Low-Noise, Low-Power Readout IC for Two-Electrode ECG Recording Using Common-Mode Charge Pump for Robust 20-VPP Common-Mode Interference
Source abstract
Abstract
A low-noise and -power readout integrated circuit (IC) for two-electrode electrocardiogram (ECG) recording is developed in this study using a common-mode charge pump (CMCP) for a robust 20-VPP common-mode interference (CMI). Two-electrode ECG recording offers more comfort than three-electrode ECG recording. Contrasting to the three-electrode ECG recording, the two-electrode ECG recording is affected by CMI during measurements; the intervention of a large CMI will distort the ECG signal measurement. To achieve robustness for the CMI, the proposed ECG readout IC adopts CMCP—it uses switched capacitors that store and subtract CMI by control logic. In this paper, a window comparator structure is applied to CMCP to obtain a signal with less distortion. The window voltage ranges were set between the input common-mode ranges in which IA can operate. Therefore, a signal with less distortion was obtained by stopping the operation of CMCP between the window voltage ranges. It also reduced additional current consumption. To achieve this, the proposed circuit is implemented using a chopper stabilization technique. The chopper implemented in the amplifier can reduce low-frequency noise components, such as 1/f noise, and it comprises a CMCP, current feedback instrumentation amplifier, QRS peak detector, relaxation oscillator, voltage reference, timing generator, and serial peripheral interface on a single chip. The proposed circuit was designed using a standard 0.18 μm CMOS process with an active area of 0.54 mm2. The proposed CMCP achieves a CMI robustness of 20 VPP at 60 Hz. The measured input-referred noise level was 119 nV/√Hz at 1 Hz, and the power consumption was 23.83 μW with a 1.8 V power supply.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Analog and Mixed-Signal Circuit Design · ECG Monitoring and Analysis · Neuroscience and Neural Engineering
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Electronics Engineering, Chungnam National University, Daejeon 34134, Republic of Korea
Funding
Ministry of Health and Welfare · MSIT (Ministry of Science and ICT) · National Research Foundation of Korea (NRF) · ABOV semiconductor, Inc.
References (31)
Show source-provided references
1. Prognostic Value of ECG findings for total, cardiovascular disease, and Coronary Heart Disease Death in men and Women10.1136/hrt.80.6.570 ↗
2. A New Method to Identify Coronary Artery Disease with ECG signals and Time-Frequency Concentrated Antisymmetric Biorthogonal Wavelet Filter Bank10.1016/j.patrec.2019.04.014 ↗
3. Portable Out-of-Hospital Electrocardiography: A Review of Current Technologies10.1002/joa3.12035 ↗
4. A 24.88-μW Biopotential Amplifier Tolerant to 15-VPP Common-Mode Interference for Two-Electrode ECG Recording in 180-nm CMOS10.1109/JSSC.2020.3005768 ↗
5. Koo, N., and Cho, S. (2019, January 17–21). A 27.8 μW Biopotential Amplifier Tolerant to 30 VPP Common-Mode Interference for Two-electrode ECG Recording in 0.18 μm CMOS. Proceedings of the IEEE International Solid-State Circuits Conference (ISSCC), San Francisco, CA, USA.10.1109/ISSCC.2019.8662373 ↗
6. Ground-Free ECG Recording with Two Electrodes10.1109/TBME.1980.326595 ↗
7. Jedari, E., Rashidzadeh, R., Mirhassani, M., and Ahmadi, M. (May, January 30). Two-Electrode ECG Measurement Circuit Using a Feed Forward CMRR Enhancement Method. Proceedings of the IEEE 30th Canadian Conference on Electrical and Computer Engineering (CCECE), Windsor, ON, Canada.
8. You, D., Heo, H., Kim, H., Kwon, Y., Lee, S., and Ko, H. (2020). Power Line Interference Reduction Technique with a Current-Reused Current-Feedback Instrumentation Amplifier for ECG Recording. Appl. Sci., 10.10.3390/app10238478 ↗
9. Reduction of Interference Due to Common Mode Voltage in Biopotential Amplifiers10.1109/TBME.1983.325167 ↗
10. Sun, X. (2021, January 22–24). Low Noise, High Accuracy Analog Electrocardiogram (ECG) Signal Front End Amplifier for Wearable Equipment. Proceedings of the 2021 International Conference on Electronics, Circuits and Information Engineering (ECIE), Zhengzhou, China.10.1109/ECIE52353.2021.00025 ↗
11. Chen, L., He, J., and Ge, F. (2021, January 23–25). A Low Power Low Noise ECG Amplifier Design Based on OTA. Proceedings of the 2021 3rd International Conference on Intelligent Control, Measurement and Signal Processing and Intelligent Oil Field (ICMSP), Xi’an, China.10.1109/ICMSP53480.2021.9513222 ↗
12. A Neuromorphic Processing System with Spike-Driven SNN Processor for Wearable ECG Classification10.1109/TBCAS.2022.3189364 ↗
13. An Electronic Neuromorphic System for Real-Time Detection of High Frequency Oscillations (HFO) in Intracranial EEG10.1038/s41467-021-23342-2 ↗
14. Automatic diagnosis of the 12-lead ECG using a deep neural network10.1038/s41467-020-15432-4 ↗
15. Scalable Digital Neuromorphic Architecture for Large-Scale Biophysically Meaningful Neural Network with Multi-Compartment Neurons10.1109/TNNLS.2019.2899936 ↗
16. Neuromorphic Context-Dependent Learning Framework with Fault-Tolerant Spike Routing10.1109/TNNLS.2021.3084250 ↗
17. BiCoSS: Toward Large-Scale Cognition Brain with Multigranular Neuromorphic Architecture10.1109/TNNLS.2020.3045492 ↗
18. CerebelluMorphic: Large-Scale Neuromorphic Model and Architecture for Supervised Motor Learning10.1109/TNNLS.2021.3057070 ↗
19. Indirect Current Feedback Instrumentation Amplifier with a Common-Mode Input Range that Includes the Negative Roll10.1109/4.222171 ↗
20. A Chopper Stabilized Current-Feedback Instrumentation Amplifier for EEG Acquisition Applications10.1109/ACCESS.2019.2892502 ↗
21. A Chopper Current-Feedback Instrumentation Amplifier with a 1 mHz 1/f Noise Corner and an AC-Coupled Ripple Reduction Loop10.1109/JSSC.2009.2032710 ↗
22. Wu, R., Makinwa, K.A.A., and Huijsing, J.H. (2008, January 18–21). The Design of a Chopped Current-Feedback Instrumentation Amplifier. Proceedings of the 2008 IEEE International Symposium on Circuits and Systems (ISCAS), Seattle, WA, USA.
23. Dry-Contact and Noncontact Biopotential Electrodes: Methodological Review10.1109/RBME.2010.2084078 ↗
24. Chopper-Stabilized Low-Noise Multipath Operational Amplifier with Dual Ripple Rejection Loops
25. A 30 μW Analog Signal Processor ASIC for Portable Biopotential Signal Monitoring10.1109/JSSC.2010.2085930 ↗
26. Kwon, Y., Kim, H., You, D., Heo, H., Ko, H., and Lee, S. (2020, January 21–24). A 28.4 nV/√Hz Chopper Stabilized Current Feedback Instrumentation Amplifier with Auto Offset Calibration DAC for Resistive Bridge Sensor. Proceedings of the 2020 International SoC Design Conference (ISOCC), Yeosu, South Korea.10.1109/ISOCC50952.2020.9332997 ↗
27. Kim, J., Kim, H., Han, K., You, D., Heo, H., Kwon, Y., Cho, D.-I., and Ko, H. (2020). Low-Noise Chopper-Stabilized MultiPath Operational Amplifier with Nested Miller Compensation for High-Precision Sensors. Appl. Sci., 10.10.3390/app10010281 ↗
28. Design Methodology and Advances in Nested-Miller Compensation10.1109/TCSI.2002.800463 ↗
29. A Configurable and Low-Power Mixed Signal SoC for Portable ECG Monitoring Applications10.1109/TBCAS.2013.2260159 ↗