A Comparison of Polyethylene and Polyurethane Blocks on the Stability of Dental Implants: An In Vitro Study

İbrahim Doğru, Levent Ciğerim

Applied Sciences· Volume 16· Issue 9

Abstract

The long-term success of dental implants is significantly influenced by primary stability, which is commonly assessed through insertion torque (IT) and removal torque (RT) measurements in vitro. While polyurethane (PU) blocks are accepted by the American Society for Testing and Materials (ASTM) as the standard bone analog material for biomechanical testing, the use of polyethylene (PE) as a bone model material for dental implant research remains limited and not well established. This operator-blinded, in vitro study compared the IT and RT values of dental implants placed in PE and PU blocks of identical density (60 pounds per cubic foot [pcf]; 0.96 g/cm3). A total of 60 tapered dental implants (4.2 × 12 mm, RBM surface, platform switching) were placed into PE (n = 30) and PU (n = 30) blocks by a calibrated operator blinded to the material type. Implant sockets were prepared by an independent surgeon following the manufacturer’s drilling protocol. IT and RT values were recorded using a physiodispenser with torque measurement capability (5–80 N·cm). Statistical analysis was performed using Student’s t-test (α = 0.05), with Mann–Whitney U tests reported as a sensitivity analysis for non-normally distributed variables. No statistically significant difference was observed in IT between PE and PU groups (58.50 ± 8.42 vs. 58.17 ± 9.60 N·cm; p = 0.887; Cohen’s d = 0.04; 95% CI of mean difference: −4.33 to 5.00 N·cm). However, RT was significantly higher in the PU group compared to the PE group (71.17 ± 7.15 vs. 64.33 ± 9.17 N·cm; p = 0.002; Cohen’s d = 0.83; 95% CI: −11.08 to −2.58 N·cm; Mann–Whitney U sensitivity analysis p = 0.004). Under the specific high-density (60 pcf) conditions tested, the absence of a statistically significant IT difference does not constitute formal evidence of equivalence or non-inferiority, and the significantly higher RT in PU indicates that PE and PU are not interchangeable bone analogs. Further studies across a range of densities, implant macrogeometries, and using formal equivalence testing are required before PE can be considered for in vitro dental implant stability research.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Facial Trauma and Fracture Management · Dental Radiography and Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Van Yüzüncü Yıl University, 65090 Van, Turkey

References (31)

Show source-provided references
1. The Future of Dental Implants: A Narrative Review of Trends, Technologies, and Patient Considerations
2. Dental Implants—Are They Better than Natural Teeth?10.1111/eos.12543 ↗
3. Bone Stability around Dental Implants: Treatment Related Factors10.1016/j.aanat.2015.02.004 ↗
4. Factors Influencing Implant Stability10.4103/jdi.jdi_14_18 ↗
5. Correlation between Implant Geometry, Implant Surface, Insertion Torque, and Primary Stability: In Vitro Biomechanical Analysis10.11607/jomi.6285 ↗
6. Model Selection for Determination of Primary Stability of Dental Implants in Experimental Studies10.17116/stomat202410306253 ↗
7. Rosas-Díaz, J., Guerrero, M.E., Córdova-Limaylla, N., Galindo-Gómez, M., García-Luna, M., and Cayo-Rojas, C. (2024). The Influence of the Degree of Dental Implant Insertion Compression on Primary Stability Measured by Resonance Frequency and Progressive Insertion Torque: In Vitro Study. Biomedicines, 12.10.3390/biomedicines12122878 ↗
8. Dura Haddad, C., Andreatti, L., Zelezetsky, I., Porrelli, D., Turco, G., Bevilacqua, L., and Maglione, M. (2024). Primary Stability of Implants Inserted into Polyurethane Blocks: Micro-CT and Analysis In Vitro. Bioengineering, 11.10.3390/bioengineering11040383 ↗
9. Evaluation of a Vibration Modeling Technique for the In-Vitro Measurement of Dental Implant Stability10.1016/j.jmbbm.2024.106537 ↗
10. Evaluation of the Insertion Torque, Implant Stability Quotient and Drilled Hole Quality for Different Drill Design: An in Vitro Investigation10.1111/clr.12808 ↗
11. Orhan, Z.D., and Ciğerim, L. (2024). Evaluation of Effect of Different Insertion Speeds and Torques on Implant Placement Condition and Removal Torque in Polyurethane Dense D1 Bone Model. Polymers, 16.10.20944/preprints202404.0942.v1 ↗
12. Santos, M., Mariz, M., Tiago, I., Alarico, S., and Ferreira, P. (2025). Bio-Based Polyurethane Foams: Feedstocks, Synthesis, and Applications. Biomolecules, 15.10.3390/biom15050680 ↗
13. Zerdzicki, K., Znaczko, A., Kondrusik, A., and Korbut, W. (2024). Compressive and Tensile Properties of Polyurethane Foam Mimicking Trabecular Tissue in Artificial Femoral Head Bones. Front. Bioeng. Biotechnol., 12.10.3389/fbioe.2024.1482165 ↗
14. Soft Material Drilling: A Thermo-Mechanical Analysis of Polyurethane Foam for Biomimetic Bone Scaffolds and Optimization of Process Parameters Using Taguchi Method10.1016/j.heliyon.2024.e37465 ↗
15. Influence of Implant Length and Insertion Depth on Primary Stability of Short Dental Implants: An In Vitro Study of a Novel Mandibular Artificial Bone Model10.1016/j.jds.2023.05.019 ↗
16. Patient-Specific Mechanical Analysis of Pedicle Screw Insertion in Simulated Osteoporotic Spinal Bone Models Derived from Medical Images10.31616/asj.2024.0121 ↗
17. Synthesis and Deconstruction of Polyethylene-Type Materials10.1021/acs.chemrev.3c00587 ↗
18. Hussain, M., Naqvi, R.A., Abbas, N., Khan, S.M., Nawaz, S., Hussain, A., Zahra, N., and Khalid, M.W. (2020). Ultra-High-Molecular-Weight-Polyethylene (UHMWPE) as a Promising Polymer Material for Biomedical Applications: A Concise Review. Polymers, 12.10.3390/polym12020323 ↗
19. UHMWPE for Arthroplasty: Past or Future?
20. Implementing Machine Learning Approaches for Accelerated Prediction of Bone Strain in Acetabulum of a Hip Joint10.1016/j.jmbbm.2024.106495 ↗
21. Effect of Matrix Material Property on the Composite Tibia Fracture Plate: A Biomechanical Study10.1088/2057-1976/ad38c4 ↗
22. Effects of Dental Implant Diameter and Tapered Body Design on Stress Distribution in Rigid Polyurethane Foam During Insertion10.1016/j.medengphy.2024.104181 ↗
23. Comparison of Epoxy Resin and Rigid Polyurethane Foam as a Bone Substitute to Study Implant Biomechanics
24. Measurement of Bone Deformation and Insertion Torque During Dental Implant Installation10.1007/s10006-024-01294-3 ↗
25. Custom-Made Polyurethane-Based Synthetic Bones Mimic Screw Cut-Through of Intramedullary Nails in Human Long Bones10.1016/j.jmbbm.2021.104405 ↗
26. Comuzzi, L., Tumedei, M., D’Arcangelo, C., Piattelli, A., and Iezzi, G. (2021). An In Vitro Analysis on Polyurethane Foam Blocks of the Insertion Torque (IT) Values, Removal Torque Values (RTVs), and Resonance Frequency Analysis (RFA) Values in Tapered and Cylindrical Implants. Int. J. Environ. Res. Public Health, 18.10.3390/ijerph18179238 ↗
27. Comparison of the Primary Stabilities of Conical and Cylindrical Endosseous Dental Implants: An In-Vitro Study
28. Prediction of Primary Stability via the Force Feedback of an Autonomous Dental Implant Robot10.1016/j.prosdent.2023.11.008 ↗
29. An Analytical Model to Measure Dental Implant Stability with the Advanced System for Implant Stability Testing (ASIST)10.1016/j.jmbbm.2023.106238 ↗
30. Bone Density and Proximal Support Effects on Dental Implant Stability—Finite Element Analysis and In Vitro Experiments10.1016/j.jormas.2023.101512 ↗
31. Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant10.11607/jomi.9769 ↗