journal article·2004-12-15

Comparison of plasma‐sprayed hydroxyapatite coatings and zirconia‐reinforced hydroxyapatite composite coatings: In vivo study

T. M. Lee, C. Y. Yang, E. Chang, R. S. Tsai

Publisher page

Abstract

Abstract The clinical use of plasma‐sprayed hydroxyapatite (HA) coatings on metal implants has been widely adopted because the HA coating can achieve the firmly and directly biological fixation with the surrounding bone tissue. However, the long‐term mechanical properties of HA coatings has been concern for the long‐term clinical application. Previous research showed that the concept of adding ZrO2 as second phase to HA significantly increased the bonding strength of plasma‐sprayed composite material. The present work aimed to explore the biological properties, including the histological responses and shear strength, between the plasma‐sprayed HA and HA/ZrO2 coating, using the transcortical implant model in the femora of canines. After 6 and 12 weeks of implantation, the HA coating revealed the direct bone‐to‐coating contact by the backscattered electron images (BEIs) of scanning electron microscope (SEM), but the osseointegration was not observed at the surface of HA/ZrO2 coating. For new bone healing index (NBHI) and apposition index (AI), the values for HA implants were significantly higher than that for HA/ZrO2 coatings throughout all implant periods. After push‐out test, the shear strength of HA‐coated implants were statistically higher than HA/ZrO2 coated implants at 6‐ and 12‐week implantation, and the failure mode of HA/ZrO2 coating was observed at the coating‐bone interface by SEM. The results indicate that the firm fixation between bone and HA/ZrO2 has not been achieved even after 12‐week implantation. Consequently, the addition of ZrO2 could improve the mechanical properties of coatings, while the biocompatibility was influenced by the different material characteristics of HA/ZrO2 coating compared to HA coatings. © 2004 Wiley Periodicals, Inc. J Biomed Mater Res 71A: 652–660, 2004
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Bone Tissue Engineering Materials · Dental Implant Techniques and Outcomes · Orthopaedic implants and arthroplasty

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (33)

Show source-provided references
1. 10.1016/0278-2391(87)90270-910.1016/0278-2391(87)90270-9 ↗
2. Bonding of bone to apatite‐coated implants10.1302/0301-620X.70B1.2828374 ↗
3. Hydroxyapatite‐coated titanium for orthopedic implant applications
4. 10.1002/jbm.82022030410.1002/jbm.820220304 ↗
5. 10.3109/1745367900899352110.3109/17453679008993521 ↗
6. 10.1002/jbm.82025070810.1002/jbm.820250708 ↗
7. 10.1002/jbm.82025080510.1002/jbm.820250805 ↗
8. Histomorphometric studies of hydroxylapatite‐coated and uncoated CP titanium threaded implants in bone
9. 10.1007/BF0012219810.1007/BF00122198 ↗
10. 10.1002/jbm.82027101210.1002/jbm.820271012 ↗
11. Hydroxyapatite‐coated total hip prostheses. Two‐year clinical and roentgenographic results of 100 cases10.1097/00003086-199012000-00007 ↗
12. 10.2106/00004623-199173100-0000110.2106/00004623-199173100-00001 ↗
13. 10.1002/jbm.82021120310.1002/jbm.820211203 ↗
14. 10.1002/jbm.82029010410.1002/jbm.820290104 ↗
15. 10.1016/0142-9612(94)90241-010.1016/0142-9612(94)90241-0 ↗
16. 10.1111/j.1151-2916.1986.tb04706.x10.1111/j.1151-2916.1986.tb04706.x ↗
17. 10.1007/BF0072011510.1007/BF00720115 ↗
18. 10.1179/mst.1991.7.6.49010.1179/mst.1991.7.6.490 ↗
19. 10.1023/A:101858352232210.1023/A:1018583522322 ↗
20. 10.1023/A:101853560639310.1023/A:1018535606393 ↗
21. 10.1023/A:101401090142310.1023/A:1014010901423 ↗
22. 10.1016/0254-0584(94)90071-X10.1016/0254-0584(94)90071-X ↗
23. 10.1002/(SICI)1097-4636(199707)36:1<39::AID-JBM5>3.0.CO;2-M10.1002/(SICI)1097-4636(199707)36:1<39::AID-JBM5>3.0.CO;2-M ↗
24. 10.1016/S0142-9612(98)00067-210.1016/S0142-9612(98)00067-2 ↗
25. 10.1023/A:101518302823210.1023/A:1015183028232 ↗
26. 10.1111/j.1151-2916.2002.tb00147.x10.1111/j.1151-2916.2002.tb00147.x ↗
27. Kokubo T · 1990 · 41
28. 10.1016/0142-9612(91)90194-F10.1016/0142-9612(91)90194-F ↗
29. 10.1002/jbm.82025030710.1002/jbm.820250307 ↗
30. 10.1002/jbm.82019030810.1002/jbm.820190308 ↗
31. 10.1002/jbm.82015030210.1002/jbm.820150302 ↗
32. 10.1002/jbm.82014050710.1002/jbm.820140507 ↗
33. 10.1007/BF0005896610.1007/BF00058966 ↗