journal article·2021-11-23Open access

Biomechanical Analysis of Clavicle Hook Plates with a Range of Posterior Hook Offsets Implanted at Different Acromion Positions in the Acromioclavicular Joint: A Finite Element Analysis Study

Li-Kun Hung, Cheng-Hung Lee, Kuo-Chih Su

Applied Sciences· Volume 11· Issue 23

Abstract

The clavicle hook plate is commonly used in acromioclavicular injuries; however, the biomechanical effect of the posterior hook offset and hook position is unclear. This study applied a finite element analysis (FEA) to evaluate these parameters to improve the clinical strategy. Nine FEA models with 0-mm, 5-mm, and 10-mm posterior hook offsets implanted in the anterior, middle, and posterior acromion were established to evaluate the stress distribution and the reaction force on the acromion. The 5-mm and 10-mm posterior hook offsets at all acromion positions reduced the reaction force on the acromion but slightly increased the stress on the clavicle. The 0-mm offset increased the reaction force at all acromion positions and was relatively lower at the middle acromion. The clavicle hook plate with a posterior hook offset reduces the reaction force on the acromion, providing a flexibility of the hook position. These results provide surgeons with the biomechanical basis for the hook offset and position and engineers with the mechanical basis for the implant design.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Shoulder and Clavicle Injuries · Shoulder Injury and Treatment · Trauma Management and Diagnosis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Orthopedics, Taichung Veterans General Hospital, Taichung 407, Taiwan

College of Medicine, National Chung Hsing University, Taichung 402, Taiwan

Department of Food Science and Technology, Hungkuang University, Taichung 433, Taiwan

Department of Medical Research, Taichung Veterans General Hospital, Taichung 407, Taiwan

Department of Biomedical Engineering, Hungkuang University, Taichung 433, Taiwan

Department of Chemical and Materials Engineering, Tunghai University, Taichung 407, Taiwan

References (29)

Show source-provided references
1. The use of hook plate in type III and V acromio-clavicular Rockwood dislocations: Clinical and radiological midterm results and MRI evaluation in 42 patients10.1016/j.injury.2011.04.002 ↗
2. Hook plate fixation of acute displaced lateral clavicle fractures: Mid-term results and a brief literature overview10.1186/1749-799X-7-2 ↗
3. Hook plate fixation for acute acromioclavicular dislocations without coracoclavicular ligament reconstruction: A functional outcome study in military personnel. Strategies10.1007/s11751-015-0228-0 ↗
4. Long-term clinical and radiological results after hook plate osteosynthesis of lateral clavicle fractures10.1097/BOT.0000000000002007 ↗
5. The Radiologic Comparison of Operative Treatment Using a Hook Plate versus a Distal Clavicle Locking Plate of Distal Clavicle Fracture10.5397/cise.2018.21.4.227 ↗
6. Acromioclavicular joint separation treated with clavicular hook plate: A study of radiological and functional outcomes10.1007/s00402-020-03521-4 ↗
7. In vivo analysis of acromioclavicular joint motion after hook plate fixation using three-dimensional computed tomography10.1016/j.jse.2014.12.012 ↗
8. Transacromial erosion of a locked subacromial hook plate: Case report and review of literature10.1016/j.jse.2009.10.019 ↗
9. Acromion osteolysis and fracture after hook plate fixation for acromioclavicular joint dislocation: A case report10.1016/j.jse.2009.12.005 ↗
10. Complications with the clavicle hook plate after fixation of Neer type II clavicle fractures10.1007/s00264-018-4108-3 ↗
11. Clavicular hook plate may induce subacromial shoulder impingement and rotator cuff lesion-dynamic sonographic evaluation10.1186/1749-799X-9-6 ↗
12. Fixation failure of the clavicular hook plate: A report of three cases10.1177/230949900601400320 ↗
13. Rare complication of clavicular hook plate: Clavicle fracture at the medial end of the plate10.1016/j.jse.2011.06.005 ↗
14. Peri-implant Fractures Following Hook Plate Fixation for Unstable Distal Clavicle Fractures10.3928/01477447-20200619-01 ↗
15. Midshaft clavicle fracture following osteosynthesis with a hook plate: A retrospective case analysis10.1007/s00402-020-03397-4 ↗
16. Subacromial morphometric assessment of the clavicle hook plate10.1016/j.injury.2009.12.012 ↗
17. Anatomical research on the subacromial interval following implantation of clavicle hook plates10.1055/s-0034-1367050 ↗
18. Three-dimensional morphological analysis of acromioclavicular joint in patients with and without subacromial erosion after hook plate fixation10.1177/0300060517725973 ↗
19. Evaluation of the Acromioclavicular Joint Morphology for Minimizing Subacromial Erosion after Surgical Fixation of the Joint Using a Clavicular Hook Plate10.5397/cise.2018.21.3.138 ↗
20. Biomechanical analysis of implanted clavicle hook plates with different implant depths and materials in the acromioclavicular joint: A finite element analysis study10.1111/aor.12679 ↗
21. Biomechanical analysis of acromioclavicular joint dislocation treated with clavicle hook plates in different lengths10.1007/s00264-015-2890-8 ↗
22. Biomechanical analysis of clavicle hook plate implantation with different hook angles in the acromioclavicular joint10.1007/s00264-016-3384-z ↗
23. Relationship between the lateral acromion angle and postoperative persistent pain of distal clavicle fracture treated with clavicle hook plate10.1186/s13018-020-01737-z ↗
24. Fifteen-degree clavicular hook plate achieves better clinical outcomes in the treatment of acromioclavicular joint dislocation10.1177/0300060518786910 ↗
25. To explore the reasonable selection of clavicular hook plate to reduce the occurrence of subacromial impingement syndrome after operation10.1186/s13018-021-02325-5 ↗
26. Combined finite element and multibody musculoskeletal investigation of a fractured clavicle with reconstruction plate10.1080/10255842.2013.845175 ↗
27. Lee, H.H. (2014). Finite Element Simulations with ANSYS Workbench 15, Chuan Hwa Book Co.
28. Enderle, J., and Bronzino, J. (2012). Introduction to Biomedical Engineering, Academic Press.10.1016/B978-0-12-374979-6.00001-0 ↗
29. Advantage of Multi-Directional Stability of Spiral Clavicle Plate in Treatment of Middle One-Third Clavicle Fracture: A Finite Element Study10.1007/s40846-021-00635-x ↗