journal article·2024-11-07Open access

Image Quality and Information Parameters of Electronic Portal Imaging Devices

Marios K. Tzomakas, Vasiliki Peppa, Antigoni Alexiou, Georgios Karakatsanis, Anastasios Episkopakis, Christos Michail, Ioannis Valais, George Fountos, Nektarios Kalyvas, Ioannis S. Kandarakis

Applied Sciences· Volume 14· Issue 22

Abstract

In this study, the imaging performance of electronic portal imaging devices (EPIDs) is evaluated, comparing measurements collected from EPID images captured at 115 cm, with a field size of 15 × 15 cm2, monitor units (MUs) in the range of 2 MU-100 MU and dose rates (DRs) of 200 MU/min, 400 MU/min and 600 MU/min, using a 6 MV LINAC system and the QC-3V image quality phantom. The analysis includes the normalized contrast transfer function (CTFnorm), the noise power spectrum (NPS) and the information capacity (IC), as well as the signal-to-noise frequency response (SNFR), which can be used as a comprehensive quality index. The results of our study are compared with previously published data captured at 100 cm under similar exposure conditions. They show similar CTF curves with different source-to-phantom distances, with the lowest values observed at specific MU and DR combinations. Moreover, NPS graphs are found to decrease with respect to spatial frequency. SNFR values also display a reduction with increasing spatial frequency. In addition, irradiation with the phantom placed closer to the EPID, 115 cm from the LINAC, yields better SNFR and IC performance characteristics, indicating better delineation of the organs closer to the EPID. The testing of EPID performance may potentially benefit from our results, which may lead to improvements in the quality of radiotherapy treatments.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Medical Imaging Techniques and Applications · Advanced Radiotherapy Techniques · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Radiation Physics, Materials Technology and Biomedical Imaging Laboratory, Department of Biomedical Engineering, University of West Attica, 12210 Athens, Greece

Department of Radiotherapy, General Hospital of Athens Alexandra, 11528 Athens, Greece

Elekta, 15124 Athens, Greece

Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, 11527 Athens, Greece

References (27)

Show source-provided references
1. Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy10.17946/JRST.2015.38.4.16 ↗
2. Task Group 142 report: Quality Assurance of Medical Accelerators10.1118/1.3190392 ↗
3. A phantom based evaluation of the clinical imaging performance of electronic portal imaging devices10.1016/j.heliyon.2023.e21116 ↗
4. A quality assurance phantom for electronic portal imaging devices10.1120/jacmp.v12i2.3350 ↗
5. On site evaluation of three flat panel detectors for digital radiography10.1118/1.1569273 ↗
6. Performance of electronic portal imaging devices (EPIDs) used in radiotherapy: Image quality and dose measurements10.1118/1.1688212 ↗
7. X-ray quantum limited portal imaging using amorphous silicon flat-panel arrays10.1118/1.598252 ↗
8. Use of a line-pair resolution phantom for comprehensive quality assurance of electronic portal imaging devices based on fundamental imaging metrics10.1118/1.3099559 ↗
9. The information spectrum as a measure of radiographic image quality and system performance10.1088/0031-9155/29/4/001 ↗
10. Michail, C.M., Kalyvas, N.E., Valais, I.G., Fudos, I.P., Fountos, G.P., Dimitropoulos, N., Koulouras, G., Kandris, D., Samarakou, M., and Kandarakis, I.S. (2014). Figure of Image Quality and Information Capacity in Digital Mammography. BioMed Res. Int., 2014.10.1155/2014/634856 ↗
11. A Mathematical Theory of Communication10.1002/j.1538-7305.1948.tb01338.x ↗
12. Michail, C., Karpetas, G., Kalyvas, N., Valais, I., Kandarakis, I., Agavanakis, K., Panayiotakis, G., and Fountos, G. (2018). Information Capacity of Positron Emission Tomograpgy Scanners. Crystals, 8.10.3390/cryst8120459 ↗
13. (2022, December 14). SeeDOS Ltd. The Quality Control Phantom QC-3, Available online: http://www.seedos.co.uk/qc3_phantom.htm.
14. The influence of software filtering in digital mammography image quality10.1088/1748-0221/4/05/P05018 ↗
15. Haus, A.G. (1979). The Physics of Medical Imaging: Recording System, Measurements and Techniques, Wiley.
16. Nill, N.B. (2023, March 07). Conversion Between Sine Wave and Square Wave Spatial Frequency Response of an Imaging System, MITRE. Available online: https://apps.dtic.mil/sti/citations/ADA460454.
17. Boreman, G.D. (2001). Modulation Transfer Function in Optical and Electro-Optical Systems, SPIE.10.1117/3.419857 ↗
18. Sinusoidal modulation analysis for optical system MTF measurements10.1118/1.597840 ↗
19. Podgorsak, E.B. (2005). Radiation Oncology Physics: A Handbook for Teachers And Students, IAEA.
20. De Moura, F.M.G. (2008). Amorphous Silicon Detector Panel Damage: Correlating Physical Parameters to Clinical Usability. [Master’s Thesis, Universidade de Lisboa]. Available online: https://repositorio.ul.pt/handle/10451/1277?locale=en.
21. Quantitative and Qualitative Comparison of Standard-Dose and Low-Dose Pediatric Head Computed Tomography: A Retrospective Study Assessing the Effect of Adaptive Statistical Iterative Reconstruction10.1097/RCT.0b013e31828426de ↗
22. Noise power spectra of images from digital mammography detectors10.1118/1.598623 ↗
23. Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device10.1118/1.1637969 ↗
24. Development of a novel high quantum efficiency MV x-ray detector for image-guided radiotherapy: A feasibility study10.1002/mp.13900 ↗
25. Model of the spatial-frequency-dependent detective quantum efficiency of phosphor screens10.1118/1.596583 ↗
26. Application of information theory to the assessment of computed tomography10.1118/1.594559 ↗
27. In vivo dosimetry in external beam photon radiotherapy: Requirements and future directions for research, development, and clinical practice10.1016/j.phro.2020.08.003 ↗