journal article·2023-08-04Open access

Surface Corrosion from Implant–Abutment Couplings with Different Connection Designs Influences Osteoblasts’ Function: A Novel Technique

Ghada Alrabeah, Jonathan C. Knowles, Haralampos Petridis

Applied Sciences· Volume 13· Issue 15

Abstract

The improved peri-implant bone response demonstrated when utilizing the platform-switching concept may result from the reduced levels of metal ions released from implant–abutment surfaces to the surrounding tissues. These corrosion products may play a major role in crestal bone remodeling around dental implants. This study evaluated the effect of different implant–abutment couplings (platform-matched vs. platform-switched) on osteoblasts’ function. Titanium alloy and cobalt–chrome alloy abutments were coupled with titanium cylinders, forming either platform-switched or platform-matched groups, and were incubated in human osteoblast cultures utilizing a novel direct-exposure technique. Viability was evaluated over 21 days using Alamar Blue assay. Apoptosis was measured after 24 h using flow cytometry. The expression of genes related to bone resorption was analysed over 21 days using a real-time quantitative polymerase chain reaction assay. Cell viability was reduced from day 4 to day 21 (p < 0.05), with higher rates of early apoptosis (p < 0.05) compared to the controls. Apoptosis was higher in the platform-matched groups (p < 0.05). The tested genes’ expression was up-regulated after 1 and 3 days of exposure to implant–abutment couplings (p < 0.05). The upregulation was more pronounced in platform-matched groups (p < 0.05). Exposure of osteoblasts to implant–abutment couplings induced adverse biological responses, which were more pronounced with platform-matched couplings. These reactions might be related to the increased amounts of metal ions released from the platform-matched couplings, highlighting the possible role of corrosion products in the mediation of crestal bone loss around dental implants.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials · Dental Implant Techniques and Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Prosthetic Dental Sciences, College of Dentistry, King Saud University, Riyadh 11451, Saudi Arabia

Prosthodontic Unit, Department of Restorative Dentistry, UCL Eastman Dental Institute, University College London, London WC1E 6DE, UK

Division of Biomaterials and Tissue Engineering, UCL Eastman Dental Institute, University College London, London NW3 2PF, UK

Funding

King Saud University, Saudi Arabia

References (48)

Show source-provided references
1. Peri-implant health and disease. A systematic review of current epidemiology10.1111/jcpe.12334 ↗
2. Thanissorn, C., Guo, J., Jing Ying Chan, D., Koyi, B., Kujan, O., Khzam, N., and Miranda, L.A. (2022). Success rates and complications associated with single immediate implants: A systematic review. Dent. J., 10.10.3390/dj10020031 ↗
3. Bragger, U., and Heitz-Mayfield, L. (2015). ITI Treatment Guide, Quintessence Publishing Co., Ltd.
4. Peri-implant health, peri-implant mucositis, and peri-implantitis: Case definitions and diagnostic considerations10.1111/jcpe.12956 ↗
5. The long-term efficacy of currently used dental implants: A review and proposed criteria of success
6. Influence of the size of the microgap on crestal bone changes around titanium implants. A histometric evaluation of unloaded non-submerged implants in the canine mandible10.1902/jop.2001.72.10.1372 ↗
7. Platform switching: A new concept in implant dentistry for controlling postrestorative crestal bone levels
8. The effect of platform switching on the levels of metal ion release from different implant–abutment couples10.1038/ijos.2016.5 ↗
9. Reduction of tribocorrosion products when using the platform-switching concept10.1177/0022034518765751 ↗
10. Biologic effects of implant debris
11. The potential role of the osteoblast in the development of periprosthetic osteolysis: Review of in vitro osteoblast responses to wear debris, corrosion products, and cytokines and growth factors10.1054/arth.2001.28719 ↗
12. The role of osteoblasts in peri-prosthetic osteolysis
13. Is marginal bone loss around oral implants the result of a provoked foreign body reaction?10.1111/cid.12142 ↗
14. Foreign body reactions, marginal bone loss and allergies in relation to titanium implants
15. What is the impact of titanium particles and biocorrosion on implant survival and complications? A critical review10.1111/clr.13305 ↗
16. Is metal particle release associated with peri-implant bone destruction? An emerging concept10.1177/0022034517740560 ↗
17. Can degradation products released from dental implants affect peri-implant tissues?10.1111/jre.12479 ↗
18. Bone loss around implants—Is it metallosis?10.1002/JPER.20-0208 ↗
19. Peri-Implantitis in Relation to Titanium Corrosion: Current Status and Future Perspectives10.1007/s40735-022-00644-6 ↗
20. The effect of metal ions released from different dental implant-abutment couples on osteoblast function and secretion of bone resorbing mediators10.1016/j.jdent.2017.08.002 ↗
21. Peri-implant diseases and conditions: Consensus report of workgroup 4 of the 2017 World Workshop on the Classification of Periodontal and Peri-Implant Diseases and Conditions10.1002/JPER.17-0739 ↗
22. Electrical implications of corrosion for osseointegration of titanium implants10.1177/0022034511408428 ↗
23. The effects of proteins on metallic corrosion10.1002/jbm.820160205 ↗
24. (2009). Biological Evaluation of Medical Devices-Part 5: Tests for In Vitro Cytotoxicity (Standard No. ISO 10993-5: 2009).
25. Analysis of relative gene expression data using real-time quantitative PCR and the 2−ΔΔCT method10.1006/meth.2001.1262 ↗
26. Predicting clinical biological responses to dental materials10.1016/j.dental.2011.08.595 ↗
27. Biological evaluation of dental restorative materials—A comparison of different test methods10.1002/jbm.820170103 ↗
28. Modulation of immune-inflammatory responses through surface modifications of biomaterials to promote bone healing and regeneration10.1177/20417314211041428 ↗
29. The effects of the coupling of titanium implants and dissimilar metal abutments on osteoblast differentiation in vitro
30. Systemic metal–protein binding associated with total joint replacement arthroplasty10.1002/(SICI)1097-4636(20000305)49:3<353::AID-JBM8>3.0.CO;2-T ↗
31. Cellular responses to cobalt-chrome and CP titanium--an in vitro comparison of frameworks for implant-retained oral prostheses
32. Concentration-and composition-dependent effects of metal ions on human MG-63 osteoblasts10.1002/jbm.10106 ↗
33. The potential role of human osteoblasts for periprosthetic osteolysis following exposure to wear particles
34. Orthopedic implant cobalt-alloy particles produce greater toxicity and inflammatory cytokines than titanium alloy and zirconium alloy-based particles in vitro, in human osteoblasts, fibroblasts, and macrophages10.1002/jbm.a.34122 ↗
35. Osteoblast programmed cell death (apoptosis): Modulation by growth factors and cytokines10.1359/jbmr.1998.13.5.793 ↗
36. Gene expression analysis of osteoblastic cells contacted by orthopedic implant particles10.1002/jbm.10218 ↗
37. Induction of interleukin-6 release in human osteoblast-like cells exposed to titanium particles in vitro10.1007/s00223001125 ↗
38. The effects of biomaterial implant wear debris on osteoblasts10.3389/fcell.2020.00352 ↗
39. The interaction of osteoblasts with bone-implant materials: 1. The effect of physicochemical surface properties of implant materials10.33549/physiolres.931882 ↗
40. Dental implants stimulate expression of Interleukin-8 and its receptor in human blood—An in vitro approach10.1002/jbm.b.32693 ↗
41. Chemokine gene activation in human bone marrow-derived osteoblasts following exposure to particulate wear debris10.1002/jbm.a.30609 ↗
42. Cobalt ions induce chemokine secretion in primary human osteoblasts10.1002/jor.20837 ↗
43. COX-1 and COX-2 tissue expression: Implications and predictions
44. The release of elements of dental casting alloys into cell-culture medium10.1177/00220345910700060301 ↗
45. Correlation between cytotoxicity and the elements released by dental casting alloys
46. Biologic Width around one-and two-piece titanium implants: A histometric evaluation of unloaded nonsubmerged and submerged implants in the canine mandible10.1034/j.1600-0501.2001.120603.x ↗
47. The microbiota associated with implants restored with platform switching: A preliminary report10.1902/jop.2009.090498 ↗
48. Biomechanical analysis on platform switching: Is there any biomechanical rationale?10.1111/j.1600-0501.2007.01398.x ↗