Assessing the Reliability of 3D-Printed Custom Silicone Boluses in Radiotherapy: Thickness and Air Bubble Considerations
Source abstract
Abstract
This study aimed to evaluate the scientific reliability of 3D-printed silicone boluses fabricated with patient-specific molds, focusing on fabrication-related uncertainties such as internal air bubbles, thickness variations, and density differences, thereby providing evidence for clinical quality assurance. Custom silicone boluses were fabricated using 3D-printed molds with varying vacuum degassing times (1, 5, and 10 min). Air bubble size and depth were quantified using scanner image analysis, while density and Hounsfield unit (HU) values were compared with a commercial bolus. Dosimetric evaluation was performed using a VitalBeam linear accelerator (6 MV photons, Varian Medical Systems, Palo Alto, CA, USA) and a MatriXX 2D detector (IBA Dosimetry, Schwarzenbruck, Germany), comparing treatment planning system (TPS) calculated doses with measured doses across a 3 × 3 grid. Surface dose distributions were further analyzed using EBT3 film. Results showed that bubble size increased with longer vacuum times, interpreted as coalescence due to limited degassing and silicone viscosity. The density of 3D boluses ranged from 0.980 to 1.104 g/cm3 (commercial: 0.988 g/cm3), with HU values of +240 to +250 (commercial: −110). In point-wise comparisons, mean dose differences were less than 1% for 1- and 5 min samples and approximately 1% for 10 min, with all conditions within |Δ| ≤ 3%. Film analysis confirmed equivalent surface dose distributions. These findings demonstrate, for the first time, that microscopic bubbles in 3D-printed silicone boluses have negligible clinical impact, supporting their safe adoption without requiring complex degassing procedures.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Radiotherapy Techniques · Anatomy and Medical Technology · Surgical Simulation and Training
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Radiation Oncology, Keimyung University, Dongsan Hospital, Dalgubeol-daero, Daegu 42601, Republic of Korea
Department of Radiation Oncology, Keimyung University School of Medicine, Daegu 42601, Republic of Korea
References (25)
Show source-provided references
1. Estimation of surface dose in the presence of unwanted air gaps under the bolus in postmastectomy radiation therapy: A phantom dosimetric study10.31557/APJCP.2022.23.9.2973 ↗
2. On bolus for megavoltage photon and electron radiation therapy10.1016/j.meddos.2013.02.007 ↗
3. Bolus use in postmastectomy radiation therapy for breast cancer: A systematic literature review10.1177/15330338251344521 ↗
4. Effects on skin dose from unwanted air gaps under bolus in photon beam radiotherapy10.1016/S1350-4487(99)00276-0 ↗
5. Monte Carlo investigation on the effect of air gap under bolus in post-mastectomy radiotherapy10.1016/j.ejmp.2018.10.023 ↗
6. Technical note: Evaluation of a silicone-based custom bolus for radiation therapy of a superficial pelvic tumor10.1002/acm2.13538 ↗
7. 3D-printed boluses for radiotherapy: Influence of geometrical and printing parameters on dosimetric characterization and air gap evaluation10.1007/s12194-024-00782-1 ↗
8. Innovative Approach for Generating Soft Silicone Bolus Using 3-Dimensional Printing for Electron Treatment of Skin Cancers in Areas with Irregular Contours
9. Development and clinical implementation of 3D printed bolus in radiation therapy10.1016/j.prro.2017.11.001 ↗
10. Comprehensive clinical implementation, workflow, and FMEA of bespoke silicone bolus cast from 3D printed molds using open-source resources10.1002/acm2.14498 ↗
11. Son, J., Kang, S., Jin, J., Park, H., Lee, I., Huh, Y., Choi, C.H., Kim, J.-I., and Wu, H.-G. (2025). Clinical application of patient-specific bolus based on molding and casting method in radiotherapy. J. Clin. Med., 14.10.3390/jcm14113796 ↗
12. Evaluations of patient-specific bolus fabricated by mold-and-cast method using computer numerical control machine tools10.1093/jrr/rrad075 ↗
13. Fabrication and dosimetric characteristics of silicon elastomer-based bolus using external beam radiotherapy10.31557/APJCP.2023.24.1.141 ↗
14. Yu, S., Ahn, S.H., Choi, S.H., Ahn, W.S., and Jung, I.H. (2023). Clinical application of a customized 3D-printed bolus in radiation therapy for distal extremities. Life, 13.10.3390/life13020362 ↗
15. Wang, X., Wang, X., Xiang, Z., Zeng, Y., Liu, F., Shao, B., He, T., Ma, J., Yu, S., and Liu, L. (2021). The clinical application of 3D-printed boluses in superficial tumor radiotherapy. Front. Oncol., 11.10.3389/fonc.2021.698773 ↗
16. Huang, C.-Y., Yang, B., Lam, W.W., Tang, K.K., Li, T.C., Law, W.K., Cheung, K.Y., and Yu, S.K. (2021). Effects on skin dose from unwanted air gaps under bolus in an MR-guided linear accelerator (MR-linac) system. Phys. Med. Biol., 66.10.1088/1361-6560/abe837 ↗
17. Application of 3D printing for personalized boluses in radiotherapy: A systematic review10.5603/rpor.104014 ↗
18. Smooth-On Inc (2025). Dragon Skin™ 10 NV: Technical Bulletin [Internet], Smooth-On, Inc.. Available online: https://www.smooth-on.com/tb/files/DRAGON_SKIN_10NV_TB.pdf.
19. Intrapatient study comparing 3D printed bolus versus standard vinyl gel sheet bolus for postmastectomy chest wall radiation therapy10.1016/j.prro.2017.12.008 ↗
20. Determining tolerance levels for quality assurance of 3D printed bolus for modulated arc radiotherapy of the nose10.1007/s13246-021-01054-7 ↗
21. Verification using in vivo optically stimulated luminescent dosimetry of the predicted skin surface dose in patients receiving postmastectomy radiotherapy10.1016/j.meddos.2020.10.001 ↗
22. Clinical and dosimetric implications of air gaps between bolus and skin surface during radiation therapy10.4236/jct.2013.47147 ↗
23. Kim, J., Park, J., Park, B., Park, B., and Kim, T.G. (2023). Feasibility of a patient-specific bolus using the life-casting method for radiation therapy. Appl. Sci., 13.10.3390/app13179977 ↗
24. Development and dosimetric verification of 3D customized bolus in head and neck radiotherapy10.1093/jrr/rrac013 ↗
25. Geometric evaluation of patient-specific 3D bolus from 3D printed mold and casting method for radiation therapy10.14316/pmp.2019.30.1.32 ↗