journal article·2022-07-25Open access

Biomechanical Assessment of Cannulated Nails for the Treatment of Proximal Femur Fractures

Karel Frydrýšek, Tomáš Halo, Daniel Čepica, Vojtěch Machalla, Kateřina Šimečková, Ondřej Skoupý, Roman Madeja, Miroslav Havlíček, Kamila Dostálová, Antonín Trefil, Leopold Pleva, Zuzana Murčinkova, Pavel Krpec, Josef Hlinka

Applied Sciences· Volume 12· Issue 15

Abstract

This article focuses on a type of surgical implant used in orthopaedics and traumatology—cannulated femoral nails. Femoral nails are used in medical treatment for purposes of osteosynthesis, i.e., when treating various types of complicated fractures, in this case fractures of the femur. The article investigates cases in which a nail has been implanted in the proximal part of the femur for a short time (with the fracture still not healed), compared with cases in which the bone has already healed. According to AO classification, examined fractures are described as AO 31B3 AO 32A3. The main focus is on strength-deformation analysis using the finite element method (FEM), which makes it possible to determine the behaviour of the femur-implant system. FEM analysis was used to compare 1.4441 steel nails made by two manufacturers, Medin (Czech Republic) and Tantum (Germany). Boundary conditions including external loading, prescribed supports and elastic foundation are defined. There were solved FEM analyses for five cases of healed femur and five cases of broken femur both including implants with prescribed collo-diaphyseal angles. The results of the analysis were used to assess stress-deformation states from the perspective of appropriateness for clinical treatment, biomechanical reliability and safety. All examined femoral nails are compared, safe and suitable for patient treatment.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Bone fractures and treatments · Hip and Femur Fractures

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Faculty of Mechanical Engineering, VSB—Technical University of Ostrava, 17. listopadu 15/2172, 70800 Ostrava-Poruba, Czech Republic

Faculty of Medicine, University of Ostrava, Syllabova 19, 70300 Ostrava-Vítkovice, Czech Republic

Trauma Centre, University Hospital Ostrava, 17. listopadu 1790, 70852 Ostrava-Poruba, Czech Republic

Medin, a.s, Vlachovicka 619, 59231 Nové Město na Moravě, Czech Republic

Centre for Advanced Innovative Technologies, VSB—Technical University of Ostrava, 17. listopadu 15/2172, 70800 Ostrava-Poruba, Czech Republic

Department of Design and Monitoring of Technical Systems, Technical University of Košice, Letná 9, 04200 Košice, Slovakia

V-NASS, a.s., Halasova 2938/1a, 70300 Ostrava-Vítkovice, Czech Republic

Faculty of Materials Science and Technology, VSB—Technical University of Ostrava, 17. listopadu 15/2172, 70800 Ostrava-Poruba, Czech Republic

Funding

Ministry of Education Youth and Sports · European Union and Government of the Czech Republic

References (29)

Show source-provided references
1. Strength Analyses of Screws for Femoral Neck Fractures10.1007/s40846-018-0378-x ↗
2. Frydrýšek, K., Šír, M., Pleva, L., Szeliga, J., Stránský, J., Čepica, D., Kratochvíl, J., Koutecký, J., Madeja, R., and Dědková, K.P. (2022). Stochastic Strength Analyses of Screws for Femoral Neck Fractures. Appl. Sci., 12.10.3390/app12031015 ↗
3. Fracture and Dislocation Classification Compendium—201810.1097/BOT.0000000000001063 ↗
4. Proximal femoral fractures: Principles of management and review of literature10.1016/j.jcot.2012.04.001 ↗
5. The treatment of trochanteric fractures of the femur10.1302/0301-620X.31B2.190 ↗
6. Low-Angle Fixation in Fractures of the Femoral Neck10.1302/0301-620X.43B4.647 ↗
7. Locking Compression Plate for Distal Tibia Fractures10.4028/www.scientific.net/AMM.827.359 ↗
8. Stýskala, V., Kolosov, D., Snášel, V., Karakeyev, T., and Abraham, A. (2016). Finite element modelling of T-plate for treatment of distal radius. Intelligent Systems for Computer Modelling. Advances in Intelligent Systems and Computing, Springer.10.1007/978-3-319-27644-1 ↗
9. The history of internal fixation of proximal femur fractures Ernst Pohl—The genius behind10.1007/s00264-014-2320-3 ↗
10. Treatment of Complex Proximal Femoral Fractures with the Proximal Femur Locking Compression Plate10.3928/01477447-20070801-18 ↗
11. The AO/ASIF-proximal femoral nail (PFN): A new device for the treatment of unstable proximal femoral fractures10.1016/S0020-1383(99)00091-1 ↗
12. Finite Element Modelling and Simulation for Lower Limb of Human Bone: A Review
13. Jamari, J., Ammarullah, M., Saad, A., Syahrom, A., Uddin, M., van der Heide, E., and Basri, H. (2021). The Effect of Bottom Profile Dimples on the Femoral Head on Wear in Metal-on-Metal Total Hip Arthroplasty. J. Funct. Biomater., 12.10.3390/jfb12020038 ↗
14. Analysis of Locking Self-Taping Bone Screws for Angularly Stable Plates10.1007/s40846-017-0279-4 ↗
15. Drápala, J., Kostiuková, G., and Losertová, M. (2018, January 23–25). Influence of heat treatment on microstructure and mechanical properties of SUS 316L alloy. Proceedings of the 27th International Conference on Metallurgy and Materials (METAL), Brno, Czech Republic.
16. Hlinka, J., Dostalova, K., Dedkova, K.P., Madeja, R., Frydrysek, K., Koutecky, J., Sova, P., and Douglas, T.E.L. (2021). Complex Material and Surface Analysis of Anterolateral Distal Tibial Plate of 1.4441 Steel. Metals, 12.10.3390/met12010060 ↗
17. Measurement of Collo-Diaphyseal Angle and Femoral Neck Anteversion Angle of Femur Bone
18. MEDIN (2018). Instruments and Implants for Traumatology, MEDIN. Available online: https://www.medin.cz/media/cache/file/ce/medin-traumatology-catalogue-2018-10-CS-EN_LQ.pdf.
19. Tantum (2022, April 21). Systemübersicht PLATON. Available online: https://tantum-ag.de/wp-content/uploads/2018/09/b_02_11.1_platon_broschuere.pdf.
20. Ansys (2022, April 21). Ansys Workbench 2020 R2. Available online: https://www.ansys.com/products/ansys-workbench.
21. The fixion proximal femur nailing system: Biomechanical properties of the nail and a cadaveric study10.1016/j.jbiomech.2004.03.014 ↗
22. Biomechanical comparison of the femoral neck system and the dynamic hip screw in basicervical femoral neck fractures10.1038/s41598-022-11914-1 ↗
23. A comparative biomechanical study of proximal femoral nail (InterTAN) and proximal femoral nail antirotation for intertrochanteric fractures10.1007/s00264-013-2120-1 ↗
24. Frydrýšek, K., Čepica, D., Halo, T., Skoupý, O., Pleva, L., Madeja, R., Pometlová, J., Losertová, M., Koutecký, J., and Michal, P. (2022). Biomechanical Analysis of Staples for Epiphysiodesis. Appl. Sci., 12.10.3390/app12020614 ↗
25. Mechanical validation of whole bone composite femur models10.1016/0021-9290(95)00084-4 ↗
26. Finite Element Analysis of the Proximal Femur and Hip Fracture Risk in Older Men10.1359/jbmr.081201 ↗
27. Experimental validation of finite element model for proximal composite femur using optical measurements10.1016/j.jmbbm.2013.02.006 ↗
28. Reconstruction of human femur bone from CT scan images using CAD techniques10.1088/1757-899X/455/1/012103 ↗
29. Materialise (2022, April 21). Mimics Innovation Suite. Medical Image Analysis Software. Available online: https://www.materialise.com/en.