Restoration of biomechanical and biological conditions for the finger implant to properly assess the suitability of the implant
Source abstract
Abstract
Abstract The aim of this work was to correlate biomechanical tests and simulations of implants dedicated to finger amputees. The implants were produced from Ti-6Al-4V (Ti64) alloy and ceramic ZrO2-Al2O3 (Alumina Toughened Zirconia 20, ATZ20). Research included finite element modeling, experimental simplified shear, tensile test, long-term cyclical tests of finger joint implant and their finite element models. Wear material collected during the experiment was evaluated for in vitro toxicity towards human fibroblast cells. Simulated fracture modes were verified by mechanical tests. During shear loading samples broke at ~ 55 N, whereas tensile break points were close to 110 N. Tested implant joints withstood one million cycle long-term test, however, partial wear of the anodized coating was noted. Wear debris from the joint was collected with a cellulose filter and analyzed for cytotoxic effects. Lactate dehydrogenase activity levels of cells incubated alongside collected wear material were only slightly higher than the control sample. Tested materials show promising potential for further research in terms of application.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopedic Surgery and Rehabilitation · Additive Manufacturing Materials and Processes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (42)
Show source-provided references
1. Bregoli C, Biffi CA, Morellato K, Gruppioni E, Primavera M, Rampoldi M, Lando M, Adani R, Tuissi A. Osseointegrated metallic implants for finger amputees: a review of the literature. Orthop Surg. 2022;14:1019–33. https://doi.org/10.1111/OS.13296.10.1111/OS.13296 ↗
2. Guerrero EM, Mastracci JC, Gart MS, Garcia RM, Loeffler BJ, Gaston RG. Soft tissue management of partial hand amputation. J Hand Surg Am. 2023;48:612–8. https://doi.org/10.1016/J.JHSA.2023.01.006.10.1016/J.JHSA.2023.01.006 ↗
3. Workshop-“Tecnologie bioniche e disabilità: le sfide della protesica d’arto” INAIL. 2019. https://www.inail.it/cs/internet/comunicazione/news-ed-eventi/eventi/evento-workshop-dcod-protesica-2019.html. Accessed 04 Jan 2024.
4. Reid DBC, Shah KN, Eltorai AEM, Got CC, Daniels AH. Epidemiology of Finger Amputations in the United States From 1997 to 2016. J Hand Surg Glob Online. 2019;1(2019):45–51. https://doi.org/10.1016/J.JHSG.2019.02.001.10.1016/J.JHSG.2019.02.001 ↗
5. Brenes C, Bencharit S, Fox T. Evaluation of prosthetic outcomes and patient satisfaction with 3D-printed implant-supported fixed prosthesis. Cureus. 2023. https://doi.org/10.7759/CUREUS.42537.10.7759/CUREUS.42537 ↗
6. Adani R, Woo SH. Microsurgical thumb repair and reconstruction. J Hand Surg Eur. 2017;42:771–88. https://doi.org/10.1177/1753193417723310.10.1177/1753193417723310 ↗
7. Zang CW, Zhang JL, Meng ZZ, Liu LF, Zhang WZ, Chen YX, Cong R. 3D printing technology in planning thumb reconstructions with second toe transplant. Orthop Surg. 2017;9:215–20. https://doi.org/10.1111/OS.12326.10.1111/OS.12326 ↗
8. Li Y, Kulbacka-Ortiz K, Caine-Winterberger K, Brånemark R. Thumb amputations treated with osseointegrated percutaneous prostheses with up to 25 years of follow-up. J Am Acad Orthop Surg Glob Res Rev. 2019. https://doi.org/10.5435/JAAOSGLOBAL-D-18-00097.10.5435/JAAOSGLOBAL-D-18-00097 ↗
9. Huang HF, Matschke J. Digit replantation under wide-awake local anesthesia. Plast Reconstr Surg. 2022;149:143–9. https://doi.org/10.1097/PRS.0000000000008609.10.1097/PRS.0000000000008609 ↗
10. Eberlin KR. Personalized treatment for finger amputations-what should we do with the thumb? JAMA Netw Open. 2020. https://doi.org/10.1001/JAMANETWORKOPEN.2019.21689.10.1001/JAMANETWORKOPEN.2019.21689 ↗
11. Wang H, Yang X, Huo Y, Hao R, Wang B, Wang W. Reverse homodigital dorsal wraparound flap for reconstruction of distal thumb. J Plast Surg Hand Surg. 2023;57:336–42. https://doi.org/10.1080/2000656X.2022.2088542.10.1080/2000656X.2022.2088542 ↗
12. Hoellwarth JS, Tetsworth K, Akhtar MA, Al Muderis M. Transcutaneous osseointegration for amputees: lessons from the past of relevance to the future. Bone Jt Res. 2021;10:690. https://doi.org/10.1302/2046-3758.1010.BJR-2021-0235.R2.10.1302/2046-3758.1010.BJR-2021-0235.R2 ↗
13. Forte AJ, Maita KC, Torres-Guzman RA, Avila FR, Safa B, Buncke G, Buntic R, Watt A. Great Toe Transplantation. Semin Plast Surg. 2022;36:243–52. https://doi.org/10.1055/S-0042-1758689.10.1055/S-0042-1758689 ↗
14. Naked Prosthetics | It’s All About Function | Prosthetic Fingers. 2023. https://www.npdevices.com/. Accessed 04 Jan 2024.
15. Shaterian A, Sayadi LR, Anderson A, Ng WKY, Evans GRD, Leis A. Characteristics of secondary procedures following digit and hand replantation. J Hand Microsurg. 2019;11:127. https://doi.org/10.1055/S-0039-1681981.10.1055/S-0039-1681981 ↗
16. Pamuk Ç. Supermicrosurgical experience pertaining to replantation of tamai zone 1 fingertip amputations. Handchir Mikrochir Plast Chir. 2023;55:28–34. https://doi.org/10.1055/A-1894-7320.10.1055/A-1894-7320 ↗
17. Lu YM, Te Lin Y, Tsai CH, Pan CH, Chen HH, Lee MC. Prognostic factors for attempted finger replantation and revascularisation after traumatic amputation: a 16-year retrospective cohort study. J Hand Surg Asian Pac. 2023;28:149–55. https://doi.org/10.1142/S242483552350025X.10.1142/S242483552350025X ↗
18. Stearns SA, Weidman AA, Engmann TF, Valentine L, Hernandez Alvarez A, Foppiani J, Lin SJ, Dowlatshahi S. Hand and wrist amputation: a demographic analysis using the national inpatient sample. Hand (N Y). 2023. https://doi.org/10.1177/15589447231167583.10.1177/15589447231167583 ↗
19. Byrski A, Kopernik M, Major Ł, Kasperkiewicz K, Dyner M, Lackner JM, Lumenta D, Major R. Characterization of biomaterials with reference to biocompatibility dedicated for patient-specific finger implants. Acta Bioeng Biomech. 2023;25:3–17. https://doi.org/10.37190/ABB-02156-2022-02.10.37190/ABB-02156-2022-02 ↗
20. Major R, Grajoszek A, Byrski A, Szawiraacz K, Barski JJ, Major Ł, Gawlikowski M, Kopernik M, Kot M, Dyner A, Lackner JM. Evaluation of in vivo biocompatibility in preclinical studies of a finger implant medical device correlated with mechanical properties and microstructure. ACS Appl Mater Interfaces. 2023. https://doi.org/10.1021/ACSAMI.3C16742.10.1021/ACSAMI.3C16742 ↗
21. J.A. Bannantine, J.J. Comer, J.L. Handrock, Fundamentals of metal fatigue analysis, Prentice Hall. 1990. https://www.worldcat.org/title/18836034. Accessed 07 June 7 2023.
22. Carrion PE, Shamsaei N, Daniewicz SR, Moser RD. Fatigue behavior of Ti-6Al-4V ELI including mean stress effects. Int J Fatigue. 2017;99:87–100. https://doi.org/10.1016/J.IJFATIGUE.2017.02.013.10.1016/J.IJFATIGUE.2017.02.013 ↗
23. Neuber H. Theory of stress concentration for shear-strained prismatical bodies with arbitrary nonlinear stress-strain law. J Appl Mech. 1961;28:544–50. https://doi.org/10.1115/1.3641780.10.1115/1.3641780 ↗
24. Faudot B, Milan JL, Goislard de Monsabert B, Le Corroller T, Vigouroux L. Estimation of joint contact pressure in the index finger using a hybrid finite element musculoskeletal approach. Comput Methods Biomech Biomed Engin. 2020;23:1225–35. https://doi.org/10.1080/10255842.2020.1793965.10.1080/10255842.2020.1793965 ↗
25. Butz KD, Merrell G, Nauman EA. A biomechanical analysis of finger joint forces and stresses developed during common daily activities. Comput Methods Biomech Biomed Engin. 2012;15:131–40. https://doi.org/10.1080/10255842.2010.517525.10.1080/10255842.2010.517525 ↗
26. Chagnon G, Achouri K, Connesson N, Gerbelot J, Courvoisier A. Development and operation of an experimental system to measure the moments generated in the finger joints. Bioengineering. 2022;2022(9):184. https://doi.org/10.3390/BIOENGINEERING9050184.10.3390/BIOENGINEERING9050184 ↗
27. Joyce TJ, Unsworth A. NeuFlex metacarpophalangeal prostheses tested in vitro. Proc Inst Mech Eng H. 2005;219:105–10. https://doi.org/10.1243/095441105X9192.10.1243/095441105X9192 ↗
28. Joyce TJ, Unsworth A. A test procedure for artificial finger joints. Proc Inst Mech Eng H. 2002;216:105–10. https://doi.org/10.1243/0954411021536324.10.1243/0954411021536324 ↗
29. T. Pylios, A New Metacarpophalangeal Joint Replacement Prosthesis, 2010.
30. ISO 10993–5:2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity. 2009. https://www.iso.org/standard/36406.html. Accessed 04 Jan 2024.
31. Kopernik M, Trȩbacz L, Pietrzyk M. Modelling of fatigue behaviour of hard multilayer nanocoating system in nanoimpact test. Comput Methods Appl Sci. 2008;9:137–59. https://doi.org/10.1007/978-1-4020-6975-8_8.10.1007/978-1-4020-6975-8_8 ↗
32. Vidhya MS, Merlyn-Christina KV. Fatigue life, fatigue damage, fatigue factor of safety, fatigue sensitivity, biaxiality indication and equivalent stress of a radial connecting rod. Int Res J Eng Technol. 2020;7:1499–502.
33. Fass M, Hecht M, Baumgartner J, Bauer N. Evaluation of the approach based on the maximum principal stress from the IIW-Recommendation for welded joints under proportional, multiaxial stress states. Weld World. 2023;67:2323–32. https://doi.org/10.1007/S40194-023-01571-X.10.1007/S40194-023-01571-X ↗
34. Carrion PE, Sausto F, Beretta S, Shamsaei N. Multiaxial fatigue behavior and modelling of additive manufactured Ti-6Al-4V parts: the effects of layer orientation and surface texture. Int J Fatigue. 2023;176: 107860. https://doi.org/10.1016/J.IJFATIGUE.2023.107860.10.1016/J.IJFATIGUE.2023.107860 ↗
35. Jones R, Michopoulos JG, Iliopoulos AP, Singh Raman RK, Phan N, Nguyen T. Representing crack growth in additively manufactured Ti-6Al-4V. Int J Fatigue. 2018;116:610–22. https://doi.org/10.1016/J.IJFATIGUE.2018.07.019.10.1016/J.IJFATIGUE.2018.07.019 ↗
36. Simonelli M, Tse YY, Tuck C. Effect of the build orientation on the mechanical properties and fracture modes of SLM Ti–6Al–4V. Mater Sci Eng, A. 2014;616:1–11. https://doi.org/10.1016/J.MSEA.2014.07.086.10.1016/J.MSEA.2014.07.086 ↗
37. Nicoletto G. Anisotropic high cycle fatigue behavior of Ti–6Al–4V obtained by powder bed laser fusion. Int J Fatigue. 2017;94:255–62. https://doi.org/10.1016/J.IJFATIGUE.2016.04.032.10.1016/J.IJFATIGUE.2016.04.032 ↗
38. Barricelli L, Patriarca L, du Plessis A, Beretta S. Orientation-dependent fatigue assessment of Ti6Al4V manufactured by L-PBF: size of surface features and shielding effect. Int J Fatigue. 2023;168: 107401. https://doi.org/10.1016/J.IJFATIGUE.2022.107401.10.1016/J.IJFATIGUE.2022.107401 ↗
39. ISO 16428:2005 - Implants for surgery—Test solutions and environmental conditions for static and dynamic corrosion tests on implantable materials and medical devices. 2005. https://www.iso.org/standard/30280.html. Accessed 04 Jan 2024.
40. Crisco JJ, Wolfe SW. Mechanical testing of orthopedic implants: Hand and wrist. In: Mechanical testing of orthopaedic implants. Elsevier; 2017. p. 63–98. https://doi.org/10.1016/B978-0-08-100286-5.00005-6.10.1016/B978-0-08-100286-5.00005-6 ↗
41. Guo M, Qi B, Li J, Shi X, Ni H, Shi H, Ren J, Zhou X, Ye T, Yao L, Xu Y, Zhang M, Li C. Mechanical properties evaluation of metacarpophalangeal joint prosthesis with new titanium-nickel memory alloy: a cadaver study. BMC Musculoskelet Disord. 2023;24:1–10. https://doi.org/10.1186/S12891-023-06859-Z/TABLES/4.10.1186/S12891-023-06859-Z/TABLES/4 ↗
42. Du J, Ding Y, Su X, He L, Hao W, Wang T. Surface modification and biocompatibility of medical nickel-titanium shape memory alloys. Chin J Tissue Eng Res. 2012;16:4686. https://doi.org/10.3969/J.ISSN.1673-8225.2012.25.026.10.3969/J.ISSN.1673-8225.2012.25.026 ↗