In Silico Investigation of an Innovative Cone-Beam CT Configuration for Quantitative Imaging

Antonio Sarno, Ivan Veronese, Paolo Mauriello, Immacolata Vanore, Antonio Minopoli, Carlos Maximiliano Mollo, Silvio Pardi, Gianfranco Paternò, Mariagabriella Pugliese, Riccardo de Asmundis, Paolo Cardarelli

Applied Sciences· Volume 16· Issue 3

Abstract

Quantitative evaluations in 3D images acquired via Cone-Beam Computed Tomography (CBCT) are limited by the scatter abundance and cone-beam artifacts. This work investigates benefits in using an innovative scanning geometry in CBCT (eCT), which replaces each projection of the conventional scanning protocol with a series of collimated projections (Np) acquired over an oscillating trajectory, realized either with an oscillating source or a multi-spot array. In silico tests employed a cylindrical water phantom embodying inserts of four biological materials. 1 mm-thick bone slabs were sandwiched between 9 mm water slabs to evaluate the image conspicuity. eCT improved the Hounsfield Unit (HU) accuracy, with a direct relation with Np. eCT with Np = 10 reduced the bias of the estimated HU more than two times when compared to CBCT. Increasing the Np presented a large impact on the image conspicuity for portions of the field of view (FOV) distant from the central axial plane, with the signal-to-noise ratio between water and bone slabs increasing by a factor of 18 for Np = 10 compared to CBCT. The proposed eCT configuration is expected to be adopted in applications without strict demand for scanning time and projection number, such as dentomaxillofacial and intrasurgical imaging, imaging of the extremities, and image-guided radiotherapy.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Medical Imaging Techniques and Applications · Dental Radiography and Imaging · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Dipartimento di Fisica “Aldo Pontemoli”, Università degli Studi di Milano, Via Celoria 16, 20133 Milano, Italy

INFN Sezione di Milano, Via Celoria 16, 20133 Milano, Italy

Dipartimento di Fisica “Ettore Pancini”, Università di Napoli Federico II, Complesso Universitario di Monte Sant’Angelo, Via Cinthia 21, 80126 Napoli, Italy

INFN Sezione di Napoli, Complesso Universitario di Monte Sant’Angelo, Via Cinthia 21, 80126 Napoli, Italy

INFN Sezione di Ferrara, Via G. Saragat 1, 44122 Ferrara, Italy

Funding

National Recovery and Resilience Plan · European Union—NextGenerationEU– · HPC IBiSCo Cluster

References (49)

Show source-provided references
1. Flat-panel conebeam CT in the clinic: History and current state10.1117/1.JMI.8.5.052115 ↗
2. Distefano, S., Cannarozzo, M.G., Spagnuolo, G., Bucci, M.B., and Lo Giudice, R. (2023). The “dedicated” CBCT in dentistry. Int. J. Environ. Res. Public Health, 20.10.3390/ijerph20115954 ↗
3. Polizzi, A., Serra, S., and Leonardi, R. (2024). Use of CBCT in orthodontics: A scoping review. J. Clin. Med., 13.10.3390/jcm13226941 ↗
4. A systematic review and meta-analysis on the concept of bone quality in dento-maxillofacial Cone Beam Computed Tomography10.1007/s11547-025-02052-5 ↗
5. Webster, M., Podgorsak, A., Li, F., Zhou, Y., Jung, H., Yoon, J., Lemus, O.D., and Zheng, D. (2025). New approaches in radiotherapy. Cancers, 17.10.3390/cancers17121980 ↗
6. Lintz, F., de Cesar Netto, C., Belvedere, C., Leardini, A., Bernasconi, A., and The International Weight-Bearing CT Society (2024). Recent Innovations Brought about by Weight-Bearing CT Imaging in the Foot and Ankle: A Systematic Review of the Literature. Appl. Sci., 14.10.3390/app14135562 ↗
7. Cone-Beam CT of the Extremities in Clinical Practice10.1148/rg.230143 ↗
8. Perspectives of cone-beam computed tomography in interventional radiology: Techniques for planning, guidance, and monitoring10.1016/j.tvir.2023.100912 ↗
9. Fluoroscopy and Cone Beam CT Guidance in Robotic Interventions10.1016/j.tvir.2024.101007 ↗
10. Emerging clinical applications for cone beam breast CT: Changing the breast imaging paradigm10.1007/s12609-024-00535-4 ↗
11. Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of X-ray scatter10.1118/1.1339879 ↗
12. Shading artifact correction in breast CT using an interleaved deep learning segmentation and maximum-likelihood polynomial fitting approach10.1002/mp.13599 ↗
13. Reduction of scatter in breast CT yields improved microcalcification visibility10.1088/1361-6560/abae07 ↗
14. A dedicated cone-beam CT system for musculoskeletal extremities imaging: Design, optimization, and initial performance characterization10.1118/1.3611039 ↗
15. Cone beam CT multisource configurations: Evaluating image quality, scatter, and dose using phantom imaging and Monte Carlo simulations10.1088/1361-6560/abc306 ↗
16. Can Dental Cone Beam Computed Tomography Assess Bone Mineral Density?10.11005/jbm.2014.21.2.117 ↗
17. Automated planning of breast radiotherapy using cone beam CT imaging10.1118/1.4905111 ↗
18. Repeatability of CBCT radiomic features and their correlation with CT radiomic features for prostate cancer10.1002/mp.14787 ↗
19. Quantitative Cone-Beam CT Imaging in Radiotherapy: Parallel Computation and Comprehensive Evaluation on the TrueBeam System10.1109/ACCESS.2019.2902168 ↗
20. How accurate is CBCT in measuring bone density? A comparative CBCT-CT in vitro study10.1111/cid.12027 ↗
21. QCBCT-NET for direct measurement of bone mineral density from quantitative cone-beam CT: A human skull phantom study10.1038/s41598-021-94359-2 ↗
22. Mason, J.H. (2018). Quantitative Cone-Beam Computed Tomography Reconstruction for Radiotherapy Planning. [Ph.D. Thesis, The University of Edinburgh]. Available online: http://hdl.handle.net/1842/33193.
23. Assessing the reproducibility of CBCT-derived radiomics features using a novel three-dimensional printed phantom10.1002/mp.15043 ↗
24. Can radiomics features be reproducibly measured from CBCT images for patients with non-small cell lung cancer?10.1118/1.4934826 ↗
25. CBCT quantitative evaluation of mandibular lingual concavities in dental implant patients10.1007/s00276-015-1493-9 ↗
26. Quantitative CBCT evaluation of maxillary and mandibular cortical bone thickness and density variability for orthodontic miniplate placement
27. Quantitative Cone-Beam CT Imaging in Radiation Therapy Using Planning CT as a Prior: First Patient Studies10.1118/1.3693050 ↗
28. Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system10.1007/s00330-014-3546-6 ↗
29. Cone-beam computed tomography with ultralow dose protocols for pre-implant radiographic assessment: An in vitro study
30. Comparing the precision of panoramic radiography and cone-beam computed tomography in avoiding anatomical structures critical to dental implant surgery: A retrospective study10.5624/isd.2018.48.4.269 ↗
31. The accuracy of cone-beam computed tomography for evaluating bone density and cortical bone thickness at the implant site: Micro-computed tomography and histologic analysis10.1097/SCS.0000000000004672 ↗
32. A mobile isocentric C-arm for intraoperative cone-beam CT: Technical assessment of dose and 3D imaging performance10.1002/mp.13983 ↗
33. Validation of radiological reduction criteria with intraoperative cone beam CT in unstable syndesmotic injuries10.1007/s00068-020-01299-z ↗
34. Generalizable cone beam CT esophagus segmentation using physics-based data augmentation10.1088/1361-6560/abe2eb ↗
35. Evaluation of the feasibility of a multisource CBCT for maxillofacial imaging10.1088/1361-6560/acea17 ↗
36. Li, B., Spronk, D., Luo, Y., Puett, C., Inscoe, C.R., Tyndall, D.A., Lee, Y.Z., Lu, J., and Zhou, O. (2022). Feasibility of dual-energy CBCT by spectral filtration of a dual-focus CNT X-ray source. PLoS ONE, 17.10.1371/journal.pone.0262713 ↗
37. Volumetric computed tomography with carbon nanotube X-ray source array for improved image quality and accuracy10.1038/s44172-023-00123-x ↗
38. Image quality and dose for a multisource cone-beam CT extremity scanner10.1002/mp.12659 ↗
39. Multisource cone beam computed tomography using a carbon nanotube X-ray source array
40. X-ray scatter correction for multi-source interior computed tomography10.1002/mp.12022 ↗
41. Noise power properties of a cone-beam CT scanner with unconventional scanning geometry10.1016/j.ejmp.2024.104888 ↗
42. ICRP (2009). ICRP Publication 110: Adult Reference Computational Phantoms. Ann. ICRP, 39, Available online: https://www.icrp.org/publication.asp?id=icrp%20publication%20110.
43. Accelerating Monte Carlo simulations of photon transport in a voxelized geometry using a massively parallel graphics processing unit10.1118/1.3231824 ↗
44. A model for a linear a-Se detector in simulated X-ray breast imaging with Monte Carlo software10.1109/TRPMS.2024.3349563 ↗
45. Minopoli, A., and Sarno, A. (2026, January 20). CBCT/ExoCT Reconstruction Software. Available Under Creative Common Attribution 4.0 License. Available online: https://www.openaccessrepository.it/records/wxdh6-2y475.
46. Improving the accuracy of bone mineral density using a multisource CBCT10.1038/s41598-024-54529-4 ↗
47. Bosmans, H., Chen, G.-H., and Gilat Schmidt, T. (2019). Multi-X-ray source array for stationary tomosynthesis or multi-cone angle cone beam CT. Medical Imaging 2019: Physics of Medical Imaging, SPIE.
48. A multiple X-ray-source array (MXA) system with a planar two-dimensional source distribution for digital breast tomosynthesis10.1002/mp.17452 ↗
49. Hu, Y., Li, B., Xu, S., Inscoe, C.R., Tyndall, D.A., Lee, Y.Z., Lu, J., and Zhou, O. (2026). A Dual-Energy CBCT With Reduced Scatter and Cone Beam Artifacts Using an X-Ray Source Array and Interlaced Spectral Filters. IEEE Trans. Biomed. Eng., 1–11.10.1109/TBME.2026.3651411 ↗