Simultaneous Verification of Multi-Ultrasonic-Flowmeters in Tandem Arrangement for Natural Gas
Source abstract
Abstract
A combined experiment and numerical analysis was performed to determine whether the flow field has a significant impact on the simultaneous verification of multi-ultrasonic flowmeters (UFMs) in tandem arrangement (SVUFMST) and whether the SVUFMST is practical for flowrate measurement of pressurized natural gas. In practical testing, SVUFMST can highly improve the efficiency of UFMs’ calibration. Two types of UFMs of different nominal diameter were studied: DN250 and DN150. When it comes to DN250, data obtained from the actual-flow experiments were first used to verify the simulated results for n ≤ 3 (number of tested UFMs) and then the verified numerical method was extended to simulate the internal flow of four and five tested UFMs. Comparison analysis shows that all the numerical results agree remarkably well with the available experimental data and more than two tested UFMs can induce an overall shift in the E–Q curve (relative measurement error E plotted as a function of flowrate Q). For the slight difference among the tested UFMs, the practicability of the SVUFMST at n ≤ 4 (DN250) and n ≤ 3 (DN150) are thereby proved, whereas the striking difference in E (~1.0%) between the most downstream UFM and the other UFMs for n = 5 (DN250), associated with the outstanding collapse onto the velocity profile along the acoustic paths, results in the invalidation of the SVUFMST.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Flow Measurement and Analysis · Advanced Sensor Technologies Research · Electrical and Bioimpedance Tomography
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Zhejiang Provincial Key Laboratory of Flow Measurement Technology, China Jiliang University, Hangzhou 310018, China
Petrochina West-East Natural Gas Pipeline Company, Nanjing 210058, China
College of Control Science and Engineering, Zhejiang University, Hangzhou 310027, China
LEO Group Co., Ltd., Wenling 317500, China
Funding
S&T Projects of Petrochina West-East Natural Gas Pipeline Company · the National Natural Science Foundation of China
References (21)
Show source-provided references
1. Application of extreme learning machine to gas flow measurement with multipath acoustic transducers10.1016/j.flowmeasinst.2016.03.003 ↗
2. Calibration procedures and uncertainty analysis for a thermal mass gas flowmeter of a new generation10.1016/j.measurement.2016.03.073 ↗
3. Funck, B., and Baldwin, P. (2012). Challenges for Ultrasonic Flow Meters in Wet Gas Applications, FLEXIM GmbH.
4. West to East Gas Transmission Project and Development of China Natural Gass Industry
5. Xie, L. (2016). Study on Natural Gas Marketing Strategy of West to East Gas Pipeline Company, Nanjing University. (In Chinese).
6. National Technical Committee for Measurement of Flowrate and Capacity (China) (2007). JJG1030-2007 Verification Regulation of Ultrasonic Flowmeters.
7. (2019, September 03). Technical Committee ISO/TC 30, ISO 17089-1:2010(en) Measurement of Fluid Flow in Closed Conduits Ultrasonic Meters for Gas: Part 1: Meters for Custody Transfer and Allocation Measurement. Available online: https://www.iso.org/standard/41235.html.
8. Wan, Y. (2012). Feasibility analysis of simultaneous verification of gas ultrasonic flowmeter in tightly tandem arrangement. China Metrol., 119–121. (In Chinese).
9. Study of acoustic transducer protrusion and recess effects on ultrasonic flowmeter measurement by numerical simulation10.1016/j.flowmeasinst.2011.08.003 ↗
10. CFD Aided Investigation of Multipath Ultrasonic Gas Flow Meter Performance Under Complex Flow Profile10.1109/JSEN.2013.2290863 ↗
11. The influence of flattened flow profiles to the airflow velocity measurement by ultrasound, Ultragarsas
12. Effect of wall roughness changes on ultrasonic gas flowmeters10.1016/S0955-5986(01)00014-0 ↗
13. Effects of pulsating flow on an ultrasonic gas flowmeter10.1016/0955-5986(94)90042-6 ↗
14. Yunus, A.C., and Cimbala, J.M. (2006). Fluid Mechanics Fundamentals and Applications, International ed., McGraw Hill Publication.
15. Freysa, K.-E., and Lunde, P. (2001, January 28–31). A ray theory approach to investigate the influence of flow velocity profiles on transit times in ultrasonic flow meters for gas and liquid. Proceedings of the 24th Scandinavian Symposium on Physical Acoustics, Ustaoset, Norway.
16. Yeh, T.T., and Mattingly, G.E. (2000, January 4–8). Ultrasonic technology: Prospects for improving flow measurements and standards. Proceedings of the 10th IMEKO TC9 Conference on Flow Measurement (FLOMEKO 2000), Salvador, Brazil.
17. Investigation of the flow velocity profile in a metering section of an invasive ultrasonic flowmeter10.1016/j.flowmeasinst.2005.11.004 ↗
18. Zanker, K. (2003, January 25–28). Diagnostic Ability of the Daniel Four-Path Ultrasonic Flow Meter. Proceedings of the 2nd International South East Asia Hydrocarbon Flow Measurement Workshop, Kuala Lumpur, Malaysia.
19. Zanker, K. (2000, January 4–8). Installation effects on single and multi-path ultrasonic meters. Proceedings of the Flomeko 2000—IMEKO TC9 Conference, Salvador, Brazil.
20. Ultrasonic transit-time flowmeters modelled with theoretical velocity profiles: Methodology10.1088/0957-0233/11/12/321 ↗
21. Actual flow calibration of a feedwater flowmeter using a high Reynolds number facility at NMIJ10.1016/j.nucengdes.2009.03.009 ↗