In Vitro Estimation of Relative Compliance during High-Frequency Oscillatory Ventilation
Source abstract
Abstract
High-frequency oscillatory ventilation (HFOV), which uses a small tidal volume and a high respiratory rate, is considered a type of protective lung ventilation that can be beneficial for certain patients. A disadvantage of HFOV is its limited monitoring of lung mechanics, which complicates its settings and optimal adjustment. Recent studies have shown that respiratory system reactance (Xrs) could be a promising parameter in the evaluation of respiratory system mechanics in HFOV. The aim of this study was to verify in vitro that a change in respiratory system mechanics during HFOV can be monitored by evaluating Xrs. We built an experimental system consisting of a 3100B high-frequency oscillatory ventilator, a physical model of the respiratory system with constant compliance, and a system for pressure and flow measurements. During the experiment, models of different constant compliance were connected to HFOV, and Xrs was derived from the impedance of the physical model that was calculated from the spectral density of airway opening pressure and spectral cross-power density of gas flow and airway opening pressure. The calculated Xrs changed with the change of compliance of the physical model of the respiratory system. This method enabled monitoring of the trend in the respiratory system compliance during HFOV, and has the potential to optimize the mean pressure setting in HFOV in clinical practice.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Respiratory Support and Mechanisms · Neonatal Respiratory Health Research · Inhalation and Respiratory Drug Delivery
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, nam. Sitna 3105, CZ-272 01 Kladno, Czech Republic
Funding
České Vysoké Učení Technické v Praze
References (47)
Show source-provided references
1. Severity of Hypoxemia and Effect of High-Frequency Oscillatory Ventilation in Acute Respiratory Distress Syndrome10.1164/rccm.201609-1938OC ↗
2. High frequency oscillation in patients with acute lung injury and acute respiratory distress syndrome (ARDS): Systematic review and meta-analysis10.1136/bmj.c2327 ↗
3. High-Frequency Oscillatory Ventilation in Adults With ARDS: Past, Present, and Future10.1016/j.chest.2017.06.025 ↗
4. High-frequency oscillatory ventilation: Still a role?10.1097/MCC.0000000000000387 ↗
5. High-Frequency Oscillation for Adult Patients with Acute Respiratory Distress Syndrome. A Systematic Review and Meta-Analysis10.1513/AnnalsATS.201704-341OT ↗
6. The impact of high frequency oscillatory ventilation on mortality in paediatric acute respiratory distress syndrome10.1186/s13054-020-2741-x ↗
7. The harm of high-frequency oscillatory ventilation (HFOV) in ARDS is not related to a high baseline risk of acute cor pulmonale or short-term changes in hemodynamics10.1007/s00134-019-05806-8 ↗
8. The effect of high-frequency oscillatory ventilation or airway pressure release ventilation on children with acute respiratory distress syndrome as a rescue therapy10.21037/tp-19-178 ↗
9. High-frequency oscillation in early acute respiratory distress syndrome10.1056/NEJMoa1215554 ↗
10. High-frequency oscillation for acute respiratory distress syndrome10.1056/NEJMoa1215716 ↗
11. Is high-frequency oscillatory ventilation more effective and safer than conventional protective ventilation in adult acute respiratory distress syndrome patients? A meta-analysis of randomized controlled trials10.1186/cc13900 ↗
12. High-frequency ventilation does not provide mortality benefit in comparison with conventional lung-protective ventilation in acute respiratory distress syndrome: A meta-analysis of the randomized controlled trials10.1097/ALN.0000000000000306 ↗
13. Feasibility of an alternative, physiologic, individualized open-lung approach to high-frequency oscillatory ventilation in children10.1186/s13613-019-0492-0 ↗
14. Physiologic responses to a staircase lung volume optimization maneuver in pediatric high-frequency oscillatory ventilation10.1186/s13613-020-00771-8 ↗
15. Optimal mean airway pressure during high-frequency oscillatory ventilation in an experimental model of acute respiratory distress syndrome: EIT-based method10.1186/s13613-020-0647-z ↗
16. Efficacy of lung volume optimization maneuver monitored by optoelectronic pletismography in the management of congenital diaphragmatic hernia
17. Effect of frequency on pressure cost of ventilation and gas exchange in newborns receiving high-frequency oscillatory ventilation10.1038/pr.2017.151 ↗
18. Do We Really Know How to Use High-Frequency Oscillatory Ventilation in Critically Ill Children?10.1164/rccm.201512-2418LE ↗
19. Any trial can (almost) kill a good technique10.1007/s00134-016-4215-9 ↗
20. Assessment of dynamic mechanical properties of the respiratory system during high-frequency oscillatory ventilation10.1097/CCM.0b013e31828cf3ea ↗
21. A method for determining optimal mean airway pressure in high-frequency oscillatory ventilation10.1007/s00408-012-9434-4 ↗
22. Oxygenation index, an indicator of optimal distending pressure during high-frequency oscillatory ventilation?10.1007/s00134-002-1368-5 ↗
23. Optimal mean airway pressure during high-frequency oscillation: Predicted by the pressure-volume curve10.1097/00000542-200105000-00026 ↗
24. Optimal high-frequency oscillatory ventilation settings by nonlinear lung mechanics analysis10.1164/rccm.200205-398OC ↗
25. High-frequency oscillatory ventilation guided by transpulmonary pressure in acute respiratory syndrome: An experimental study in pigs10.1186/s13054-018-2028-7 ↗
26. Setting mean airway pressure during high-frequency oscillatory ventilation according to the static pressure-volume curve in surfactant-deficient lung injury: A computed tomography study10.1097/00000542-200312000-00012 ↗
27. Indicators of optimal lung volume during high-frequency oscillatory ventilation in infants10.1097/CCM.0b013e31826a427a ↗
28. Optimal mean airway pressure during high-frequency oscillatory ventilation determined by measurement of respiratory system reactance10.1038/pr.2013.251 ↗
29. The effect of airway pressure and oscillation amplitude on ventilation in pre-term infants10.1183/09031936.00138311 ↗
30. Tidal volume, recruitment and compliance in HFOV: Same principles, different frequency10.1183/09031936.00020012 ↗
31. Kung, S.C., Hung, Y.L., Chen, W.L., Wang, C.M., Chang, H.C., and Liu, W.L. (2019). Effects of Stepwise Lung Recruitment Maneuvers in Patients with Early Acute Respiratory Distress Syndrome: A Prospective, Randomized, Controlled Trial. J. Clin. Med., 8.10.3390/jcm8020231 ↗
32. The Role of Airway Shunt Elastance on the Compartmentalization of Respiratory System Impedance
33. Measuring lung function using sound waves: Role of the forced oscillation technique and impulse oscillometry system10.1183/20734735.020514 ↗
34. Peripheral bronchial obstruction evaluation in patients with asthma by lung sound analysis and impulse oscillometry10.1016/j.alit.2016.06.008 ↗
35. Airway resistance and reactance are affected in systemic sclerosis10.3402/ecrj.v2.28667 ↗
36. Bhattarai, P., Myers, S., Chia, C., Weber, H.C., Young, S., Williams, A.D., and Sohal, S.S. (2020). Clinical Application of Forced Oscillation Technique (FOT) in Early Detection of Airway Changes in Smokers. J. Clin. Med., 9.10.3390/jcm9092778 ↗
37. A protocol for high-frequency oscillatory ventilation in adults: Results from a roundtable discussion10.1097/01.CCM.0000269026.40739.2E ↗
38. Roubik, K. (2014). Measuring and evaluating system designed for high-frequency oscillatory ventilation monitoring. Biomed. Tech.
39. The use of fast Fourier transform for the estimation of power spectra: A method based on time averaging over short, modified periodograms10.1109/TAU.1967.1161901 ↗
40. Pulmonary mechanics by spectral analysis of forced random noise10.1172/JCI108198 ↗
41. Damping of the dynamic pressure amplitude in the ventilatory circuit during high-frequency oscillatory ventilation
42. Models of PaO2 response to the continuous distending pressure maneuver during high frequency oscillatory ventilation in healthy and ARDS lung model pigs10.3109/01902148.2016.1145307 ↗
43. Optimum endexpiratory airway pressure in patients with acute pulmonary failure10.1056/NEJM197502062920604 ↗
44. Regional tidal ventilation and compliance during a stepwise vital capacity manoeuvre10.1007/s00134-010-1995-1 ↗
45. Trends in mechanical ventilation: Are we ventilating our patients in the best possible way?10.1183/20734735.007817 ↗
46. Design and control of a demand flow system assuring spontaneous breathing of a patient connected to an HFO ventilator10.1109/TBME.2011.2165541 ↗
47. Demand flow facilitates spontaneous breathing during high-frequency oscillatory ventilation in a pig model10.1097/CCM.0b013e318196153b ↗