fracture features aligned well with both previous retrieval analysis and in vitro baseplate fatigue tests. On a micron scale, we identified striations on each baseplate, demonstrating fatigue as the fracture mechanism. In vivo fatigue fracture processes depended on both the alloy (Ti‐6Al‐4V vs. CoCrMo) and the microstructure of the alloy formed during manufacturing. For Ti‐6Al‐4V, the presence of equiaxed or acicular microstructure influenced the fatigue crack propagation, the latter arising from large prior β grains and a Widmanstatten microstructure, degrading fatigue strength. CoCrMo alloy fatigue cracks propagated linearly, crystallographically influenced by planar slip. However, we did not document any features of overload or fast fracture, suggesting a high cycle, low stress fatigue regime. Ultimately, the crack profiles we present here may provide insight into fatigue fractures of modern tibial baseplates.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Titanium Alloys Microstructure and Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Implant Research Core School of Biomedical Engineering, Science, and Health Systems, Drexel University Philadelphia Pennsylvania USA
Department of Bioengineering Clemson University Clemson South Carolina USA
The Clemson University‐Medical University of South Carolina Bioengineering Program Charleston South Carolina USA
Department of Orthopaedic Surgery Hackensack University Medical Center Hackensack New Jersey USA
Tennessee Orthopedic Clinics Knoxville Tennessee USA
References (47)
Show source-provided references
1. 10.1155/2013/97070310.1155/2013/970703 ↗
2. 10.1001/2012.jama.1115310.1001/2012.jama.11153 ↗
3. 10.1016/j.arth.2013.07.02410.1016/j.arth.2013.07.024 ↗
4. 10.1016/j.ijscr.2013.11.01110.1016/j.ijscr.2013.11.011 ↗
5. 10.1016/j.artd.2022.02.00210.1016/j.artd.2022.02.002 ↗
6. 10.1155/2017/408081610.1155/2017/4080816 ↗
7. 10.2106/JBJS.CC.20.0010910.2106/JBJS.CC.20.00109 ↗
8. 10.1016/j.knee.2019.06.00910.1016/j.knee.2019.06.009 ↗
9. 10.1016/j.arth.2004.09.03310.1016/j.arth.2004.09.033 ↗
10. 10.3390/ma1507257510.3390/ma15072575 ↗
11. 10.1016/j.artd.2019.09.00210.1016/j.artd.2019.09.002 ↗
12. 10.1016/j.artd.2022.08.00210.1016/j.artd.2022.08.002 ↗
13. 10.1007/s00167-016-4160-810.1007/s00167-016-4160-8 ↗
14. 10.1098/rsif.2008.015110.1098/rsif.2008.0151 ↗
15. Fracture of the Metal Tibial Tray After Kinematic Total Knee Replacement. A Common Cause of Early Aseptic Failure
16. 10.1016/S0883-5403(99)90212-710.1016/S0883-5403(99)90212-7 ↗
17. 10.1016/j.arth.2004.02.04410.1016/j.arth.2004.02.044 ↗
18. 10.1016/S0021-9290(99)00094-910.1016/S0021-9290(99)00094-9 ↗
19. 10.1016/S0021-9290(03)00255-010.1016/S0021-9290(03)00255-0 ↗
20. Ahir S. P. · 2004 · 329
21. 10.1016/j.msea.2014.09.11510.1016/j.msea.2014.09.115 ↗
22. 10.1111/jmi.1254710.1111/jmi.12547 ↗
23. A.H. Committee “Fractography ASM International ”1987.
24. 10.1016/j.engfracmech.2020.10712910.1016/j.engfracmech.2020.107129 ↗
25. 10.1361/1547702052292410.1361/15477020522924 ↗
26. 10.1007/s11661-009-0064-210.1007/s11661-009-0064-2 ↗
27. 10.1007/s11661-002-0160-z10.1007/s11661-002-0160-z ↗
28. 10.1016/j.actamat.2019.07.02510.1016/j.actamat.2019.07.025 ↗
29. Kannan R. · 1982
30. 10.1016/j.prosdent.2022.05.01110.1016/j.prosdent.2022.05.011 ↗
31. Failed Femoral Stems in Total Hip Prostheses. A Report of Six Cases
32. 10.1007/s11668-022-01534-210.1007/s11668-022-01534-2 ↗
33. 10.1016/j.artd.2017.12.00510.1016/j.artd.2017.12.005 ↗
34. 10.1002/jor.2475010.1002/jor.24750 ↗
35. Long‐Term Survivorship of Hip and Knee Arthroplasty
36. 10.2106/JBJS.RVW.20.0002610.2106/JBJS.RVW.20.00026 ↗
37. Long‐Term Implant Survivorship of Cementless Total Knee Arthroplasty: A Systematic Review of the Literature and Meta‐Analysis
38. 10.1007/BF0057038710.1007/BF00570387 ↗
39. 10.1007/BF0266320310.1007/BF02663203 ↗
40. 10.1016/j.actamat.2015.05.05010.1016/j.actamat.2015.05.050 ↗
41. 10.1361/10599490277034431210.1361/105994902770344312 ↗
42. 10.1016/S0142-1123(99)00024-910.1016/S0142-1123(99)00024-9 ↗
43. 10.1016/S0921-5093(01)00982-010.1016/S0921-5093(01)00982-0 ↗
44. 10.1007/s40735-021-00623-310.1007/s40735-021-00623-3 ↗
45. 10.1016/j.jallcom.2012.07.02210.1016/j.jallcom.2012.07.022 ↗
46. 10.1186/s42836-022-00124-910.1186/s42836-022-00124-9 ↗
47. 10.1302/0301-620X.103B6.BJJ-2020-2096.R110.1302/0301-620X.103B6.BJJ-2020-2096.R1 ↗