In Vitro Micro-CT Assessment of a Novel Implant–Abutment Connection Under Static and Cyclic Loading

Marco Tallarico, Dario Melodia, Lukasz Zadrozny, Carlotta Cacciò, Silvio Mario Meloni, Aurea Immacolata Lumbau, Santo Catapano, Riccardo Baldari, Rafał Molak, Jakub Jaroszewicz, Gabriele Cervino

Applied Sciences· Volume 16· Issue 5

Abstract

The implant–abutment connection (IAC) is a critical determinant of the mechanical and biological performance of dental implants. Connection design and insertion torque may influence fatigue resistance, micromovement, and microgap formation, thereby affecting long-term implant success. This in vitro study evaluated a novel conical implant–abutment connection under controlled mechanical loading conditions. Methods: A sequential in vitro protocol was applied. Mechanical testing was conducted according to ISO 14801:2016 and included static and cyclic loading tests of the KS implant system inserted at two different torque values (35 Ncm and 70 Ncm). High-resolution micro-computed tomography (micro-CT) was performed after mechanical loading to evaluate implant–abutment interface integrity, microstructural alterations, and microgap behavior. Results: Static and cyclic loading tests revealed no observable differences between implants inserted at 35 Ncm and 70 Ncm, with all specimens completing the loading protocols without mechanical failure. Micro-CT analysis showed no evidence of microfractures, permanent deformation, or clinically relevant alterations at the implant–abutment interface. A stable and well-sealed connection was observed for both torque values following mechanical loading. Conclusions: Within the limitations of this in vitro study, the investigated conical implant–abutment connection demonstrated stable mechanical performance and preserved interface integrity after static and cyclic loading, regardless of whether implants were placed at 35 Ncm or 70 Ncm. These findings indicate that, under the present experimental conditions, both torque levels were associated with comparable structural integrity and mechanical stability of the investigated implant–abutment connection. This study should be interpreted as a preliminary experimental investigation, designed to provide descriptive and mechanistic insights rather than statistically powered comparative conclusions. Further long-term clinical trials are required to confirm these preliminary results.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Periodontal Regeneration and Treatments · Bone Tissue Engineering Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Medicine, Surgery, and Pharmacy, University of Sassari, 07100 Sassari, Italy

Department of Dental Propaedeutics and Prophylaxis, Faculty of Dental Medicine, Medical University of Warsaw, 59 Nowogrodzka Street, 02-006 Warsaw, Poland

Independent Researcher, 07100 Sassari, Italy

Department of Prosthodontics, University of Ferrara, Via Luigi Borsari 46, 44121 Ferrara, Italy

Independent Researcher, Rhein83, 40128 Bologna, Italy

Faculty of Materials Science and Engineering, Warsaw University of Technology, Woloska 141, 02-507 Warsaw, Poland

Department of Biomedical and Dental Sciences and Morphofunctional Imaging, University of Messina, Via Consolare Valeria, 1, 98125 Messina, Italy

Funding

OSSTEM

References (17)

Show source-provided references
1. Long-term evaluation of ANKYLOS® dental implants, part i: 20-year life table analysis of a longitudinal study of more than 12,500 implants10.1111/cid.12154 ↗
2. Timing of single implant placement and long-term observation of marginal bone levels
3. Kim, J.J., Lee, J.H., Kim, J.C., Lee, J.B., and Yeo, I.L. (2019). Biological responses to the transitional area of dental implants: Material- and structure-dependent responses of peri-implant tissue to abutments. Materials, 13.10.3390/ma13010072 ↗
4. Evaluation of the three methods of bacterial decontamination on implants with three different surfaces10.17219/acem/112606 ↗
5. A two-piece dental implant comprising a replaceable thin titanium sleeve may be a new approach to re-osseointegration following peri-implant disease10.1016/j.mehy.2018.09.029 ↗
6. Fatigue induced changes in conical implant-abutment connections10.1016/j.dental.2015.09.004 ↗
7. Association Between Implant-Abutment Microgap and Implant Circularity to Bacterial Leakage: An In Vitro Study Using Tapered Connection Implants10.11607/jomi.5836 ↗
8. Persistent acute inflammation at the implant-abutment interface10.1177/154405910308200316 ↗
9. Microbial leakage and marginal fit of the implant-abutment interface
10. Bacterial leakage into and from prefabricated screw-retained implant-borne crowns in vitro10.1046/j.1365-2842.1998.00261.x ↗
11. Fixture-abutment connection surface and micro-gap measurements by 3D micro-tomographic technique analysis
12. Micro-CT Analysis of Microgap at a Novel Two-Piece Dental Implant Comprising a Replaceable Sleeve In Vitro10.11607/jomi.8563 ↗
13. La connessione implantare: Meccanismi e implicazioni cliniche
14. Tribst, J.P.M., Özkara, N., Blom, E.J., Kleverlaan, C.J., Ausiello, P., Bruhnke, M., Feilzer, A.J., and Dal Piva, A.M.d.O. (2025). The Effect of Implant–Abutment Contact Area on the Stress Generation of Bone-Level and Tissue-Level Implants. Appl. Sci., 15.10.3390/app15052744 ↗
15. Impact of implantoplasty on strength of the implant-abutment complex10.11607/jomi.3227 ↗
16. In vitro evaluation of the implant-abutment bacterial seal: The locking taper system
17. Manfredini, M., Ghizzoni, M., Cusaro, B., Beretta, M., Maiorana, C., Souza, F.Á., and Poli, P.P. (2025). High Insertion Torque-Clinical Implications and Drawbacks: A Scoping Review. Medicina, 61.10.3390/medicina61071187 ↗