journal article·2023-03

In vitro and in vivo assessment of a novel ultra-flexible ventriculoamniotic shunt for treating fetal hydrocephalus

Stephen P Emery, Stephanie Greene, Moataz Elsisy, Kaitlin Chung, Sang-Ho Ye, Seungil Kim, William R Wagner, Nika Hazen, Youngjae Chun

Publisher page

Abstract

Fetal aqueductal stenosis (AS) is one of the most common causes of congenital hydrocephalus, which increases intracranial pressure due to partial or complete obstruction of cerebrospinal fluid (CSF) flow within the ventricular system. Approximately 2–4 infants per 10,000 births develop AS, which leads to progressive hydrocephalus, which enlarges the head often necessitating delivery by cesarean section. Most babies born with AS are severely neurologically impaired and experience a lifetime of disability. Therefore, a new device technology for venticuloamniotic shunting is urgently needed and has been studied to ameliorate or prevent fetal hydrocephalus development, which can provide a significant impact on patients and their family’s quality of life and on the decrease of the healthcare dollars spent for the treatment. This study has successfully validated the design of shunt devices and demonstrated the mechanical performance and valve functions. A functional prototype shunt has been fabricated and subsequently used in multiple in vitro tests to demonstrate the performance of this newly developed ventriculoamniotic shunt. The shunt contains a main silicone-nitinol composite tube, a superelastic 90° angled dual dumbbell anchor, and an ePTFE valve encased by a stainless-steel cage. The anchor will change its diameter from 1.15 mm (collapsed state) to 2.75 mm (deployed state) showing up to 1.4-fold diameter change in human body temperature. Flow rates in shunts were quantified to demonstrate the valve function in low flow rates mimicking the fetal hydrocephalus condition showing “no backflow” for the valved shunt while there is up to 15 mL/h flow through the shunt with pressure difference of 20 Pa. In vivo ovine study results show the initial successful device delivery and flow drainage with sheep model.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Cerebrospinal fluid and hydrocephalus · Fetal and Pediatric Neurological Disorders · Spinal Dysraphism and Malformations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Obstetrics, Gynecology & Reproductive Sciences, Divisions of Maternal-Fetal Medicine, Magee Womens Hospital of UPMC, Pittsburgh, PA, USA

Department of Neurological Surgery, Division of Neurosurgery, Children’s Hospital of Pittsburgh of UPMC, Pittsburgh, PA, USA

Mechanical Design and Production Department, Cairo University, Giza, Egypt

Department of Biengineering, University of Pittsburgh, Pittsburgh, PA, USA

Department of Surgery, UPMC, Pittsburgh, PA, USA

Center for Preclinical Studies, University of Pittsburgh McGowan Institute for Regenerative Medicine, Pittsburgh, PA, USA

Department of Industrial Engineering, University of Pittsburgh, Pittsburgh, PA, USA

McGowan Institute for Regenerative Medicine, Pittsburgh, PA, USA

Funding

Pitt Seed Project · UPMC Clinical Translational Science Institute PinCh Award · The Beckwith Institute · University of Pittsburgh Center for Medical Innovation · The 25 Club of Pittsburgh · RK Mellon Foundation · The Pediatric Device Initiative Seed Funding · Hearst Foundations

References (48)

Show source-provided references
1. 10.1016/j.siny.2012.06.00110.1016/j.siny.2012.06.001 ↗
2. 10.1159/00038201510.1159/000382015 ↗
3. 10.3171/2021.2.PEDS2082410.3171/2021.2.PEDS20824 ↗
4. 10.1016/j.wneu.2018.03.12310.1016/j.wneu.2018.03.123 ↗
5. 10.1159/00009566810.1159/000095668 ↗
6. 10.1002/pd.552710.1002/pd.5527 ↗
7. 10.1016/j.siny.2012.06.00110.1016/j.siny.2012.06.001 ↗
8. 10.1016/j.pediatrneurol.2011.03.00910.1016/j.pediatrneurol.2011.03.009 ↗
9. 10.1007/s00381-003-0775-410.1007/s00381-003-0775-4 ↗
10. 10.1007/BF0141847010.1007/BF01418470 ↗
11. 10.1056/NEJM19820603306220210.1056/NEJM198206033062202 ↗
12. 10.1056/NEJM19860731315053210.1056/NEJM198607313150532 ↗
13. 10.1080/17434440.2019.164140410.1080/17434440.2019.1641404 ↗
14. 10.7863/jum.2002.21.11.125710.7863/jum.2002.21.11.1257 ↗
15. 10.1016/j.ajog.2011.08.01910.1016/j.ajog.2011.08.019 ↗
16. 10.3109/03091902.2016.115461710.3109/03091902.2016.1154617 ↗
17. 10.1016/S0921-5093(99)00294-410.1016/S0921-5093(99)00294-4 ↗
18. 10.1097/00006123-199804000-0011310.1097/00006123-199804000-00113 ↗
19. Colas A · Biomat Sci Introduc Mater Med · 2004 · 2 · 80
20. Herdman R · 2000
21. 10.1002/pi.127310.1002/pi.1273 ↗
22. 10.1002/jbm.82013040810.1002/jbm.820130408 ↗
23. 10.1002/jbm.b.3492810.1002/jbm.b.34928 ↗
24. 10.1002/jbm.b.3418610.1002/jbm.b.34186 ↗
25. 10.1159/00002865910.1159/000028659 ↗
26. 10.1007/s00381-019-04096-110.1007/s00381-019-04096-1 ↗
27. Marc Oria SD · J Neurosurg · 2019 · 24 · 652
28. 27FDA. De Novo Classification Request for Medtronic Neurosurgery DUET External Drainage and Monitoring System (EDMS), U.S. Food & Drug, 2012, 1–13.
29. 10.3171/jns.1987.66.2.021310.3171/jns.1987.66.2.0213 ↗
30. 10.1007/s00381-003-0772-710.1007/s00381-003-0772-7 ↗
31. 10.1007/s00381-020-04738-910.1007/s00381-020-04738-9 ↗
32. 10.1159/00049736010.1159/000497360 ↗
33. 10.1109/ICBBE.2008.75610.1109/ICBBE.2008.756 ↗
34. 10.1080/10255842.2012.68342710.1080/10255842.2012.683427 ↗
35. 10.1016/j.jmbbm.2012.02.00610.1016/j.jmbbm.2012.02.006 ↗
36. McClean DR · Rev Cardiovasc Med · 2003 · 3 · S16
37. 10.3174/ajnr.A324610.3174/ajnr.A3246 ↗
38. 10.1016/j.jocn.2013.07.00610.1016/j.jocn.2013.07.006 ↗
39. 10.1002/adma.20180545210.1002/adma.201805452 ↗
40. 10.1007/BF0154086910.1007/BF01540869 ↗
41. 40Grunerbel L, Leikam B., Schrag G. Experimental Characterization Method for Passive Microvalves in Diaphragm Pumps for Medical Applications. Germany: MikroSystemTechnik Kongress, 2021, 204–207.
42. 10.1016/S0924-4247(03)00363-710.1016/S0924-4247(03)00363-7 ↗
43. 10.1109/JMEMS.2011.214815810.1109/JMEMS.2011.2148158 ↗
44. 10.1016/j.msea.2003.10.32610.1016/j.msea.2003.10.326 ↗
45. 10.1007/s40242-021-1177-410.1007/s40242-021-1177-4 ↗
46. 10.1016/0142-9612(93)90183-310.1016/0142-9612(93)90183-3 ↗
47. 10.1039/D1BM00233C10.1039/D1BM00233C ↗
48. 10.1159/00026355510.1159/000263555 ↗