journal article·2025-11-02Open access

Parametric Study and Hemocompatibility Assessment of a Centrifugal Blood Pump Based on CFD Simulation and Experimental Validation

Yiwen Wang, Libo Xin, Qinghong Weng

Applied Sciences· Volume 15· Issue 21

Abstract

The heart is the body’s core pump. Heart failure impairs the heart’s ability to pump blood, leading to circulatory disorders. The artificial heart (blood pump) is an important mechanical circulatory support device that can partially or completely substitute cardiac pumping function, potentially improving hemodynamic performance and alleviating symptoms of heart failure. A combination of computational fluid dynamics simulation and hydraulic performance testing was used to study key parameters of the impeller, including blade count, blade wrap angle, impeller flow path, and diversion cone height. The goal was to reduce hemolysis risk and enhance pumping efficiency. Increasing the blade count raised the head, with optimal efficiency achieved at seven blades. A larger blade wrap angle decreased the head but improved efficiency. Synchronizing the flow path and diversion cone height at 4.1 mm maximized the head. Under various rotational speeds, the studied hemolysis index remained well below 0.1 g/100 L. Both experimental and simulation data were validated against each other, meeting the required error tolerances. The studied blood pump meets the design specifications. At an operating condition of 5 L/min flow rate and 2800 rpm, the pump achieves the required head and hemolysis criteria with a margin of safety.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Mechanical Circulatory Support Devices · Cavitation Phenomena in Pumps · Hydraulic and Pneumatic Systems

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Mechanical and Power Engineering, Harbin University of Science and Technology, Harbin 150080, China

Funding

Natural Science Foundation of Heilongjiang Province of China

References (31)

Show source-provided references
1. Wang, S., Jia, T., Liu, G., Lu, X., Yang, Q., and Zhu, C. (2022). Mapping the research trends and hot topics of ventricular arrhythmia: A bibliometric analysis from 2001 to 2020. Front. Cardiovasc. Med., 9.10.3389/fcvm.2022.856695 ↗
2. Inhibition of MEG3 ameliorates cardiomyocyte apoptosis and autophagy by regulating the expression of miRNA-129-5p in a mouse model of heart failure10.1080/13510002.2023.2224607 ↗
3. Dandel, M. (2023). Cardiological Challenges Related to Long-Term Mechanical Circulatory Support for Advanced Heart Failure in Patients with Chronic Non-Ischemic Cardiomyopathy. J. Clin. Med., 12.10.3390/jcm12206451 ↗
4. Optimization of hemocompatibility metrics in ventricular assist device design using machine learning and CFD-based response surface analysis10.1177/03913988251346712 ↗
5. Oran, E., Abo-Serie, E., Jewkes, J., Henry, M., and Oran, B. (2024). Design and optimisation of an Intra-Aortic Shrouded rotor axial pump. J. Biomech., 162.10.1016/j.jbiomech.2023.111858 ↗
6. Numerical study of a centrifugal blood pump with different impeller profiles10.1097/MAT.0b013e3181c8f066 ↗
7. Miniaturized soft centrifugal pumps with magnetic levitation for fluid handling10.1126/sciadv.abi7203 ↗
8. Deng, Q., Pei, J., Wang, W., Lin, B., Zhang, C., and Zhao, J. (2021). Energy Loss and Radial Force Variation Caused by Impeller Trimming in a Double-Suction Centrifugal Pump. Entropy, 23.10.3390/e23091228 ↗
9. Channel impeller design for centrifugal blood pump in hybrid pediatric total artificial heart: Modeling, magnet integration, and hydraulic experiments10.1111/aor.14480 ↗
10. Analysis of fluid forces impacting on the impeller of a mixed flow blood pump with computational fluid dynamics10.1177/03913988241293003 ↗
11. Ji, D., Lu, W., Xu, B., Xu, L., and Lu, L. (2023). Study of Energy Loss Characteristics of a Shaft Tubular Pump Device Based on the Entropy Production Method. Entropy, 25.10.3390/e25070995 ↗
12. Parente, R., Fumagalli, M.R., Di Claudio, A., Rincón, C.L.C., Erreni, M., Zanini, D., Iapichino, G., Protti, A., Garlanda, C., and Rusconi, R. (2023). A Multilayered Imaging and Microfluidics Approach for Evaluating the Effect of Fibrinolysis in Staphylococcus aureus Biofilm Formation. Pathogens, 12.10.3390/pathogens12091141 ↗
13. Zhu, L., Fang, Y., Shen, Y., Li, H., Wang, Y., Zhu, X., and Lu, X. (2025). Structural characteristics of corrugated sandwich rectangular pressure vessels. Sci. Rep., 15.10.1038/s41598-025-95592-9 ↗
14. On the Significance of Systolic Flow Waveform on Aortic Valve Energy Loss10.1007/s10439-018-2102-y ↗
15. Wall Shear Stress Estimation for 4D Flow MRI Using Navier-Stokes Equation Correction10.1007/s10439-022-02993-2 ↗
16. Effect of microplastics deposition on human lung airways: A review with computational benefits and challenges10.1016/j.heliyon.2024.e24355 ↗
17. Research on a portable electrical impedance tomography system for evaluating blood compatibility of biomaterials
18. Hemodynamic evaluation and in vitro hemolysis evaluation of a novel centrifugal pump for extracorporeal membrane oxygenation10.21037/atm-21-1135 ↗
19. Yun, Z., Yao, J., Wang, L., Tang, X., and Feng, Y. (2023). The design and evaluation of the outflow structures of an interventional microaxial blood pump. Front. Physiol., 14.10.3389/fphys.2023.1169905 ↗
20. Optimizing the design of a pediatric blood pump through orthogonal experimentation10.1016/j.heliyon.2025.e42502 ↗
21. Investigation of the influence of blade configuration on the hemodynamic performance and blood damage of the centrifugal blood pump10.1111/aor.14265 ↗
22. Wang, J., Nakata, T., and Liu, H. (2019). Development of Mixed Flow Fans with Bio-Inspired Grooves. Biomimetics, 4.10.3390/biomimetics4040072 ↗
23. Integrated long-term multifunctional pediatric mechanical circulatory assist device
24. Medina, R., Sánchez, R.-V., Cabrera, D., Cerrada, M., Estupiñan, E., Ao, W., and Vásquez, R.E. (2024). Scale-Fractal Detrended Fluctuation Analysis for Fault Diagnosis of a Centrifugal Pump and a Reciprocating Compressor. Sensors, 24.10.3390/s24020461 ↗
25. Peng, G., Chen, Q., Zhou, L., Pan, B., and Zhu, Y. (2020). Effect of Blade Outlet Angle on the Flow Field and Preventing Overload in a Centrifugal Pump. Micromachines, 11.10.3390/mi11090811 ↗
26. Yu, N., Yu, X., Cai, Z., Yang, X., Tian, L., and Li, S. (2025). Research on multi-condition optimization of centrifugal compressor impeller meridian profile. PLoS ONE, 20.10.1371/journal.pone.0318783 ↗
27. A mathematical model for predicting cardiovascular responses at rest and during exercise in demanding environmental conditions10.1152/japplphysiol.00619.2021 ↗
28. Coronary perfusion pressure is associated with adverse outcomes in advanced heart failure10.1177/02676591221118693 ↗
29. Reymond, P., Bendjelid, K., Giraud, R., Richard, G., Murith, N., Cikirikcioglu, M., and Huber, C. (2021). To Balloon or Not to Balloon? The Effects of an Intra-Aortic Balloon-Pump on Coronary Artery Flow during Extracorporeal Circulation Simulating Normal and Low Cardiac Output Syndromes. J. Clin. Med., 10.10.3390/jcm10225333 ↗
30. The history of continuous-flow blood pumps10.1046/j.1525-1594.2000.06652.x ↗
31. Transfusion thresholds for guiding red blood cell transfusion