Design of a Motor-Torsion Driven 3D-Printed Left Ventricular Mock Circulation System and Hemocompatibility Assessment
Source abstract
Abstract
In vitro testing of ventricular assist devices, constructing a mock circulation system that reproduces physiological cardiac function, is critical. However, current ventricular simulators often lack biomimetic fidelity and may introduce hemolysis and coagulation risks during prolonged operation, affecting hemocompatibility assessment. This study proposes a motor-driven torsional 3D-printed left ventricular simulator to reconstruct the hemodynamics of severe heart failure and related pathological conditions. The system integrates a 3D-printed elastic ventricular model with programmable torsional actuation, allowing the simulation of various cardiac conditions by adjusting the motor torsion angle and rotational speed, peripheral resistance and compliance. Fresh porcine blood was circulated for 4 h in a closed-loop system, with periodic measurements of plasma-free hemoglobin (PfHb), thrombin–antithrombin complex (TAT), and P-selectin. The results show that the system successfully reproduces typical hemodynamic features of severe heart failure, while hemolysis and coagulation markers remain low. After 4 h, PfHb was below 20 mg/dL, with no significant platelet activation or thrombosis. This study demonstrates that the proposed system enhances biomimicry while maintaining excellent hemocompatibility, offering a reliable platform for in vitro performance and safety evaluation of ventricular assist devices.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Mechanical Circulatory Support Devices · Cardiac Structural Anomalies and Repair · Cardiomyopathy and Myosin Studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
School of Automation and Intelligent Sensing, Shanghai Jiao Tong University, Shanghai 200240, China
Funding
the National Key R&D Program of China
References (34)
Show source-provided references
1. Nakamura, M., Imamura, T., Hida, Y., and Kinugawa, K. (2022). Pulmonary Artery Pulsatility Index and Hemolysis during Impella-Incorporated Mechanical Circulatory Support. J. Clin. Med., 11.10.3390/jcm11051206 ↗
2. Attinger-Toller, A., Bossard, M., Cioffi, G.M., Tersalvi, G., Madanchi, M., Bloch, A., Kobza, R., and Cuculi, F. (2022). Ventricular Unloading Using the ImpellaTM Device in Cardiogenic Shock. Front. Cardiovasc. Med., 9.10.3389/fcvm.2022.856870 ↗
3. What Is Known in Pre-, Peri-, and Post-Procedural Anticoagulation in Micro-Axial Flow Pump Protected Percutaneous Coronary Intervention?10.1093/eurheartjsupp/suac066 ↗
4. Mock Circulatory Loops Used for Testing Cardiac Assist Devices: A Review of Computational and Experimental Models10.1177/03913988211045405 ↗
5. Pathophysiology of Heart Failure10.21037/cdt-20-302 ↗
6. Baturalp, T.B., and Bozkurt, S. (2024). Design and Analysis of a Polymeric Left Ventricular Simulator via Computational Modelling. Biomimetics, 9.10.3390/biomimetics9050269 ↗
7. Soft Robotic Artificial Left Ventricle Simulator Capable of Reproducing Myocardial Biomechanics10.1126/scirobotics.ado4553 ↗
8. An Organosynthetic Dynamic Heart Model with Enhanced Biomimicry Guided by Cardiac Diffusion Tensor Imaging10.1126/scirobotics.aay9106 ↗
9. Robotic Right Ventricle Is a Biohybrid Platform That Simulates Right Ventricular Function in (Patho)Physiological Conditions and Intervention10.1038/s44161-023-00387-8 ↗
10. The Use of Computational Fluid Dynamics in the Development of Ventricular Assist Devices10.1016/j.medengphy.2010.10.014 ↗
11. Gao, X., Xu, Z., Chen, C., Hao, P., He, F., and Zhang, X. (2023). Full-Scale Numerical Simulation of Hemodynamics Based on Left Ventricular Assist Device. Front. Physiol., 14.10.3389/fphys.2023.1192610 ↗
12. MRI-Based CFD Analysis of Flow in a Human Left Ventricle: Methodology and Application to a Healthy Heart10.1007/s10439-008-9627-4 ↗
13. Comunale, G., Di Micco, L., Boso, D.P., Susin, F.M., and Peruzzo, P. (2021). Numerical Models Can Assist Choice of an Aortic Phantom for In Vitro Testing. Bioengineering, 8.10.3390/bioengineering8080101 ↗
14. Kameneva, M.V., Burgreen, G.W., Kono, K., Repko, B., Antaki, J.F., and Umezu, M. (2019). Effects of Turbulent Stresses on Mechanical Hemolysis: Experimental and Computational Analysis. American Society for Artificial Internal Organs (ASAIO) Platinum 70th Anniversary Special Edition, CRC Press.
15. Model Reference Adaptive Control for Aortic Pressure Regulation in Ex Vivo Heart Perfusion10.1109/TCST.2020.2978448 ↗
16. Reproduction of Human Blood Pressure Waveform Using Physiology-Based Cardiovascular Simulator10.1038/s41598-023-35055-1 ↗
17. Wu, P., Bai, Y., Du, G., Zhang, L., and Zhao, X. (2023). Resistance Valves in Circulatory Loops Have a Significant Impact on in Vitro Evaluation of Blood Damage Caused by Blood Pumps: A Computational Study. Front. Physiol., 14.10.3389/fphys.2023.1287207 ↗
18. Comprehensive Review of Hemolysis in Ventricular Assist Devices10.4330/wjc.v12.i7.334 ↗
19. Borchers, P., Winnersbach, P., Kraemer, S., Beckers, C., Buhl, E., Leonhardt, S., Rossaint, R., Walter, M., Breuer, T., and Bleilevens, C. (2022). Comparison of the Hemocompatibility of an Axial and a Centrifugal Left Ventricular Assist Device in an In Vitro Test Circuit. J. Clin. Med., 11.10.3390/jcm11123431 ↗
20. Citrate-Buffered Dialysis Solution (Citrasate) Allows Avoidance of Anticoagulation During Intermittent Hemodiafiltration—At the Cost of Decreased Performance and Systemic Biocompatibility10.1111/aor.12851 ↗
21. Soluble P-Selectin in Human Plasma: Effect of Anticoagulant Matrix and Its Levels in Patients with Cardiovascular Disorders10.1177/107602960000600204 ↗
22. Zhu, Y., Wang, S., Liu, Y., Yu, J., and Yang, M. (2025). The Impact of Transient Control Performance of Pulsatile Flow on Hemolysis and Coagulation in ex vivo Heart Perfusion. Artif. Organs.10.1111/aor.15051 ↗
23. Classification of Heart Failure According to Ejection Fraction10.1016/j.jacc.2021.04.070 ↗
24. Real-Time Regurgitation Estimation in Percutaneous Left Ventricular Assist Device Fully Supported Condition Using an Unscented Kalman Filter10.1088/1361-6579/ad3d29 ↗
25. Practical Implications of the Erroneous Treatment of Exposure Time in the Eulerian Hemolysis Power Law Model10.1111/aor.14543 ↗
26. High Shear Induces Platelet Dysfunction Leading to Enhanced Thrombotic Propensity and Diminished Hemostatic Capacity10.1080/09537104.2017.1384542 ↗
27. Liu, Y., Zhu, Y., Wang, S., Fu, H., Lu, Z., and Yang, M. (2024). Impact of Impeller Speed Adjustment Interval on Hemolysis Performance of an Intravascular Micro-Axial Blood Pump. Micromachines, 15.10.3390/mi15070934 ↗
28. Multi-Method Investigation of Blood Damage Induced by Blood Pumps in Different Clinical Support Modes10.1097/MAT.0000000000002116 ↗
29. Hemolysis Testing In Vitro: A Review of Challenges and Potential Improvements10.1097/MAT.0000000000001454 ↗
30. Hemolysis Associated with Impella Heart Pump Positioning: In Vitro Hemolysis Testing and Computational Fluid Dynamics Modeling10.1177/0391398820909843 ↗
31. Impella 5.5 Versus Centrimag: A Head-to-Head Comparison of Device Hemocompatibility10.1097/MAT.0000000000001283 ↗
32. Heartmate 3 Fully Magnetically Levitated Left Ventricular Assist Device for the Treatment of Advanced Heart Failure—1 Year Results from the Ce Mark Trial10.1186/s13019-017-0587-3 ↗
33. The Influence of Different Operating Conditions on the Blood Damage of a Pulsatile Ventricular Assist Device10.1097/MAT.0000000000000261 ↗
34. Schulte, D.T., Pietsch, C., Topalovic, N., Hofmann, M., Schmiady, M.O., Weisskopf, M., and Schmid Daners, M. (2025). Preliminary in Vitro Hemolysis Evaluation of MR-Conditional Blood Pumps. Front. Med. Technol., 7.10.3389/fmedt.2025.1671938 ↗