journal article·2021-10-29Open access

Bladeless Heart Pump Design: Modeling and Numerical Analysis

Vedrana Markučič, Mario Šavar, Severino Krizmanić, Bojan Biočina

Applied Sciences· Volume 11· Issue 21

Abstract

In this paper, the design of a new type of heart pump is presented. The uniqueness of this pump concept is that the mechanical energy is transferred to the fluid by the rotation of flat disks without blades. Both theoretical and numerical analysis are used to determine the pump design parameters. The pump design parameters are calculated using the Navier-Stokes equations. The application of computational fluid dynamics is used to define the geometric design. The pump head and flow have to be within strictly defined limits to ensure normal blood circulation. The negative impact of the pump on the blood must be minimal (no stagnation and recirculation zones, shear stress in the acceptable range). It is also important to achieve the smallest possible pump volume. For the pump operating point of ∆p = 65 mmHg, Q = 5.43 L/min, and ω = 6000 rpm, the design parameters are inner radius R1 = 12 mm, outer radius R2 = 15 mm, the distance between disk of h = 1 mm, and the number of disks is n = 6. The shear stress in the rotor is in range 46–108 Pa and the pump residence time is 0.0194 s.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Mechanical Circulatory Support Devices · Fuel Cells and Related Materials · Cardiac Structural Anomalies and Repair

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Fluid Mechanics, Faculty of Mechanical Engineering and Naval Architecture, University of Zagreb, Ivana Lučića 5, 10000 Zagreb, Croatia

Department of Cardiac Surgery, University Hospital Centre Zagreb, School of Medicine, University of Zagreb, 10000 Zagreb, Croatia

References (39)

Show source-provided references
1. Global epidemiology and future trends of heart failure10.21037/amj.2020.03.03 ↗
2. Progress in mechanical circulatory support: Challenges and op-portunities10.1111/aor.13500 ↗
3. Left ventricular assist devices—Current state and perspectives10.21037/jtd.2016.07.13 ↗
4. Left Ventricular Assist Devices: Chal-lenges Toward Sustaining Long-Term Patient Care10.1007/s10439-017-1858-9 ↗
5. Total artificial hearts: Past, present, and future10.1038/nrcardio.2015.79 ↗
6. Superior performance of continuous over pulsatile flow ventricular assist devices in the single ventricle circulation: A computational study10.1016/j.jbiomech.2016.12.003 ↗
7. Is Continuous Flow Superior to Pulsatile Flow in Single Ventricle Mechanical Support? Results from a Large Animal Pilot Study10.1097/MAT.0000000000000220 ↗
8. Third Annual Report from the ISHLT Mechanically As-sisted Circulatory Support Registry: A comparison of centrifugal and axial continuous-flow left ventricular assist devices10.1016/j.healun.2019.02.004 ↗
9. Kannojiya, V., Das, A.K., and Das, P.K. (2021). Comparative assessment of different versions of axial and centrifugal LVADs: A review. Artif. Organs.10.1111/aor.13914 ↗
10. Hemocompatibility of Axial Versus Cen-trifugal Pump Technology in Mechanical Circulatory Support Devices10.1111/aor.12544 ↗
11. Seventh INTERMACS annual report: 15,000 patients and counting10.1016/j.healun.2015.10.003 ↗
12. Results With Syncardia Total Artificial Heart Beyond 1 Year10.1097/MAT.0000000000000132 ↗
13. Wang, Y., Karnik, S., Smith, P.A., Elgalad, A., Frazier, O.H., and Kurita, N. (2020, January 20–24). Numerical and Experimental Approach to Characterize a BLDC Motor with Different Radial-gap to Improve Hemocompatibility Performance. Proceedings of the 2020 42nd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), Montreal, QC, Canada.10.1109/EMBC44109.2020.9175989 ↗
14. The Spiral Groove Bearing as a Mechanism for Enhancing the Secondary Flow in a Centrifugal Rotary Blood Pump10.1111/aor.12081 ↗
15. In Vitro Assessment of the Apico Aortic Blood Pump: Anatomical Positioning, Hydrodynamic Performance, Hemolysis Studies, and Analysis in a Hybrid Cardiovascular Simulator10.1111/aor.12229 ↗
16. In vitro hemocompatibility testing of biomaterials according to the ISO 10993-410.1016/S1389-0344(02)00015-1 ↗
17. A critical review on blood flow in large arteries; relevance to blood rheology, viscosity models, and physiologic conditions
18. Effects of the non-Newtonian viscosity of blood on flows in a diseased arterial vessel. Part 1: Steady flows10.3233/BIR-1991-283-415 ↗
19. Tesla-Based Blood Pump and Its Applications10.1115/1.4025982 ↗
20. Horvat, M. (2015). Heart Pump h-Q Characteristic Calculation. [Bachelor’s Thesis, Faculty of Mechanical Engineering and Naval Architecture, University of Zagreb].
21. Optimization parameters of the heart pump design10.1088/1757-899X/393/1/012125 ↗
22. Heng, Y., Han, Y., Zhang, H., Zhang, W., Bois, G., Jiang, Q., Wang, Z., and Liu, X. (2020). Tesla Bladed Pump (Disc Bladed Pump) Preliminary Experimental Performance Analysis. Energies, 13.10.3390/en13184873 ↗
23. Habhab, M.-B., Ismail, T., and Lo, J. (2016). A Laminar Flow-Based Microfluidic Tesla Pump via Lithography Enabled 3D Printing. Sensors, 16.10.3390/s16111970 ↗
24. Joyce, D.L., and Joyce, L.D. (2020). Mechanical Circulatory Support: Principles and Applications [Internet], Oxford University Press.10.1093/med/9780190909291.001.0001 ↗
25. A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis Index10.1115/1.4007092 ↗
26. Hemolysis in a Laminar Flow-Through Couette Shearing Device: An Experimental Study
27. Device hemocompatibility: New frontiers
28. Factors Related to Pump Thrombosis with the Heartmate II Left Ventricular Assist Device10.1111/jocs.12607 ↗
29. Continuous-Flow Left Ventricular Assist Device Thrombosis: A Solvable Problem10.1097/MAT.0000000000000319 ↗
30. Controlling critical Reynolds number of plate flow by boundary input parallel10.1108/00022660710780632 ↗
31. Schobeiri, M.T. (2010). Fluid Mechanics for Engineers, Springer Science and Business Media LLC.10.1007/978-3-642-11594-3 ↗
32. Numerical investigation of non-Newtonian blood flow within an artery with cone shape of stenosis in various stenosis angles10.1016/j.cmpb.2020.105434 ↗
33. Katavić, J. (2016). Improvement of Flow Pattern in Friction Pump. [Bachelor’s Thesis, Faculty of Mechanical Engineering and Naval Architecture, University of Zagreb].
34. Shear Stress-Induced Total Blood Trauma in Multiple Species10.1111/aor.12932 ↗
35. Model for a General Mechanical Blood Damage Prediction10.1111/j.1525-1594.1995.tb02385.x ↗
36. Clinical Courses of HeartMate II Left Ventricular Assist Device Thrombosis10.1097/MAT.0000000000000952 ↗
37. Experience with the HeartMate II Left Ventricular Assist Device in Patients Older than 60 Years10.1532/hsf.2297 ↗
38. Pump thrombosis in the Thoratec HeartMate II device: An update analysis of the INTERMACS Registry10.1016/j.healun.2015.10.024 ↗
39. Experimental and numerical study on convective boiling in a staggered array of micro pin-fin microgap10.1016/j.ijheatmasstransfer.2019.119203 ↗