Effects of machined/turned, TiO2‐blasted and sandblasted/acid‐etched titanium oral implant surfaces on nerve conduction: A study on isolated rat sciatic nerves
Source abstract
Abstract
Abstract The purpose of this study was to explore the clinical relevance of the effects of machined/turned, TiO2‐blasted and sandblasted/acid‐etched titanium oral implant surfaces on nerve conduction. Isolated rat sciatic nerves were placed between two suction electrodes in a pyrex bath containing a tyrode solution. Evoked compound action potentials (cAPs) of the nerves were recorded before and after contact with the implants. The mandibular incisors of randomly selected animals were extracted and changes in cAP amplitudes were used as controls. The differences in final cAP values of Astra Tech® implants and rat natural teeth were insignificant (P < 0.05), whereas the differences between other groups were significant (P < 0.05). Machined/turned‐surface implants did not cause any change in cAPs. A slight decrease in cAPs was observed for TiO2‐blasted and sandblasted/acid‐etched implants, and the natural teeth. The reductions of cAPs in latter groups were not 50% after an application time of 300 min. The cAP changes of nerves contacting TiO2‐blasted and sandblasted/acid‐etched oral implants fall within physiologic limits in vitro. Machined/turned, TiO2‐blasted, and sandblasted/acid‐etched titanium implant surfaces do not lead to irreversible neurotoxic effects. © 2003 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 67B: 772–778, 2003
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Dental Implant Techniques and Outcomes · Dental materials and restorations
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (40)
Show source-provided references
1. Producing and avoiding stress shielding10.1097/00003086-199201000-00010 ↗
2. 10.1002/jbm.82025070810.1002/jbm.820250708 ↗
3. 10.1002/jbm.82025040510.1002/jbm.820250405 ↗
4. 10.1002/(SICI)1097-4636(199704)35:1<19::AID-JBM3>3.0.CO;2-P10.1002/(SICI)1097-4636(199704)35:1<19::AID-JBM3>3.0.CO;2-P ↗
5. 10.1002/jbm.82022130710.1002/jbm.820221307 ↗
6. 10.1177/0022034500079011030110.1177/00220345000790110301 ↗
7. Jacobs R · 1998 · 75
8. Wang YH · 1998 · 3
9. Lambrichts I · 1998 · 13
10. 10.1034/j.1600-0501.2001.012003219.x10.1034/j.1600-0501.2001.012003219.x ↗
11. Intramedullary osseointegration: development of a rodent model and study of histology and neuropeptide changes around titanium implants
12. 10.1016/0306-4522(93)90588-710.1016/0306-4522(93)90588-7 ↗
13. 10.3109/1745367960899467610.3109/17453679608994676 ↗
14. 10.1080/0001647005294394610.1080/00016470052943946 ↗
15. 10.1016/0196-9781(88)90023-X10.1016/0196-9781(88)90023-X ↗
16. 10.1016/0165-1838(88)90016-110.1016/0165-1838(88)90016-1 ↗
17. 10.1007/BF0031903710.1007/BF00319037 ↗
18. Herskovits MS · 1990 · 165
19. 10.1016/S0901-5027(05)80562-410.1016/S0901-5027(05)80562-4 ↗
20. 10.1016/S1010-5182(97)80052-210.1016/S1010-5182(97)80052-2 ↗
21. 10.1016/S0278-2391(98)90407-410.1016/S0278-2391(98)90407-4 ↗
22. Endosseous implant placement in conjunction with inferior alveolar nerve transposition: An evaluation of neurosensory disturbance
23. 10.1034/j.1600-0501.1999.100206.x10.1034/j.1600-0501.1999.100206.x ↗
24. 10.1097/00006534-199707000-0000310.1097/00006534-199707000-00003 ↗
25. The use of somatosensory evoked potential during revision or reoperation for total hip arthoplasty10.1097/00003086-198906000-00020 ↗
26. Tissue regeneration adjacent to titanium implants placed with simultaneous transposition of the inferior dental nerve: A study in dogs
27. 10.1034/j.1600-0501.2003.140303.x10.1034/j.1600-0501.2003.140303.x ↗
28. 10.1111/j.1600-0765.1992.tb01668.x10.1111/j.1600-0765.1992.tb01668.x ↗
29. 10.1016/0006-8993(76)90716-210.1016/0006-8993(76)90716-2 ↗
30. 10.1097/00004770-200111000-0000610.1097/00004770-200111000-00006 ↗
31. 10.1097/00004770-200203000-0000110.1097/00004770-200203000-00001 ↗
32. 10.1016/S0099-2399(98)80116-710.1016/S0099-2399(98)80116-7 ↗
33. 10.1034/j.1600-0501.1992.030205.x10.1034/j.1600-0501.1992.030205.x ↗
34. Design and surface characteristics of 13 commercially available oral implant systems
35. 10.1113/jphysiol.1970.sp00929910.1113/jphysiol.1970.sp009299 ↗
36. 10.1007/978-1-4757-1758-710.1007/978-1-4757-1758-7 ↗
37. Osseointegration in skeletal reconstruction and rehabilitation
38. Surface analysis of four dental implant systems
39. 10.2106/00004623-196042010-0000710.2106/00004623-196042010-00007 ↗
40. 10.2106/00004623-196244020-0000810.2106/00004623-196244020-00008 ↗