journal article·2024-07-05Open access

Characterization of the First Prototype of an Angular Independent Silicon Diode Array for Quality Assurance in Stereotactic Radiosurgery

Aishah Bashiri, Sean Hood, Jessie Posar, Yashiv Dookie, Joanne McNamara, Joel Poder, Fathimat Zahra, Michael L. F. Lerch, Anatoly B. Rosenfeld, Marco Petasecca

Applied Sciences· Volume 14· Issue 13

Abstract

Quality assurance (QA) ensures the accurate and safe delivery of radiation treatment. However, there are several challenges for advanced radiotherapy techniques, such as stereotactic radiosurgery (SRS), where substantial doses of radiation with multi-directional beams and variable dose rates are delivered to specific areas. Current dosimeters lack high precision, exhibiting issues with dependency on the angle of measurement and the dose rate. This study investigates the characterization of a two-dimensional edgeless silicon diode array for QA in SRS. This detector underwent evaluation of its dose linearity, percentage depth dose (PDD), output factors (OFs), dose rate variability, and angular dependence with megavoltage linear accelerator beams. The edgeless array demonstrated a linear response in the direct detection of MV therapeutic X-rays with sensitivity of 6.95 × 10−3 ± 2.3 × 10−5 Gy/nC, and the percentage differences for PDD and OF measurements were found to be within 2% compared to the reference detector. A dose per pulse dependence of ±2% was demonstrated across the range of 0.12 to 0.39 mGy/pulse. The angular dependence was within 2% variation for irradiation angles greater than 80° and smaller than 120°; however, a maximum of 4% variation was observed with some diodes for angles between 80° and 120°. The improved performance of the edgeless array is likely to overcome limitations of the current dosimeters for SRS QA by operating without the need of any corrections.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Radiotherapy Techniques · Medical Imaging Techniques and Applications · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2522, Australia

Physics Department, Faculty of Science and Arts, Najran University, Najran 66462, Saudi Arabia

Shoalhaven Cancer Care Centre, Shoalhaven Hospital, Nowra, NSW 2541, Australia

St George Cancer Care Centre, Kogarah, NSW 2217, Australia

References (53)

Show source-provided references
1. Stereotactic Radiosurgery of Brain Metastases: A Retrospective Study10.1186/s13014-023-02389-z ↗
2. Stereotactic Radiosurgery versus Whole-Brain Radiotherapy after Resection of Solitary Brain Metastasis: A Systematic Review and Meta-Analysis10.1016/j.wnsx.2023.100170 ↗
3. Impact of Biologically Effective Dose on Tremor Decrease after Stereotactic Radiosurgical Thalamotomy for Essential Tremor: A Retrospective Longitudinal Analysis10.1007/s10143-024-02296-1 ↗
4. Kondziolka, D. (2009). Textbook of Stereotactic and Functional Neurosurgery, Springer.
5. Larson, D.A., Barani, I.J., Roach, M., and Sethi, R.A. (2023). Handbook of Evidence-Based Stereotactic Radiosurgery and Stereotactic Body Radiotherapy, Springer International Publishing.10.1007/978-3-031-33156-5_1 ↗
6. Single-Isocenter Multiple-Target Stereotactic Radiosurgery for Multiple Brain Metastases: Dosimetric Evaluation of Two Automated Treatment Planning Systems10.1186/s13014-022-02086-3 ↗
7. Single Isocenter Stereotactic Irradiation for Multiple Brain Metastases: Current Situation and Prospects10.1007/s11604-022-01333-7 ↗
8. Single-Isocenter Stereotactic Radiosurgery for Multiple Brain Metastases: Impact of Patient Misalignments on Target Coverage in Non-Coplanar Treatments10.1016/j.zemedi.2022.02.005 ↗
9. Development of a Dedicated Phantom for Multi-Target Single-Isocentre Stereotactic Radiosurgery End to End Testing10.1002/acm2.12452 ↗
10. Detection of Rotational Errors in Single-Isocenter Multiple-Target Radiosurgery: Is a Routine Off-Axis Winston–Lutz Test Necessary?10.1002/acm2.13665 ↗
11. The Use of Stereotactic Radiosurgery to Treat Functional Disorders: A Topic Discussion
12. IAEA (2017). Dosimetry of Small Static Fields Used in External Beam Radiotherapy: An IAEA-AAPM International Code of Practice for Reference and Relative Dose Determination (No. 483), IAEA.
13. A Contemporary Review of Stereotactic Radiotherapy: Inherent Dosimetric Complexities and the Potential for Detriment10.3109/0284186X.2010.551665 ↗
14. Effects of Inaccurate Small Field Dose Measurements on Calculated Treatment Doses10.1007/s13246-016-0461-y ↗
15. International Commission on Radiation Units and Measurements (2014). ICRU REPORT 91: Prescribing, Recording, and Reporting of Stereotactic Treatments with Small Photon Beams. J. ICRU, 14, 1–160.10.1093/jicru_ndw040 ↗
16. The Influence of Detector Size Relative to Field Size in Small-Field Photon-Beam Dosimetry Using Synthetic Diamond Crystals as Sensors10.1016/j.radphyschem.2015.04.005 ↗
17. On the Evaluation of Edgeless Diode Detectors for Patient-Specific QA in High-Dose Stereotactic Radiosurgery10.1016/j.ejmp.2021.07.010 ↗
18. Characterization of Small-Field Stereotactic Radiosurgery Beams with Modern Detectors10.1088/0031-9155/58/21/7595 ↗
19. Film-Based Delivery Quality Assurance for Robotic Radiosurgery: Commissioning and Validation10.1016/j.ejmp.2015.05.001 ↗
20. Reference Radiochromic Film Dosimetry: Review of Technical Aspects10.1016/j.ejmp.2016.02.008 ↗
21. Report of AAPM TG 135: Quality Assurance for Robotic Radiosurgery10.1118/1.3579139 ↗
22. Comprehensive Evaluation of the High-Resolution Diode Array for SRS Dosimetry10.1002/acm2.12696 ↗
23. Operation and Calibration of the Novel PTW 1600SRS Detector for the Verification of Single Isocenter Stereotactic Radiosurgery Treatments of Multiple Small Brain Metastases10.1259/bjr.20210473 ↗
24. Comprehensive Characterisation of the IBA MyQA SRS for SRS and SBRT Patient Specific Quality Assurance10.1007/s13246-023-01370-0 ↗
25. Modeling the Instantaneous Dose Rate Dependence of Radiation Diode Detectors10.1118/1.1602171 ↗
26. Spectral Distribution of Particle Fluence in Small Field Detectors and Its Implication on Small Field Dosimetry10.1002/mp.12042 ↗
27. The Physics of Small Megavoltage Photon Beam Dosimetry10.1016/j.radonc.2017.11.001 ↗
28. Quality Assurance of Cyberknife Robotic Stereotactic Radiosurgery Using an Angularly Independent Silicon Detector10.1002/acm2.12496 ↗
29. Padelli, F., Aquino, D., Fariselli, L., and De Martin, E. (2022). IBA MyQA SRS Detector for CyberKnife Robotic Radiosurgery Quality Assurance. Appl. Sci., 12.10.3390/app12157791 ↗
30. Characterization of an “Edgeless” Dosimeter for Angular Independent Measurements in Advanced Radiotherapy Treatments10.1109/TRPMS.2019.2906842 ↗
31. Angular Independent Silicon Detector for Dosimetry in External Beam Radiotherapy10.1118/1.4926778 ↗
32. Rozenfeld, A.B. (2014). Radiation Sensor and Dosimeter (MOSkin). Australian Patent. PCT/AU2008/000788. (N 8,742,357 B2), US Patent, China Patent ZL 200880023328.8.
33. Multichannel Data Acquisition System Comparison for Quality Assurance in External Beam Radiation Therapy10.1016/j.radmeas.2014.05.016 ↗
34. Development of a Silicon Diode Detector for Skin Dosimetry in Radiotherapy10.1002/mp.12469 ↗
35. On the Effective Point of Measurement in Megavoltage Photon Beams10.1118/1.2201147 ↗
36. IAEA (2024). Absorbed Dose Determination in External Beam Radiotherapy an International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, IAEA Safety Standards and Related Publications.
37. Analytic and Graphical Methods for Assigning Errors to Parameters in Non-Linear Least Squares Fitting10.1016/0029-554X(75)90496-6 ↗
38. The Effect of Dose Rate Dependence of P-Type Silicon Detectors on Linac Relative Dosimetry10.1118/1.597985 ↗
39. The Use of a Silicon Strip Detector Dose Magnifying Glass in Stereotactic Radiotherapy QA and Dosimetry10.1118/1.3549759 ↗
40. Alhujaili, S. (2019). Development and Characterization of Solid-State Detectors for Medical Dosimetry in Intensity Modulated Radiation Therapy and Stereotactic Radiosurgery. [Ph.D. Thesis, University of Wollongong].
41. MagicPlate-512: A 2D Silicon Detector Array for Quality Assurance of Stereotactic Motion Adaptive Radiotherapy10.1118/1.4921126 ↗
42. Dosimetric Characteristics of a 2D Silicon Diode Array for Stereotactic Radiotherapy End-to-End Patient-Specific QA10.1016/j.radphyschem.2020.108885 ↗
43. Characterization of a Two-Dimensional Liquid-Filled Ion Chamber Detector Array Used for Verification of the Treatments in Radiotherapy10.1118/1.4870439 ↗
44. High Resolution Ion Chamber Array Delivery Quality Assurance for Robotic Radiosurgery: Commissioning and Validation10.1016/j.ejmp.2016.05.060 ↗
45. Multi-Institution Validation of a New High Spatial Resolution Diode Array for SRS and SBRT Plan Pretreatment Quality Assurance10.1002/mp.14153 ↗
46. Looe, H.K., Harder, D., and Poppe, B. (2011). Experimental Determination of the Effective Point of Measurement for Various Detectors Used in Photon and Electron Beam Dosimetry. Phys. Med. Biol., 56.10.1088/0031-9155/56/14/005 ↗
47. Small Field Output Correction Factors of the MicroSilicon Detector and a Deeper Understanding of Their Origin by Quantifying Perturbation Factors10.1002/mp.14149 ↗
48. On the Correction, Perturbation and Modification of Small Field Detectors in Relative Dosimetry10.1088/0031-9155/59/19/5937 ↗
49. Charles, P.H., Crowe, S.B., Kairn, T., Kenny, J., Lehmann, J., Lye, J., Dunn, L., Hill, B., Knight, R.T., and Langton, C.M. (2012). The Effect of Very Small Air Gaps on Small Field Dosimetry. Phys. Med. Biol., 57.10.1088/0031-9155/57/21/6947 ↗
50. Radiation Damage Induced Dose Rate Non-Linearity in an n-Type Silicon Detector
51. Rikner, G., and Grusell, E. (1983). Effects of Radiation Damage on P-Type Silicon Detectors. Phys. Med. Biol., 28.10.1088/0031-9155/28/11/006 ↗
52. Dose Rate and SDD Dependence of Commercially Available Diode Detectors10.1118/1.1650563 ↗
53. Rosenfeld, A.B., Biasi, G., Petasecca, M., Lerch, M.L.F., Villani, G., and Feygelman, V. (2020). Semiconductor Dosimetry in Modern External-Beam Radiation Therapy. Phys. Med. Biol., 65.10.1088/1361-6560/aba163 ↗