Study on the Improvement of Electrical Facility System of Automated External Defibrillators by Real-Time Measurement of Thoracic Impedance

Tae-Jin Ha, Hong-Gyu Park, Su-Kang Park, Sang-Geon Park

Applied Sciences· Volume 10· Issue 9

Abstract

Sudden Cardiac Arrest (SCA) is a serious emergency disease that has increased steadily every year. To this end, an Automated External Defibrillator (AED) is placed in a public place so that even non-professional medical personnel can respond to SCA. However, the thoracic impedance of patients changes due to CardioPulmonary Resuscitation (CPR) and artificial respiration during first aid treatment. In addition, changes in chest statues due to gender, age, and accidents cause changes in thoracic impedance in real time. The change in thoracic impedance caused by this has a negative effect on the intended electrical energy of the automatic heart shocker to the emergency patient. To prove this, we divided it into adult and pediatric modes and experimented with the energy error of the AED according to the same impedance change. When the first peak current was up to 56.4 (A) and at least 8.4 (A) in the adult mode, the first peak current was up to 32.2 (A) and at least 4.8 (A), respectively, when the impedance changed, the error of the current figure occurred. In this paper, the inverse relationship between thoracic impedance and electric shock energy according to the state of the cardiac arrest patient is demonstrated through the results of the experiment, and the need for an electric facility system that can revise for changes in thoracic impedance of the cardiac arrest patient by reflecting them on electric shock energy in real time is presented.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Cardiac Arrest and Resuscitation · Non-Invasive Vital Sign Monitoring · Healthcare Technology and Patient Monitoring

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Electrical and Electronic Engineering, Silla University, 140 Baegyang-daero 700beon-gil, Sasang-gu, Busan 46958, Korea

Department of Electrical, Electronic Control Engineering, Changwon National University, 20 Changwondaehak-ro, Uichang-gu, Changwon-si 51140, Gyeongsangnam-do, Korea

Department of Korea Testing Certification, 22 Heungan-daero 27 beon-gil, Gunpo, Gyeonggi-do 15809, Korea

Department of Mechatronics Convergence, Changwon National University, 20 Changwondaehak-ro, Uichang-gu, Changwon-si, Gyeongsangnam-do 51140, Korea

References (28)

Show source-provided references
1. Kim, Y.T. (2017). Sudden Cardiac Arrest Survey 2006–2016.
2. Real-time monitoring of blood flow using mechanical impedance of chest during CPR
3. Chest compression-only cardiopulmonary resuscitation for out-of-hospital cardiac arrest with public-access defibrillation10.1161/CIRCULATIONAHA.112.109504 ↗
4. Outcomes of sudden cardiac arrest in a state-wide integrated resuscitation program: Results from the minnesota resuscitation consortium10.1016/j.resuscitation.2016.10.029 ↗
5. Analysis of bystander CPR quality during out-of-hospital cardiac arrest using data derived from automated external defibrillators10.1016/j.resuscitation.2018.05.012 ↗
6. Pediatric sudden cardiac arrest10.1542/peds.2012-0144 ↗
7. Warning symptoms are associated with survival from sudden cardiac arrest10.7326/M14-2342 ↗
8. Defibrillation energy dose during pediatric cardiac arrest: Systematic review of human and animal model studies10.1016/j.resuscitation.2019.04.028 ↗
9. Study on safety requirements and test methods for automated external defibrillators
10. Comparison of defibrillation efficacy between two pads placements in a pediatric porcine model of cardiac arrest10.1016/j.resuscitation.2011.12.010 ↗
11. Piuzzi, E., Pisa, S., Pittella, E., Podestà, L., and Sangiovanni, S. (2017, January 22–25). Comparison among low-cost portable systems for thoracic impedance plethysmography. Proceedings of the 2017 IEEE International Instrumentation and Measurement Technology Conference, Torino, Italy.10.1109/I2MTC.2017.7969747 ↗
12. A study on the electrical difference for the limbs and thoracic impedance using real-time bio-impedance measurement system
13. Park, J.Y. (2015). Enhancement of Automatic Defibrillator Simulated Algorithm by ECG and Chest Impedance Signal Analysis. [Master’s Thesis, Yonsei University].
14. Circulation detection using the electrocardiogram and the thoracic impedance acquired by defibrillation pads10.1016/j.resuscitation.2015.11.014 ↗
15. Thoracic-impedance changes measured via defibrillator pads can monitor signs of circulation10.1016/j.resuscitation.2006.10.001 ↗
16. Measurement method and principle for thoracic impedance graph
17. Reliability, validity, and diagnostic value of a pediatric bioelectrical impedance analysis scale10.1089/chi.2014.0156 ↗
18. Automatic detection of chest compression pauses using the transthoracic impedance signal
19. Reliable extraction of the circulation component in the thoracic impedance measured by defibrillation pads10.1016/j.resuscitation.2013.05.020 ↗
20. Bioelectrical impedance as a measure of change in body composition in young children10.1111/ijpo.263 ↗
21. Circulation assessment by automated external defibrillators during cardiopulmonary resuscitation10.1016/j.resuscitation.2018.04.036 ↗
22. An accurate method for real-time chest compression detection from the impedance signal10.1016/j.resuscitation.2016.04.023 ↗
23. Real-time feedback systems in CPR10.1016/j.tacc.2012.09.004 ↗
24. Mechanical impedance monitoring system of the patient’s chest during cardiopulmonary resuscitation10.3795/KSME-B.2018.42.9.583 ↗
25. Berve, P.O., Irusta, U., Kramer-Johansen, J., Skålhegg, T., Kongsgård, H.W., Brunborg, C., Aramendi, E., and Wik, L. (2019). Transthoracic impedance measured with defibrillator pads—New interpretations of signal change induced by ventilations. Clin. Med., 8.10.3390/jcm8050724 ↗
26. Monitoring mechanical impedance of the thorax with compression and decompression cardiopulmonary resuscitation device10.1007/s12206-019-0155-y ↗
27. Anterior-posterior thoracic force-deflection characteristics measured during cardiopulmonary resuscitation: Comparison to post-mortem human subject data
28. (2010). Medical Device Performance Test Guidance 2010.