journal article·2024-04-14Open access

Development of Heart Simulator and Analysis of Valve Dysfunction in Tricuspid Regurgitation

Hwan Yi Joo, Seokyoung Ahn, Su-Jin Jung, Min-Ku Chon

Applied Sciences· Volume 14· Issue 8

Abstract

Tricuspid regurgitation (TR) resulting from valve abnormalities necessitates precise diagnostic tools and interventions. We employed a simulated heart movement device to examine the performance of heart valve functions and analyzed internal pressure changes to provide a quantitative guide for TR treatment. We developed a simulator capable of replicating the flow profile, mimicking natural heart movements, with sensors installed for measuring internal pressure changes. We conducted an ex vivo experiment on a porcine heart to assess tricuspid valve functionality. An endoscope was installed, with a sensor and endoscopic images to detect abnormalities. TR became evident when the heart rate spectrum exceeded an average of 85.2 bpm (standard deviation, 1.3 bpm) and showed an amplitude higher than an average of 12.3 mmHg (standard deviation, 3.2 mmHg). This critical threshold consistently indicated TR onset. The application of the Pivot-TR attenuated this specific spectral area. We confirmed TR disappearance by reducing the intensity of the Tricuspid Regurgitation Generator or employing the Pivot-TR. The Pivot-TR’s ability to attenuate specific spectral areas associated with TR onset and its effectiveness in restoring normal heart functionality has implications for managing and treating TR, particularly that resulting from age-related structural changes in the heart.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Cardiac Valve Diseases and Treatments · Cardiovascular Function and Risk Factors · Infective Endocarditis Diagnosis and Management

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Mechanical Engineering, Pusan National University, Busan 46241, Republic of Korea

Department of Research Strategy Team, R&D Center, TAU MEDICAL Inc., Yangsan 50612, Republic of Korea

Department of Internal Medicine, School of Medicine, Pusan National University, Busan 46241, Republic of Korea

Department of Cardiology, Cardiovascular Center, Yangsan Hospital, Yangsan 50612, Republic of Korea

Funding

Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital

References (18)

Show source-provided references
1. Tricuspid regurgitation is a public health crisis10.1016/j.pcad.2019.10.009 ↗
2. Tricuspid valve surgery: The past 10 years from the Nationwide Inpatient Sample (NIS) database10.1016/j.jtcvs.2011.07.004 ↗
3. Clinical and echocardiographic outcomes after surgery for severe isolated tricuspid regurgitation10.1016/j.jtcvs.2012.04.019 ↗
4. Geometric determinants of functional tricuspid regurgitation: Insights from 3-dimensional echocardiography10.1161/CIRCULATIONAHA.106.611889 ↗
5. Right ventricular remodeling determines tricuspid valve geometry and the severity of functional tricuspid regurgitation: A real-time 3-dimensional echocardiography study10.4070/kcj.2010.40.9.448 ↗
6. Lee, C.H., Laurence, D.W., Ross, C.J., Kramer, K.E., Babu, A.R., Johnson, E.L., Hsu, M.C., Aggarwal, A., Mir, A., and Burkhart, H.M. (2019). Mechanics of the tricuspid valve-from clinical diagnosis/treatment, in-vivo, and in-vitro investigations, to patient-specific biomechanical modeling. Bioengineering, 6.10.3390/bioengineering6020047 ↗
7. Assessment of functional tricuspid regurgitation10.1093/eurheartj/ehs474 ↗
8. Anwar, A.M., Folkert, J., and Soliman, O.I. (2018). Practical Manual of Tricuspid Valve Diseases, Springer.10.1007/978-3-319-58229-0_2 ↗
9. A novel device for tricuspid regurgitation reduction featuring 3-dimensional leaflet and atraumatic anchor: Pivot-TR system10.1016/j.jacbts.2022.06.017 ↗
10. Transcatheter tricuspid valve interventions: Landscape, challenges, and future directions10.1016/j.jacc.2018.04.031 ↗
11. Tricuspid valve regurgitation decreases after MitraClip implantation: Fluid structure interaction simulation10.1016/j.mechrescom.2019.04.009 ↗
12. Tricuspid valve tethering predicts residual tricuspid regurgitation after tricuspid annuloplasty10.1161/01.CIR.0000156449.49998.51 ↗
13. Predictors of residual tricuspid regurgitation after mitral valve surgery10.1016/S0003-4975(03)00028-6 ↗
14. Transcatheter Tricuspid Valve Repair With a New Transcatheter Coaptation System for the Treatment of Severe Tricuspid Regurgitation: 1-Year Clinical and Echocardiographic Results
15. Impact of atrial fibrillation on clinical status, atrial size and hemodynamics in patients after mitral valve replacement
16. Tricuspid valve repair with an annuloplasty ring results in improved long-term outcomes10.1161/CIRCULATIONAHA.105.001263 ↗
17. Long-Term Follow-Up of Rheumatic Patients Undergoing Left-Sided Valve Replacement With Tricuspid Annuloplasty—Validity of Preoperative Echocardiographic Criteria in the Decision to Perform Tricuspid Annuloplasty10.1016/S0002-9149(98)00081-2 ↗
18. Physiological and Clinical Consequences of Right Ventricular Volume Overload Reduction After Transcatheter Treatment for Tricuspid Regurgitation