Self-Isolated Dual-Mode High-Pass Birdcage RF Coil for Proton and Sodium MR Imaging at 7 T MRI
Source abstract
Abstract
This study presents the feasibility of a dual-mode high-pass birdcage RF coil to acquire MR images at both 1H and 23Na frequencies at ultra-high-field MR scanner, 7 T. A dual-mode circuit (DMC) in the dual-mode birdcage (DMBC) RF coil operates at two frequencies, addressing the limitations of sensitivity reduction and isolation between two frequencies as in traditional dual-tuned RF coil. Finite-difference time-domain (FDTD) based electromagnetic (EM) simulations were performed to verify the RF coil at each frequency on the three-dimensional human head model. The DMBC RF coil resonated at proton (1H) and sodium (23Na) frequencies, and also single-tuned high-pass birdcage RF coils were constructed for both 1H and 23Na frequencies. The bench test performance of the RF coils was evaluated using network analysis parameters, including the measurement of scattering parameters (S-parameters) and quality factors (Q-factors). Q-factor of the DMBC coil at 1H port was 10.2% lower than that of 1H single-tuned birdcage (STBC) coil, with a modest SNR reduction of 6.5%. Similarly, the Q-factor for the DMBC coil at 23Na port was 12.3% less than that of 23Na STBC coil, and the SNR showed a minimal reduction of 5.4%. Utilizing the DMBC coil, promising 1H and 23Na MR images were acquired compared to those by using STBC coils. In conclusion, deploying a DMBC 1H/23Na coil has been demonstrated to overcome traditional constraints associated with dual-tuned RF coils, achieving this with only nominal signal attenuation across both nuclei operational frequencies.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced MRI Techniques and Applications · Advanced NMR Techniques and Applications · Atomic and Subatomic Physics Research
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Electronics and Information Engineering, Korea University, Sejong 30019, Republic of Korea
Research Institute for Advanced Industrial Technology, Korea University, Sejong 30019, Republic of Korea
Bio-Chemical Analysis Team, Korea Basic Science Institute, Cheongju 28119, Republic of Korea
Department of Radiology, University of Pittsburgh, Pittsburgh, PA 15213, USA
Funding
Institute of Information and Communications Technology Planning and Evaluation (IITP)
References (51)
Show source-provided references
1. Elevated tissue sodium concentration in malignant breast lesions detected with non-invasive 23Na MRI10.1007/s10549-006-9485-4 ↗
2. 23Na-MRI as a Noninvasive Biomarker for Cancer Diagnosis and Prognosis10.1002/jmri.27147 ↗
3. Sodium imaging intensity increases with time after human ischemic stroke10.1002/ana.21648 ↗
4. Sodium MRI and the Assessment of Irreversible Tissue Damage during Hyper-Acute Stroke10.1007/s12975-012-0168-7 ↗
5. Quantification of tissue sodium concentration in the ischemic stroke: A comparison between external and internal references for 23Na MRI10.1016/j.jneumeth.2022.109721 ↗
6. Combining sodium MRI, proton MR spectroscopic imaging, and intracerebral EEG in epilepsy10.1002/hbm.26102 ↗
7. Sodium MRI of multiple sclerosis10.1002/nbm.3289 ↗
8. Diffusely appearing white matter in multiple sclerosis: Insights from sodium (23Na) MRI10.1016/j.msard.2021.102752 ↗
9. Evaluation of Sodium (23Na) MR-imaging as a Biomarker and Predictor for Neurodegenerative Changes in Patients with Alzheimer’s Disease10.21873/invivo.12275 ↗
10. Measuring tissue sodium concentration: Cross-vendor repeatability and reproducibility of 23Na-MRI across two sites10.1002/jmri.26705 ↗
11. Meyer, M.M., Haneder, S., Konstandin, S., Budjan, J., Morelli, J.N., Schad, L.R., Kerl, H.U., Schoenberg, S.O., and Kabbasch, C. (2019). Repeatability and reproducibility of cerebral 23Na imaging in healthy subjects. BMC Med. Imaging, 19.10.1186/s12880-019-0324-6 ↗
12. Recent technical developments and clinical research-applications of sodium (23Na) MRI10.1016/j.pnmrs.2023.04.002 ↗
13. Whole body sodium MRI at 0.5 Tesla using surface coil and long echo time sequence10.1007/s00723-019-01142-8 ↗
14. Multinuclear MRI at ultrahigh fields10.1097/RMR.0000000000000201 ↗
15. Wilferth, T., Mennecke, A., and Gast, L. (2022). Quantitative 7 Tesla 23Na MRI of the human brain using a 32–channel phased-array head coil: Application to patients with secondary progressive multiple sclerosis. NMR Biomed., 35.10.1002/nbm.4806 ↗
16. Design of a nested eight-channel sodium and four-channel proton coil for 7T knee imaging10.1002/mrm.24432 ↗
17. An RF coil design to enable quintuple nuclear whole-brain MRI10.1002/mrm.29577 ↗
18. Tissue sodium concentration and sodium T1 mapping of the human brain at 3 T using a Variable Flip Angle method10.1016/j.mri.2019.01.015 ↗
19. 3D sodium (23Na) magnetic resonance fingerprinting for time-efficient relaxometric mapping10.1002/mrm.28873 ↗
20. High-resolution sodium imaging of human brain at 7 T10.1002/mrm.23225 ↗
21. Zaric, O., Beiglböck, H., Janacova, V., Szomolanyi, P., Wolf, P., Krebs, M., Trattnig, S., Krššák, M., and Juras, V. (2022). Repeatability assessment of sodium (23Na) MRI at 7.0 T in healthy human calf muscle and preliminary results on tissue sodium concentrations in subjects with Addison’s disease. BMC Musculoskelet. Disord., 23.10.1186/s12891-022-05879-5 ↗
22. Improving sodium Magnetic Resonance in humans by design of a dedicated 23Na surface coil10.1016/j.measurement.2014.01.006 ↗
23. Sodium MRI of the Human Kidney at 3 Tesla10.1002/mrm.21031 ↗
24. Yoon, J.-S., Kim, J.-M., Chung, H.-J., Jeong, Y.-J., Jeong, G.-W., Park, I., Kim, G.-W., and Oh, C.-H. (2021). Development of a proton-frequency-transparent birdcage radiofrequency coil for in vivo 13C MRS/MRSI study in a 3.0 T MRI system. Appl. Sci., 11.10.3390/app112311445 ↗
25. Design and simulation of a dual-tuned 1H/23Na birdcage coil for MRS studies in human calf10.1007/s00723-015-0720-1 ↗
26. Design and construction of an actively frequency-switchable RF coil for field-dependent Magnetisation Transfer Contrast MRI with fast field-cycling10.1016/j.jmr.2010.08.018 ↗
27. Design and construction of a novel 1H/19F double-tuned coil system using PIN-diode switches at 9.4 T10.1016/j.jmr.2017.04.005 ↗
28. Dual-frequency, dual-quadrature, birdcage RF coil design with identical B1 pattern for sodium and proton imaging of the human brain at 1.5 T10.1002/mrm.1910380507 ↗
29. Experimentally verified, theoretical design of dual-tuned, low-pass birdcage radiofrequency resonators for magnetic resonance imaging and magnetic resonance spectroscopy of human brain at 3.0 Tesla10.1002/(SICI)1522-2594(199902)41:2<268::AID-MRM9>3.0.CO;2-G ↗
30. Design and performance of a double-tuned bird-cage coil
31. Two configurations of the four-ring birdcage coil for 1H imaging and 1H-decoupled 31P spectroscopy of the human head10.1006/jmrb.1994.1017 ↗
32. Double-frequency birdcage volume coils for 4.7 T and 7T10.1002/cmr.b.20038 ↗
33. Construction and evaluation of a switch-tuned 13C-1H birdcage radiofrequency coil for imaging the metabolism of hyperpolarized 13C-enriched compounds10.1002/jmri.24458 ↗
34. A design for a double-tuned birdcage coil for use in an integrated MRI/MRS examination10.1016/0022-2364(90)90160-B ↗
35. Optimization and comparison of two practical dual-tuned birdcage configurations for quantitative assessment of articular cartilage with sodium magnetic resonance imaging
36. Numerical solution of initial boundary value problems involving Maxwell’s equations in isotropic media10.1109/TAP.1966.1138693 ↗
37. (2023, September 28). Sim4Life, ZMT Zurich MedTech AG. Available online: https://zmt.swiss/sim4life/.
38. Kumar, S., Chung, H.-J., Jeong, Y.-J., Lee, H.-K., and Oh, C.-H. (2021). Design and Implementation of Split-Leg Type Elliptical Whole-Body Birdcage RF Coil at 1.5 T MRI. Appl. Sci., 16.10.3390/app11167448 ↗
39. BirdcageBuilder: Design of specified-geometry birdcage coils with desired current pattern and resonant frequency10.1002/cmr.10030 ↗
40. Christ, A., Kainz, W., Hahn, E.G., Honegger, K., Zefferer, M., Neufeld, E., Rascher, W., Janka, R., Bautz, W., and Chen, J. (2009). The Virtual Family—Development of surface-based anatomical models of two adults and two children for dosimetric simulations. Phys. Med. Biol., 55.10.1088/0031-9155/55/2/N01 ↗
41. Common mode signal rejection methods for MRI: Reduction of cable shield currents for high static magnetic field systems
42. Oh, C.-H., Jeong, Y.-J., and Kumar, S. Construction Method and Signal Acquisition Method for Dual-Mode Radio Frequency Coil to Acquire Magnetic Resonance Signals from Two Kinds of Nuclei. Korea Patent, Pending.
43. Actual flip-angle imaging in the pulsed steady state: A method for rapid three-dimensional mapping of the transmitted radiofrequency field10.1002/mrm.21120 ↗
44. Whole-brain imaging with receive-only multichannel top-hat dipole antenna RF coil at 7 T MRI10.1007/s40042-021-00334-5 ↗
45. Saturated double-angle method for rapid B1+ mapping10.1002/mrm.20896 ↗
46. (2023, September 28). ISMRM Best Practices for Safety Testing of Experimental RF Hardware. Available online: https://www.ismrm.org/working-groups/best-practices-for-safety-testing-of-experimental-rf-hardware/.
47. Steensma, B.R., Sadeghi-Tarakameh, A., Meliadò, E.F., van den Berg, C.A., Klomp, D.W., Luijten, P.R., Metzger, G.J., Eryaman, Y., and Raaijmakers, A.J. (2023). Tier-based formalism for safety assessment of custom-built radio-frequency transmit coils. NMR Biomed., 36.10.1002/nbm.4874 ↗
48. Jackson, E., Bronskill, M., Drost, D., Och, J., Pooley, R., Sobol, W., and Clarke, G. (2010). AAPM Report MR TG-1: Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities, American Association of Physicists in Medicine.10.37206/101 ↗
49. (2015). Medical Electrical Equipment—Part 2–33: Particular Requirements for the Basic Safety and Essential Performance of Magnetic Resonance Equipment for Medical Diagnosis (Standard No. IEC 60601-2-33:2010).
50. Simultaneous proton magnetic resonance fingerprinting and sodium MRI10.1002/mrm.28073 ↗
51. Sodium MRI: Methods and applications10.1016/j.pnmrs.2014.02.001 ↗