Experimental and Geant4 Simulation Study of MURA Mask for Scintimammography
Source abstract
Abstract
Gamma rays have been extensively investigated for breast imaging using collimators; however, the coded-aperture technique needs to be investigated more. In this paper, we propose an experimental study and Geant4 simulations of MURA mask breast imaging. First, we compare the experimental data against the simulation results carried out using Geant4 (version 10.4) and accreditation phantom. Second, we virtually extend our work by changing the tumor-to-background (TBR) and lesion location parameters. We used 99mTc as a radioactive source. Good agreement has been seen for the benchmark stage, especially in terms of tumor localization. Moreover, the calculated full width at half maximum (FWHM) and contrast for decoded images (having average values of 8 and 3.5 for TBR between 2 and 10) permitted us to conclude that we can accurately localize small lesions up to lower TBR values by following the decoding procedure of deducing the image of a “blank phantom” (phantom with TBR = 1) every time within a matlab-based program. Hence, this work can be considered a continuously added value to previous investigations for scintimammography imaging.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Medical Imaging Techniques and Applications · Radiation Detection and Scintillator Technologies · Advanced Radiotherapy Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Radiological Sciences, College of Applied Medical Sciences, King Saud University, P.O. Box 10219, Riyadh 11433, Saudi Arabia
Department of Physics and Astronomy, College of Science, King Saud University, Riyadh 11451, Saudi Arabia
National Center for Nuclear Sciences and Technologies, Tunis 2020, Tunisia
References (26)
Show source-provided references
1. Radiation protection, radioactive waste management and site monitoring at the nuclear scientific experimental and educational centre IRT-Sofia at INRNE-BAS10.1093/rpd/ncu254 ↗
2. Ambrosino, F., Thinova, L., Briestensky, M., Giudicepietro, F., Roca, V., and Sabbarese, C. (2020). Analysis of geophysical and meteorological parameters influencing 222Rn activity concentration in Mladec caves (Czech Republic) and in soils of Phlegrean Fields caldera (Italy). Appl. Radiat. Isot., 160.10.1016/j.apradiso.2020.109140 ↗
3. Zhang, K., Zeng, Y., Liu, Z., Cao, H., and Zhu, W. (2019). Investigation on the radiation decontamination of lanthanum oxide during its production from ion-adsorption rare earth ores in China. Hydrometallurgy, 189.10.1016/j.hydromet.2019.105117 ↗
4. Role of scintimammography in the diagnosis of breast cancer
5. A versatile calibration procedure for portable coded aperture gamma cameras and RGB-D sensors10.1016/j.nima.2017.12.065 ↗
6. Amgarou, K., Paradiso, V., Patoz, A., Bonnet, F., Handley, J., Couturier, P., Becker, F., and Menaa, N. (2016). A comprehensive experimental characterization of the iPIX gamma imager. J. Instrum., 11.10.1088/1748-0221/11/08/P08012 ↗
7. Development of a panorama coded-aperture gamma camera for radiation detection10.1016/j.radmeas.2015.04.014 ↗
8. Far field 3D localization of radioactive hot spots using a coded aperture camera10.1016/j.apradiso.2015.10.021 ↗
9. ORIGAMIX, a CdTe-based spectro-imager development for nuclear applications, Nucl10.1016/j.nima.2014.12.111 ↗
10. Reconstruction method for gamma-ray coded-aperture imaging based on convolutional neural network10.1016/j.nima.2019.04.055 ↗
11. Kadri, O., and Alfuraih, A. (2019). Monte Carlo assessment of coded aperture tool for breast imaging: A Mura-mask case study. Nucl. Sci. Tech., 30.10.1007/s41365-019-0682-3 ↗
12. Zhang, T., Wang, L., Ning, J., Lu, W., Wang, X.F., Zhang, H.W., and Tuo, X.G. (2021). Simulation of an imaging system for internal contamination of lungs using MPA-MURA coded aperture collimator. arXiv.10.1007/s41365-021-00849-3 ↗
13. Telescope performance and image simulations of the balloon-borne coded-mask protoMIRAX experiment10.1016/j.jheap.2015.01.001 ↗
14. Near-field artifact reduction in planar coded aperture Imaging10.1364/AO.40.004697 ↗
15. LECX: A cubesat experiment to detect and localize cosmic explosions in hard X-rays10.1093/mnras/staa500 ↗
16. Kasuga, T., Odaka, H., Hatauchi, K., Takashima, S., Tamba, T., Aizawa, Y., Hashiba, S., Bamba, A., Zhou, Y., and Tamagawa, T. (2020). Artifact-less Coded Aperture Imaging in the X-ray Band with Multiple Different Random Patterns. arXiv.10.1117/1.JATIS.6.3.035002 ↗
17. Jeong, M., and Hammig, M.D. (2018). Comparison of gamma ray localization using system matrixes obtained by either MCNP simulations or ray-driven calculations for a coded-aperture imaging system. Nucl. Instrum. Methods A, 954.10.1016/j.nima.2018.10.031 ↗
18. Sun, S., Liu, Y., and Ouyang, X. (2020). Near-field high-resolution coded aperture gamma-ray imaging with separable masks. Nucl. Instrum. Methods A, 951.10.1016/j.nima.2019.163001 ↗
19. Preliminary Monte Carlo study of coded aperture imaging with a CZT gamma camera system for scintimammography10.1016/j.nima.2006.11.007 ↗
20. GEANT4-a simulation toolkit10.1016/S0168-9002(03)01368-8 ↗
21. Geant4 physics list comparison for the simulation of phase-contrast mammography (XPulse project)10.1016/j.ejmp.2019.03.026 ↗
22. Rinaldi, L., Ambrosino, F., Roca, V., D’Onofrio, A., and Sabbarese, C. (2022). Study of 222-220Rn Measurement Systems Based on Electrostatic Collection by Using Geant4+COMSOL Simulation. Appl. Sci., 12.10.3390/app12010507 ↗
23. Berger, M.J., Coursey, J.S., Zucker, M.A., and Chang, J. (2017). Stopping-Power & Range Tables for Electrons, Protons and Helium Ions, National Institute of Standards and Technology. NIST Standard Reference Database 124 2017, NISTIR 4999.
24. Recent developments in GEANT410.1016/j.nima.2016.06.125 ↗
25. Geant4-towards major release 1010.1088/1742-6596/513/2/022005 ↗
26. Coded-aperture imaging systems: Past, present and future development—A review10.1016/j.radmeas.2016.08.002 ↗