journal article·2018-12

Comparison of catheter thrombogenicity in a modified chandler loop model using goat blood

Aaron L. Olsen, Michael Long

Publisher page

Abstract

Abstract Thrombogenicity testing is a key component in the development of medical devices intended for contact with blood. The Chandler loop system has previously been used as an in vitro thrombogenicity testing method. In this study, we used a modified version of the Chandler loop model to evaluate its capacity to detect differential thrombogenic potential of different catheter materials using goat blood. We also sought to determine the optimal experimental conditions for detecting the thrombogenicity of catheter material. Using the Chandler loop system with goat blood we demonstrated that silicone catheters had a stronger thrombogenicity as compared to polyurethane catheters as evidenced by significantly larger thrombi (p < 0.001) and higher infusion pressures (p < 0.05). This is consistent with many, but not all, previous in vitro and in vivo studies comparing polyurethane to silicone catheters. The use of this modified Chandler loop system with goat blood may provide an additional in vitro testing platform for thrombogenicity testing of catheters. © 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part A: 106A: 3143–3151, 2018.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Central Venous Catheters and Hemodialysis · Venous Thromboembolism Diagnosis and Management · Hemodynamic Monitoring and Therapy

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Animal, Dairy, and Veterinary Sciences Utah State University Logan Utah

References (42)

Show source-provided references
1. 10.1056/NEJMra01188310.1056/NEJMra011883 ↗
2. 10.1186/s13756-016-0161-010.1186/s13756-016-0161-0 ↗
3. ISO 10993–4:2017 (2017): ISO document. Biological evaluation of medical devices–Part 4: Selection of tests for interaction with blood International Organization of Standardization Geneva.
4. 10.1039/C4TB01839G10.1039/C4TB01839G ↗
5. 10.1002/jbm.82019040410.1002/jbm.820190404 ↗
6. 10.1002/jbm.b.3340310.1002/jbm.b.33403 ↗
7. 10.1038/sj.bjp.070754310.1038/sj.bjp.0707543 ↗
8. In vitro thrombotic coagulation of the blood: A method for producing a thrombus
9. 10.1007/s10856-015-5628-710.1007/s10856-015-5628-7 ↗
10. 10.1002/1097-4636(200102)54:2<224::AID-JBM9>3.0.CO;2-E10.1002/1097-4636(200102)54:2<224::AID-JBM9>3.0.CO;2-E ↗
11. 10.1002/jbm.a.3527510.1002/jbm.a.35275 ↗
12. 10.1177/0310057X800080021610.1177/0310057X8000800216 ↗
13. 10.1007/s10856-011-4335-210.1007/s10856-011-4335-2 ↗
14. 10.1080/0953710060074655710.1080/09537100600746557 ↗
15. 10.1007/s11239-013-0938-410.1007/s11239-013-0938-4 ↗
16. 10.1111/j.1538-7836.2007.02710.x10.1111/j.1538-7836.2007.02710.x ↗
17. 10.1161/hq0701.09368610.1161/hq0701.093686 ↗
18. 10.1093/ajcp/86.4.48410.1093/ajcp/86.4.484 ↗
19. 10.1371/journal.pone.017509410.1371/journal.pone.0175094 ↗
20. 10.5858/arpa.2016-0088-OA10.5858/arpa.2016-0088-OA ↗
21. 10.1097/00001721-200401000-0001710.1097/00001721-200401000-00017 ↗
22. 10.1177/002367721559713710.1177/0023677215597137 ↗
23. 10.1016/S1389-0344(02)00015-110.1016/S1389-0344(02)00015-1 ↗
24. 10.1177/014860718400800439910.1177/0148607184008004399 ↗
25. 10.1111/j.1540-8159.1994.tb01364.x10.1111/j.1540-8159.1994.tb01364.x ↗
26. 10.1002/jbm.82021010910.1002/jbm.820210109 ↗
27. Thrombogenicity of central venous catheters: Prospective study of polyethylene, silicone and polyurethane catheters with phlebography or post‐mortem examination
28. 10.1016/j.ejca.2016.02.01110.1016/j.ejca.2016.02.011 ↗
29. 10.3109/1073119800911899010.3109/10731198009118990 ↗
30. 10.4269/ajtmh.1981.30.43110.4269/ajtmh.1981.30.431 ↗
31. 10.1046/j.1538-7933.2003.00576.x10.1046/j.1538-7933.2003.00576.x ↗
32. 10.1016/j.thromres.2008.07.01310.1016/j.thromres.2008.07.013 ↗
33. 10.1016/j.ajog.2012.05.02710.1016/j.ajog.2012.05.027 ↗
34. 10.1007/s11239-015-1327-y10.1007/s11239-015-1327-y ↗
35. 10.5603/AIT.a2016.005110.5603/AIT.a2016.0051 ↗
36. 10.1055/s-0034-154399410.1055/s-0034-1543994 ↗
37. 10.1055/s-0029-121415510.1055/s-0029-1214155 ↗
38. 10.1016/0741-5214(86)90016-910.1016/0741-5214(86)90016-9 ↗
39. 10.1161/STROKEAHA.110.58921810.1161/STROKEAHA.110.589218 ↗
40. 10.1016/j.thromres.2013.08.00710.1016/j.thromres.2013.08.007 ↗
41. 10.1111/jce.1304910.1111/jce.13049 ↗
42. Effects of modest anemia on systemic and coronary circulation of septic sheep