Three-Line Microstrip Array for Whole-Body MRI System at 7 T
Source abstract
Abstract
This paper proposes the use of a triple-line microstrip array for transmitting a magnetic field (|B1+|) into the whole body for magnetic resonance applications at ultra-high field strength, such as 7 T. We explored some technologies that can potentially be applied for whole-body 7 T magnetic resonance imaging, as there is ongoing research on this topic. The triple-line microstrip transmission line (t-MTL) array consists of 32 channels. Each channel has a t-MTL, comprising a main conductor line and two adjacent coupled lines. The adjacent lines are not connected directly to the source. This configuration resulted in increased intensity and a centered |B1+|-field. We compared the proposed structure and some reference radiofrequency (RF) transmitters, such as a patch antenna, using a magnet bore as a waveguide and a whole-body birdcage coil. We evaluated the performance of the t-MTL using cylindrical phantoms. We computed the |B1+|-field from each RF transmitter inside a 3D human model containing more than 200 tissues. We compared their uniformity and field intensity and proposed a t-MTL array that yielded better performance. The proposed design also showed a lower specific absorption rate compared with a patch antenna.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced MRI Techniques and Applications · Wireless Body Area Networks · Electromagnetic Fields and Biological Effects
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Medical Campus, Department of Biomedical Engineering, Gachon University, Incheon 21936, Korea
Funding
National Research Foundation of Korea · ministry of health & welfare the Republic of Korea
References (26)
Show source-provided references
1. Clinical applications of 7T MRI in the brain10.1016/j.ejrad.2011.07.007 ↗
2. Musculoskeletal MRI at 3.0 T and 7.0 T: A comparison of relaxation times and image contrast10.1016/j.ejrad.2011.09.021 ↗
3. MRI based medical image analysis: Survey on brain tumor grade classification10.1016/j.bspc.2017.07.007 ↗
4. Comparison of RF body coils for MRI at 3T: A simulation study using parallel transmission on various anatomical targets10.1002/nbm.3378 ↗
5. Orzada, S., Solbach, K., Gratz, M., Brunheim, S., Fiedler, T.M., Johst, S., Bitz, A.K., Shooshtary, S., Abuelhaija, A., and Voelker, M.N. (2019). A 32-channel parallel transmit system add-on for 7 T MRI. PLoS ONE, 14.10.1371/journal.pone.0222452 ↗
6. Shielded microstrip array for 7T human MR imaging
7. A 16-channel combined loop-dipole transceiver array for 7 Tesla body MRI10.1002/mrm.26153 ↗
8. Whole-body magnetic resonance imaging (WB-MRI) for cancer screening in asymptomatic subjects of the general population: Review and recommendations10.1186/s40644-020-00315-0 ↗
9. Whole-body magnetic resonance imaging (WB-MRI) in oncology: Recommendations and key uses10.1007/s11547-018-0955-7 ↗
10. Whole-body MRI: Current applications in oncology10.2214/AJR.17.17984 ↗
11. Technical assessment of whole body angiography and cardiac function within a single MRI examination10.1016/j.crad.2015.02.003 ↗
12. Efficient high-frequency body coil for high-field MRI10.1002/mrm.20177 ↗
13. Whole body traveling wave magnetic resonance imaging at high field strength: Homogeneity, efficiency, and energy deposition as compared with traditional excitation mechanisms10.1002/mrm.23107 ↗
14. Coaxial waveguide for travelling wave MRI at ultrahigh fields10.1002/mrm.24496 ↗
15. Parallel traveling-wave MRI: A feasibility study10.1002/mrm.23073 ↗
16. Pang, Y., Wu, B., Jiang, X., Vigneron, D.B., and Zhang, X. (2014). Tilted microstrip phased arrays with improved electromagnetic decoupling for ultrahigh-field magnetic resonance imaging. Medicine, 93.10.1097/MD.0000000000000311 ↗
17. Analysis of an integrated 8-channel Tx/Rx body array for use as a body coil in 7-Tesla MRI10.3389/fphy.2017.00017 ↗
18. Design equations for coupled microstrip lines10.1080/00207217908938683 ↗
19. The design and performance of three-line microstrip couplers10.1109/TMTT.1976.1128928 ↗
20. On the analysis of symmetrical three-line microstrip circuits10.1109/TMTT.1977.1129202 ↗
21. Accurate wide-range design equations for the frequency-dependent characteristic of parallel coupled microstrip lines10.1109/TMTT.1984.1132616 ↗
22. Hammerstad, E., and Jensen, O. (1980, January 28). Accurate Models for Microstrip Computer-Aided Design. Proceedings of the 1980 IEEE MTT-S International Microwave Symposium Digest, Washington, DC, USA.
23. IT’IS Foundation (2016). Tissue Properties Database V3.1, IT’IS Foundation.
24. Christ, A., Kainz, W., Hahn, E.G., Honegger, K., Zefferer, M., Neufeld, E., Rascher, W., Janka, R., Bautz, W., and Chen, J. (2009). The Virtual Family—Development of surface-based anatomical models of two adults and two children for dosimetric simulations. Phys. Med. Biol., 55.10.1088/0031-9155/55/2/N01 ↗
25. Specific absorption rate implications of within-scan patient head motion for ultra-high field MRI10.1002/mrm.28276 ↗
26. Intersubject local SAR variation for 7T prostate MR imaging with an eight-channel single-side adapted dipole antenna array10.1002/mrm.24794 ↗