An Automated Approach for Calibrating Gafchromic EBT3 Films and Mapping 3D Doses in HDR Brachytherapy

Labinot Kastrati, Burim Uka, Polikron Dhoqina, Gezim Hodolli, Sehad Kadiri, Behar Raci, Faton Sermaxhaj, Kjani Guri, Hekuran Sejdiu

Applied Sciences· Volume 15· Issue 19

Abstract

The accurate calibration of radiochromic films is critical for high dose rate (HDR) brachytherapy dosimetry. Conventional workflows frequently rely on manually determined regions of interest (ROIs), which might increase operator variability. In this investigation, Gafchromic EBT3 films were irradiated under clinical settings at nominal doses of 0–10 Gy and evaluated using a MATLAB (R2024b)-based tool that allows for both manual and automated ROI selection. The calibration curves were modeled with a second-order polynomial and rational model, and performance was assessed using statistical measures. The study found that the rational model fits better than the polynomial model. Additionally, the automatic ROI approach outperformed the manual method in both models, resulting in higher calibration accuracy and reproducibility (R2 = 0.999, RMSE = 0.118 Gy, MAE = 0.103 Gy vs. R2 = 0.986, RMSE = 0.448 Gy, MAE = 0.388 Gy). Although manual ROI occasionally produced greater dose–response slopes at higher doses, it was more susceptible to operator bias and film non-uniformity. In contrast, automatic ROI reduced variability by consistently picking homogeneous sections, resulting in steady curve fitting across the entire dose range. Furthermore, a companion module transformed calibrated films into 2D false-color maps and 3D dosage surfaces, allowing for visual assessment of dose uniformity, detection of scanner-related aberrations, and quantitative verification for quality assurance. These findings demonstrate that automated ROI selection provides a more stable and reproducible foundation for film calibration in HDR brachytherapy, minimizing operator dependency while facilitating routine clinical quality assurance.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Radiotherapy Techniques · Medical Imaging Techniques and Applications · Radiation Therapy and Dosimetry

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Faculty of Mechanical and Computer Engineering, University of Mitrovica “Isa Boletini”, 40000 Mitrovica, Kosovo

Clinical Oncology, University Clinical Center of Kosovo, 10000 Prishtina, Kosovo

Faculty of Medical Sciences, University for Business and Technology, 10000 Prishtina, Kosovo

Faculty of Natural Sciences, University of Tirana, 1000 Tirana, Albania

Faculty of Veterinary and Agriculture, University of Prishtina, 10000 Prishtina, Kosovo

Faculty of Radiology, AAB College, 10000 Prishtina, Kosovo

Funding

“AL TRADE CENTER” LLC

References (17)

Show source-provided references
1. Pathak, P., Thomas, J.J., Baghwala, A., Li, C., Teh, B.S., Butler, E.B., and Farach, A.M. (2024). Personalized Brachytherapy: Applications and Future Directions. Cancers, 16.10.3390/cancers16193424 ↗
2. Current status of brachytherapy in cancer treatment—Short overview10.5114/jcb.2017.72607 ↗
3. Applications and benefits of using gradient percentage depth dose instead of percentage depth dose for electron and photon beams in radiotherapy10.2478/pjmpe-2021-0004 ↗
4. A comparative study between MC simulation and TLD measurements of radiation doses to other parts of the body during gynaecological brachytherapy10.1007/s12553-024-00905-z ↗
5. In Vivo Dosimetry for Superficial High Dose Rate Brachytherapy with Optically Stimulated Luminescence Dosimeters: A Comparison Study with Metal-Oxide-Semiconductor Field-Effect Transistors10.3390/radiation2040026 ↗
6. Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array10.5114/jcb.2018.75599 ↗
7. Radiochromic film–based quality assurance for CT-based high-dose-rate brachytherapy10.1016/j.brachy.2015.02.192 ↗
8. Dosimetry evaluation of 192Ir Flexisource using gafchromic EBT3 film10.1017/S1460396924000177 ↗
9. Gafchromic film dosimetry of a new HDR brachytherapy source10.1120/jacmp.v17i2.6005 ↗
10. Evaluating dosimetric accuracy of the 6 MV calibration on EBT3 film in the use of Ir-192 high dose rate brachytherapy10.1002/acm2.13571 ↗
11. Dosimetric evaluation and Monte Carlo simulation of a new proposed surface brachytherapy mould10.5114/jcb.2024.143609 ↗
12. A system for mailed dosimetric audits of 192Ir and 60Co HDR brachytherapy combining OSLD and radiochromic film10.1016/j.radmeas.2024.107101 ↗
13. High spatial resolution dosimetry with uncertainty analysis using Raman micro-spectroscopy readout of radiochromic films10.1002/mp.15000 ↗
14. Rieker, V.F., Bateman, J.J., Wroe, L.M., Caloz, M., Grilj, V., Aguiar, Y.Q., Schüller, A., Bailat, C., Farabolini, W., and Gilardi, A. (2025). Radiochromic Film Dosimetry for VHEE and UHDR: Considerations for the CLEAR Facility and Comparisons with Alanine, RPL and Dosimetry Phantoms. arXiv.
15. Implementation of safety and security standards for high dose rate brachytherapy sealed radioactive source used in a medical radiation facility10.1016/j.radmp.2020.11.004 ↗
16. Calibration of the EBT3 Gafchromic Film Using HNN Deep Learning10.1155/2021/8838401 ↗
17. Calibration of GafChromic EBT3 for absorbed dose measurements in 5 MeV proton beam and (60)Co γ-rays10.1118/1.4926558 ↗